Description
1. Product Presentation
The absorbable multifilament suture made of polyglycolic acid (PGA) in Prime quality, reference P118HR1340 (SKU AMP118HR1340), is the synthetic resorbable solution your operating theatre needs for the approximation of soft tissues, ligatures and general surgery procedures with high safety requirements. Manufactured from 100 % polyglycolic acid and coated with polycaprolactone and calcium stearate, it offers smooth passage through tissue, excellent initial tensile strength and firm, secure knot tying.
This presentation incorporates a ½ circle needle with cylindrical body and conical Taperpoint tip of 13 mm (HR 13), USP gauge 4/0 (metric 1.5) and violet fluted thread of 70 cm, in a professional box format of 12 units, designed for stock management in operating theatres, clinics and purchasing centres, with special B2B conditions for hospitals and volume buyers.
2. Technical explanation of the material and procedure
Polyglycolic acid (PGA) is a synthetic absorbable polymer presented in a braided multifilament structure, which gives it excellent handling, low memory and smooth passage through tissue. The polycaprolactone and calcium stearate coating reduces friction and facilitates precise knot placement, even in fine gauges such as USP 4/0.
Absorption occurs by hydrolysis, a uniform process independent of the patient’s enzymatic activity, with complete absorption between 60 and 90 days: during the first weeks of healing the thread retains the necessary tensile strength (75 % at 2 weeks, 50 % at 3 and 40 % at 4), just enough time for the tissue to recover its own resistance.
The ½ circle needle with cylindrical body and conical tip (Taperpoint geometry) of 13 mm separates the fibres without cutting effect: it is the most atraumatic configuration for delicate and friable tissues, with controlled passage and minimal injury along the penetration path. The violet fluted colour ensures optimal visibility on the surgical field. The product is sterilised by ethylene oxide (EO), is single-use and cannot be re-sterilised.
3. Technical Specifications
3.1. Technical data sheet for reference AMP118HR1340
| Characteristic | Specification |
|---|---|
| Reference (SKU) | AMP118HR1340 |
| Factory reference | P118HR1340 |
| Quality | Prime |
| Suture type | Absorbable synthetic multifilament |
| Composition | 100 % polyglycolic acid (PGA) |
| Coating | Polycaprolactone and calcium stearate |
| Thread structure | Braided multifilament, grooved |
| USP / metric gauge | 4/0 (metric 1.5) |
| Approximate diameter | 0.15 – 0.199 mm (USP/EP standard) |
| Thread colour | Violet |
| Thread length | 70 cm |
| Needle shape | ½ circle |
| Needle body | Cylindrical |
| Needle tip | Conical Taperpoint, atraumatic |
| Needle size | 13 mm (HR 13) |
| Absorption mechanism | Hydrolysis |
| Total absorption | 60–90 days |
| Sterilisation | Ethylene oxide (EO), sterile, single-use, not re-sterilisable |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures and general surgery |
| Presentation | Box of 12 units |
| Minimum order | 30 boxes |
| Regulatory standard | USP/EP |
3.2. Tensile strength retention and absorption profile
| Period after implantation | Retained tensile strength |
|---|---|
| 2 weeks | 75 % of initial tensile strength |
| 3 weeks | 50 % of initial tensile strength |
| 4 weeks | 40 % of initial tensile strength |
| Total absorption | 60–90 days, by hydrolysis |
Predictable and controlled absorption profile, with minimal tissue reaction, excellent tensile strength and optimal knot tying. Price per box of 12 units.
4. Equivalences and Cross-References
Coated braided polyglycolic acid is the most well-established absorbable multifilament material of medium absorption in surgical practice. This Prime reference, with hydrolysis absorption of 60–90 days and tensile strength retention of 75/50/40 % at 2/3/4 weeks, is comparable at category level (not interchangeable 1:1) with the main homologous PGA lines from other manufacturers. The verified correspondence against official sources (manufacturer catalogue and GUDID database) is detailed below:
| Manufacturer / Brand | Family / Model | Reference Code / SKU | Equivalence Level | Technical Differences / Remarks |
|---|---|---|---|---|
| B. Braun | Safil® Violet (coated braided polyglycolic acid) | G1048008 / X1048008 – Safil Violet 4/0 (1.5) 70 cm HR13 (verified in GUDID, DI 04038653944002 / 04038653965656) | Comparable by category (not interchangeable 1:1) | Same 100 % PGA, same HR13 needle geometry (½ circle, cylindrical body, conical tip) and same measurements (USP 4/0, 70 cm thread, violet); coating and force retention profile of the B. Braun brand versus polycaprolactone/calcium stearate of this reference. |
| SMI (Belgium) | Surgicryl® PGA (braided multifilament coated with polycaprolactone and calcium stearate, violet) | 11150113 – USP 4/0, HR 13 mm needle, 75 cm violet (verified in official SMI data sheet) | Comparable by category (not interchangeable 1:1) | Same PGA material and same HR13 ½ circle needle with conical tip in gauge 4/0; the official SMI format is supplied with 75 cm thread (not 70 cm), so substitution must be validated code by code. |
Other homologous PGA lines verified at category level in manufacturer sources, without an exact 4/0 + HR13 + 70 cm combination confirmed in the public catalogue: DemeTECH – DemeSORB® PGA (USA), Assut Sutures – AssuCryl® PGA (Switzerland), Medtronic – Dexon® / Dexon II® (historical range) and Laboratorio Aragó – PGA range. In all cases it is essential to confirm the specific code with each commercial company.
There is no equivalence table validated by the manufacturer; consult the technical department before use in tenders. The indicated correspondences are functional comparisons verified against official sources (manufacturer catalogue and GUDID) and must be confirmed code by code — gauge, needle geometry, length and presentation — before a formal substitution of references.
5. Selection tips and variants of interest
USP gauge 4/0 (metric 1.5) is indicated for fine tissues with moderate tension: mucosa, subcutaneous tissue, delicate planes and precision approximations, with wide application in general surgery and minor ambulatory surgery techniques. The ½ circle needle with conical Taperpoint tip of 13 mm is the most atraumatic option for soft and friable tissues; if a greater penetration arc is required, 17 mm needles from the same family can be considered.
Variants and alternatives from the same Cahnos catalogue:
- AM118HR1340: same HR 13 mm configuration, USP 4/0, violet, 70 cm in standard quality (factory reference 118HR1340), for protocols with equivalent requirements and cost optimisation.
- AMP118HR1330 / AMP118HR1320 / AMP118HR1310: same HR 13 mm needle in USP gauges 3/0, 2/0 and 0 respectively, same 60–90 day absorption (larger or smaller diameter depending on required tension).
- P1L18HR1340: alternative in polyglactin 910 (PGLA) with the same HR 13 mm needle and USP 4/0 gauge, absorption of 56–72 days.
- AMP418HR1340: alternative in polydioxanone (PDO) monofilament with the same HR 13 mm needle and USP 4/0 gauge, for longer healing times (180–210 days).
All references are supplied in boxes of 12 units; the minimum order is 30 boxes to ensure continuous supply to your healthcare facility. As this is sterile single-use material, check the integrity of the packaging before each procedure, respect the expiry date and store in a cool, dry place, protected from sunlight. For special B2B conditions by volume or equivalents from other commercial houses, contact our technical-commercial team.





Reviews