Description
1. Commercial Introduction
The PGA Prime Multifilament Absorbable Suture is the reference solution for approximation of soft tissues and ligatures in general surgery and precision specialties. It features a ½ circle needle with cylindrical body and 6.5 mm taperpoint conical point on a braided thread of 100 % absorbable polyglycolic acid (PGA), in violet and 30 cm length. Professional presentation in a box of 12 sterile units with batch traceability, ideal for supplier qualification, public procurement tenders and hospital supply agreements.
2. Technical Explanation
The thread is made of polyglycolic acid (PGA), a synthetic polyester that degrades by hydrolysis with fully predictable kinetics independent of the patient’s enzymatic activity: mass absorption is completed between 60 and 90 days, while tensile strength is maintained during the critical healing phase (75 % at 2 weeks, 50 % at 3 weeks and 40 % at 4 weeks). The polycaprolactone and calcium stearate coating reduces friction and the sawing effect on tissue, improves thread glide and facilitates knot seating with excellent knot security and minimal tissue reaction. The taperpoint conical point needle separates fibres without cutting them, minimising trauma and bleeding at the passage point; USP 7/0 (metric EP 0.5) is the usual fineness for delicate planes and the violet colour ensures immediate visibility on the surgical field. Product sterile by ethylene oxide (EO), single-use and non-resterilisable.
3. Technical Specifications
3.1. General characteristics
| Parameter | Specification |
| Brand / quality | PREMISUT PRIME (premium range of multifilament suture) |
| Thread material | Polyglycolic acid (PGA) 100 % synthetic |
| Structure | Braided multifilament, non-grooved surface (Prime) |
| Coating | Polycaprolactone and calcium stearate |
| Absorption type | By hydrolysis; complete mass absorption in 60-90 days |
| Gauge / size | USP 7/0 (metric EP 0.5) |
| Thread diameter | 0.050 – 0.069 mm (USP/EP Pharmacopoeia) |
| Needle shape | ½ circle (1/2 circular) |
| Tip type | Cylindrical body, conical tip (taperpoint) |
| Needle size | 6.5 mm (HR 6.5 Prime profile) |
| Swaged | Single needle (1 thread with needle) |
| Thread length | 30 cm |
| Thread colour | Violet |
| Sterilisation | Ethylene oxide (EO); single-use, not re-sterilisable |
| Indications | Soft tissue approximation, ligatures, general surgery and precision microsurgery |
| Commercial presentation | Box of 12 units (sterile individual pouches); minimum order 10 boxes |
| Regulatory standard | Conforms to USP/EP Pharmacopoeia for sterile absorbable sutures |
| Factory reference | P114HR6,570 |
| SKU Cahnos | AMP114HR6,570 |
3.2. Tensile strength and absorption profile
| Period | Remaining tensile strength | Absorption |
| 2 weeks | Approximately 75 % | — |
| 3 weeks | Approximately 50 % | — |
| 4 weeks | Approximately 40 % | — |
| 60-90 days | — | Complete absorption by hydrolysis |
3.3. Gauge equivalence according to USP/EP Pharmacopoeia
| USP size | Metric measurement (EP) | Approximate thread diameter |
| 8/0 | 0.4 | 0.040 – 0.049 mm |
| 7/0 (this product) | 0.5 | 0.050 – 0.069 mm |
| 6/0 | 0.7 | 0.070 – 0.099 mm |
| 5/0 | 1.0 | 0.100 – 0.149 mm |
4. Equivalences and Cross-References
The reference P114HR6,570 corresponds to the official nomenclature of the PREMISUT PRIME range (HR 6.5 profile: ½ circle needle with cylindrical body and 6.5 mm taperpoint conical point, USP 7/0 gauge, 30 cm violet thread, single needle configuration) and Cahnos markets it without altering characters: no digit may be added, removed or reordered when ordering or tendering. There is no manufacturer-validated equivalence table cross-referencing this specific configuration (100 % PGA, USP 7/0, 6.5 mm taperpoint conical needle in ½ circle, single needle, 30 cm) with references from other brands; the following row is offered as a functional comparison verified against the manufacturer’s official catalogue and GUDID/FDA database, and must be validated against the IFU before any substitution. As it is another brand, it does not constitute an exact reference equivalence (reserved for the same manufacturer).
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| B. Braun (Aesculap) | SAFIL® – Coated braided PGA, violet, USP 7/0 (0.5) gauge, DR10 needle with round body and conical point, 45 cm thread | C1048101 (verified in GUDID/FDA and official catalogue B. Braun) | Comparable by category (not interchangeable 1:1) | They match in base material (pure polyglycolic acid), USP 7/0 (EP 0.5) gauge, violet colour, conical point needle geometry (DR profile) and single needle configuration. They differ in needle size (10 mm vs 6.5 mm) and thread length (45 cm vs 30 cm): not interchangeable without protocol validation. |
There is no manufacturer-validated equivalence table cross-referencing this specific configuration (100 % PGA, 6.5 mm taperpoint conical needle in ½ circle, USP 7/0 gauge, 30 cm thread, Prime quality, single needle) with references from other brands; consult the technical department before use in tenders. Other homologous families of medium-duration braided polyglycolic acid — e.g. Dexon II (PGA), VICRYL® (polyglactin 910, a copolymer different from PGA homopolymer and with its own absorption kinetics) or the PGLA range such as Novosyn / Polysorb — must be reviewed property by property (material, coating, USP/metric gauge, curvature, point family, needle size and number, thread length, colour and CE/MDR regulatory classification) before any substitution. This needle geometry (HR 6.5, USP 7/0, 30 cm, single needle) is only fully safely interchangeable within the PREMISUT system itself.
5. Selection tips and variants of interest
Tissue to be sutured: USP 7/0 (metric EP 0.5, diameter 0.050-0.069 mm) is the precision choice for delicate planes and low tension; the 6.5 mm taperpoint conical point needle in ½ circle dissects the plane by separating fibres without cutting them, ideal for lax, well-irrigated tissues and fine vascular structures, as well as for paediatric surgery and mucosal approximation. If your protocol requires continuous suture without rethreading with the same polymer and ½ curvature, the spatulated PGA Prime range offers double needle formats (e.g. 2×HSP) in 7/0 and 8/0 gauges; for maximum fineness, consult the 8/0 gauges of the same conical family and in the spatulated DSP 6.5 profile (single needle of 30 cm, SKU AMP114DSP6,570, and double needle, SKU AMP111DSP6,570-2). If a polyglactin 910 absorption profile is required (56-72 days instead of 60-90 days), the PREMISUT range offers the same needle assembly in PGLA (SKU P1L14DSP6,570): remember that the polymer composition changes, so it is not interchangeable for indication purposes without clinical validation. Sterility verification: product sterile by ethylene oxide and single-use; check the blister integrity and expiry date before each use, keep the reference P114HR6,570 together with the batch number for traceability in the operating theatre and store below 30 °C, protected from moisture and direct sunlight. For special B2B conditions by volume, hospital supply agreements or validation of equivalents from other commercial brands, contact the technical-commercial team of Cahnos.





Reviews