Description
1. Commercial Introduction
At Cahnos we offer you the absorbable multifilament polyglycolic acid (PGA) suture in Prime quality from the Premisut range, reference P118HR4001 (SKU AMP118HR4001), a high-performance professional solution for the B2B supply of operating theatres, hospitals, private clinics and purchasing centres. This presentation combines a ½ circle needle with a cylindrical body and conical point of 40 mm (HR 40), USP 1 gauge (metric 4) and a fluted violet thread of 70 cm, with excellent value for money and the warranty of professional distribution from Cahnos. It is the ideal choice for the approximation of tissues with moderate-to-high tension, ligatures and general surgery, with predictable absorption that avoids the need for suture removal. It is supplied sterile, ready for immediate use in the operating theatre, in a box of 12 units.
2. Technical explanation of the product and procedure
Polyglycolic acid (PGA) is a 100 % absorbable synthetic polymer, structured as a braided multifilament with a fluted finish and coated with polycaprolactone and calcium stearate. This combination ensures smooth passage through tissue, low suture memory, reduced sawing effect and firm, secure knot seating, even in larger diameter gauges such as USP 1.
Absorption occurs by hydrolysis, i.e., by degradation of the polymer in contact with organic fluids, over a predictable interval of 60 to 90 days, independent of the patient’s enzymatic activity. The suture retains 75 % of its tensile strength at 2 weeks, 50 % at 3 and 40 % at 4, an ideal profile for tissues requiring mechanical support during 2 to 4 weeks of healing.
The ½ circle needle with cylindrical body and conical point (Taperpoint geometry) of 40 mm passes through tissue separating the fibres without a cutting effect: it is the most atraumatic configuration for soft and friable tissues, ideal for deep suturing, muscle planes, fascia and ligatures of medium-calibre vessels. The violet colour ensures optimal visibility of the suture in the surgical field. The product is sterilised by ethylene oxide (EO), is single-use and cannot be re-sterilised.
3. Technical Specifications
3.1. Technical data sheet for reference P118HR4001
| Characteristic | Specification |
| Reference (SKU Cahnos) | AMP118HR4001 |
| Tariff reference / manufacturer | P118HR4001 |
| Range / Quality | Premisut – Prime |
| Material | 100 % synthetic absorbable polyglycolic acid (PGA) |
| Thread structure | Braided multifilament with grooved finish |
| Coating | Polycaprolactone and calcium stearate |
| Gauge | USP 1 (metric 4) |
| Approximate diameter | 0.40 – 0.499 mm (indicative according to Pharmacopoeia) |
| Thread colour | Violet (also available colourless in the family) |
| Thread length | 70 cm, 1 needle per suture |
| Needle shape | ½ circle (1/2 circular) |
| Needle body | Cylindrical |
| Needle tip | Conical Taperpoint (tapered, atraumatic) |
| Needle size | HR 40 Prime (40 mm) |
| Absorption mechanism | By hydrolysis, between 60 and 90 days |
| Sterilisation | Ethylene oxide (single-use, not re-sterilisable) |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures and general surgery |
| Presentation | Box of 12 sterile units |
| Regulatory standard | USP / EP |
3.2. Tensile strength retention profile after implantation
| Time point | Tensile strength retention |
| 2 weeks | 75% of initial strength |
| 3 weeks | 50% of initial strength |
| 4 weeks | 40% of initial strength |
| Complete absorption | 60 – 90 days (hydrolysis) |
3.3. Gauge equivalence according to USP/EP Pharmacopoeia
| USP size | Metric size | Approximate diameter |
| 1 (this product) | 4.0 | 0.40 – 0.499 mm (indicative according to Pharmacopoeia) |
3.4. Presentation and ordering conditions
| Value | Value |
| Presentation | Box of 12 units (individual sterile unit) |
| Minimum order | 30 units |
| Format | Sterile, single-use |
4. Equivalences and Cross-References
Braided and coated multifilament polyglycolic acid (PGA) is the most established medium-absorption synthetic absorbable on the market (classic ‘Dexon’ type family). This Prime reference, with hydrolysis absorption of 60-90 days, tensile strength retention of 75/50/40 % at 2/3/4 weeks and a ½ circle Taperpoint conical point needle of 40 mm, is comparable at category level (not interchangeable 1:1) with the following homologous lines verified against official sources or authorised distribution:
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Ethicon (Johnson & Johnson MedTech) | Coated VICRYL® – braided coated polyglactin 910, violet, USP 1, 90 cm, CT-2 taper point needle of 40 mm, ½ circle | J331H | Comparable by category (not interchangeable 1:1) | Confirmed in authorised distribution: USP 1 gauge, violet colour, braided structure, ½ circle curvature, atraumatic conical point and needle size 40 mm coincide; the polymer (polyglactin 910 vs 100 % PGA), suture length (90 cm vs 70 cm), coating and absorption interval (56-72 days vs 60-90 days) differ. |
| SMI (Belgium) | Surgicryl® PGA – cylindrical body, Taperpoint conical point, USP 1, ½ circle | SM1140 0140 (HR 40 mm) | Comparable by category (not interchangeable 1:1) | Confirmed in authorised distribution: same 100 % PGA braided coated with polycaprolactone and calcium stearate, same USP 1 gauge, ½ circle curvature, conical point and needle size 40 mm (HR 40); verify the strand length and colour in its official data sheet before any substitution. |
There is no equivalence table validated by the manufacturer; consult the technical department before use in tenders. As homologous families of 100 % braided coated PGA to be verified reference by reference in their official data sheet before substitution, the following lines are cited: Safil® (B. Braun), Dexon II (Medtronic/Covidien), Surgicryl® PGA (SMI), DemeSORB® PGA (DemeTECH) and Atramat (Laboratorio Aragó), always checking USP 1 gauge, ½ circle curvature, 40 mm conical point, 70 cm suture length and violet colour. The copolymeric polyglactin 910 ranges (VICRYL®, Polysorb®, Novosyn®) are chemically different from PGA homopolymer and should not be treated as interchangeable 1:1 without validation from the pharmacy service or technical department. The cited brands belong to their respective manufacturers and are mentioned only as equivalence reference.
5. Selection tips and variants of interest
USP 1 gauge (metric 4) is indicated for tissues with moderate-to-high tension: fascia, aponeurosis, abdominal wall, deep subcutaneous tissue, muscle planes and ligatures of medium-calibre vessels. The ½ circle needle with Taperpoint conical point separates the fibres without cutting them, ideal for soft tissues; if the tissue is very delicate, a blunt point (HRN) is recommended, while for fibrous structures a trocar point is considered. Select the gauge according to tissue tension: smaller gauges (2/0, 3/0) for fine tissues and larger (2) for planes of higher tension.
| Reference | Variant |
| AMP118HR4001 | PGA Prime USP 1, HR 40 mm needle, violet, 70 cm (current product) |
| AMP112HR4001H | PGA Prime Heavy USP 1, HR 40 heavy needle, violet, 75 cm (reinforced needle) |
| AMP118HR4002 | PGA Prime USP 2, same HR 40 mm needle, violet, 70 cm (higher tension) |
| AMP118HR4010 | PGA Prime USP 0, same HR 40 mm needle, violet, 70 cm (smaller gauge) |
| AMP118HR4030 | PGA Prime USP 3/0, same HR 40 mm needle, violet, 70 cm (fine gauge) |
| AMP113HR4010 | PGA Prime USP 0, HR 40 mm needle, violet, 90 cm (longer suture length) |
| P1L18HR4001 | PGLA Prime USP 1, HR 40 mm needle, violet, 70 cm (absorption 56-72 days) |
| P1L13HR4001 | PGLA Prime USP 1, HR 40 mm needle, violet, 90 cm (absorption 56-72 days) |
| AMP412HR4010 | PDO Prime USP 0, HR 40 mm needle, violet, 75 cm (absorption 180-210 days) |
All references are supplied in boxes of 12 units; the minimum order is 30 units. If you need longer strength retention (180-210 days), consider the polydioxanone (PDO) family. As this is a sterile single-use product, check the integrity of the pouch and the expiry date before each procedure, and keep the reference together with the batch number for traceability in the operating theatre. For B2B volume conditions, hospital supply contracts or equivalents from other commercial houses, contact the technical-commercial team at Cahnos.