Description
1. Commercial Introduction
The absorbable multifilament PGA suture reference 113HS3602 is the surgical closure solution your centre needs for general surgery procedures that require controlled reabsorption, excellent handling and irreproachable knot security. It combines a ½ circle needle with a reverse cutting triangular point of 36 mm with a braided thread of polyglactic acid (PGA) in USP 2 gauge, fluted violet and 90 cm in length.
Presented in B2B format in a box of 12 units with a minimum order of 30 boxes, it offers an outstanding price-quality ratio for operating theatres, clinics, distributors and purchasing centres for surgical material.
2. Technical Explanation
Polyglactic acid (PGA) is a 100 % absorbable synthetic polymer that is presented in a braided multifilament structure coated with polycaprolactone and calcium stearate. This coating reduces friction when passing through tissues, improves thread sliding and facilitates firm and secure knotting, with minimal tissue reaction.
Absorption occurs by hydrolysis, a predictable process independent of the patient’s enzymatic activity, completed between 60 and 90 days. Tensile strength retention maintains the necessary support during the first 2 to 4 weeks of healing (75 % at 2 weeks, 50 % at 3 and 40 % at 4), while the tissue recovers its own resistance.
The HS36 needle has a ½ circle curvature with a triangular body and reverse cutting point: the cutting edge is oriented towards the inside of the arc, which facilitates penetration into firm and poorly distensible tissues (aponeurosis, fascia, skin, abdominal wall) without enlarging the puncture bed. Its 36 mm length makes it suitable for planes of some depth. Product sterile by ethylene oxide, single-use and non-resterilisable.
3. Technical Specifications
3.1. Technical data sheet for reference 113HS3602
| Characteristic | Specification |
| Factory reference | 113HS3602 |
| Reference Cahnos (SKU) | AM113HS3602 |
| Thread material | 100 % polyglactic acid (PGA), absorbable by hydrolysis |
| Thread structure | Braided multifilament with grooved finish |
| Coating | Polycaprolactone and calcium stearate |
| Gauge / size | USP 2 (metric 5), conforming to USP/EP Pharmacopoeia |
| Thread colour | Fluted violet (colourless version available in the family) |
| Thread length | 90 cm |
| Needle shape | ½ circle (1/2 circular) |
| Needle tip | Triangular reverse cutting (reverse cutting) |
| Needle size | HS 36 mm, 1 needle per suture |
| Absorption mechanism | Hydrolysis, completed between 60 and 90 days |
| Sterilisation | Ethylene oxide (EtO); sterile, single-use, not re-sterilisable |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures and general surgery |
| Commercial presentation | Box of 12 units; minimum order of 30 boxes |
3.2. Tensile strength retention profile
| Period after implantation | Retained tensile strength |
| 2 weeks | 75 % of initial tensile strength |
| 3 weeks | 50 % of initial tensile strength |
| 4 weeks | 40 % of initial tensile strength |
| Total absorption | 60-90 days, by hydrolysis |
3.3. Reference decoding
Official reference 113HS3602: 113 (90 cm thread) + HS36 (½ circle needle with reverse cutting triangular point of 36 mm) + 02 (USP 2 gauge). Cahnos markets the official reference without altering characters: no digit may be added, removed or reordered when ordering or bidding.
4. Equivalences and Cross-References
Polyglactic acid (PGA) is one of the most well-established absorbable synthetic multifilament suture raw materials on the market. There is no equivalence table validated by the manufacturer that cross-references this reference with other commercial houses; the following matches are functional comparisons at category level and should not be treated as 1:1 interchangeable without prior technical validation.
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence Level | Technical Differences / Remarks |
| Ethicon (Johnson & Johnson MedTech) | VICRYL / VICRYL Plus (braided polyglactin 910, violet) | Not published (verify exact configuration) | Comparable by category (not interchangeable 1:1) | They match in multifilament structure, USP 2 gauge and ½ circle needle with cutting point; they differ in the polymer (polyglactin 910, absorption 56-72 days, versus PGA of 60-90 days for this reference), the braid coating and the strength retention at 4 weeks (Vicryl below 40 %). Verify absorption profile before substituting. |
| B. Braun (Aesculap) | Safil (braided polyglactin 910, violet) | Not published (verify exact configuration) | Comparable by category (not interchangeable 1:1) | Same clinical indication and absorbable braided structure; they differ in composition (PGLA versus PGA) and available needle diameter. Review gauge and needle geometry in the official data sheet. |
| Medtronic / Covidien | Polysorb (absorbable braided range) | Not published (verify exact configuration) | Comparable by category (not interchangeable 1:1) | Absorbable braided range of the segment, with similar but not identical absorption and retention profiles; validate property by property before substitution. |
| SMI / Assut | Surgicryl / Assucryl (polyglactic acid / polyglactin 910) | Not published (verify exact configuration) | Comparable by category (not interchangeable 1:1) | Homologous families of the PGA/PGLA segment; review exact material, USP 2 gauge, HS36 needle size and assembly of 1 needle in their technical data sheets. |
There is no equivalence table validated by the manufacturer; consult the technical department before using in tendering. The cited brands belong to their respective manufacturers and are mentioned only as comparative reference; no unverified cross-reference codes are published.
5. Selection tips and variants of interest
USP 2 gauge (metric 5) is indicated for tissues that withstand high tension: aponeurosis, fascia, muscular planes, abdominal wall and ligatures of medium-calibre vessels. The ½ circle needle with reverse cutting triangular point (HS36) is the appropriate choice for firm and poorly distensible tissues; if you need to pass through soft tissues with minimal injury, consider the conical points (HR/taperpoint) or trocar (HRT) from the same PGA family.
| Variant (SKU) | Description |
| AM113HS3602 (113HS3602) | PGA – HS36 reverse cutting – USP 2 violet – 90 cm (this product) |
| AMP113HS3601 (P113HS3601) | PGA Prime – HS36 reverse cutting – USP 1 violet – 90 cm (same geometry, gauge and Prime range) |
| AM112HS3620 (112HS3620) | PGA – HS36 reverse cutting – USP 2/0 violet – 75 cm (smaller gauge and less thread) |
| AMP118HS3610 (P118HS3610) | PGA Prime – HS36 reverse cutting – USP 0 violet – 70 cm (fine gauge, Prime range) |
| AMP112DS3602 (P112DS3602) | PGA Prime – DS36 reverse cutting 3/8 – USP 2 violet – 75 cm (another needle curvature) |
If you need a longer strength retention (180-210 days), consider the polydioxanone (PDO) family from Cahnos. Sterile single-use product: check the integrity of the pouch and the expiry date before each procedure and keep reference 113HS3602 together with the batch number for traceability in the operating theatre. For B2B volume conditions or equivalences from other commercial houses, contact the technical-commercial team of Cahnos.





Reviews
There are no reviews yet.