Description
1. Commercial Introduction
The absorbable multifilament polyglycolic acid (PGA) suture reference 122HS2220 is the B2B supply solution for operating theatres, general surgery departments, purchasing centres and distributors that need an absorbable thread with predictable behaviour, excellent handling and proven security in knot tying. It combines a 100% absorbable synthetic polymer with a ½ circle reverse cutting triangular point needle of 22 mm, USP 2/0 gauge and undyed grooved thread of 75 cm: firm penetration in medium-density tissues with maximum aesthetic discretion on the surgical field.
2. Technical Explanation
Polyglycolic acid (PGA) is an absorbable synthetic polymer presented in a braided multifilament structure with a grooved finish and a coating of polycaprolactone and calcium stearate. This configuration provides a regular surface that facilitates the passage of the thread through tissue, smooth handling and firm, secure knot tying, even in fine gauges such as USP 2/0. Absorption occurs by hydrolysis, i.e., by degradation of the polymer in contact with organic fluids, over a predictable period of 60 to 90 days and independent of the patient’s enzymatic activity.
The ½ circle needle with triangular body and reverse cutting point of 22 mm concentrates the cutting edge on the outer curvature: it penetrates skin, fasciae, aponeuroses and mucous membranes with a clean, controlled cut, without enlarging the puncture bed and minimising the risk of tearing. The undyed (uncoloured) thread is practically invisible on the surgical bed, ideal for deep planes and aesthetic finishes. Product sterile by ethylene oxide, single-use and non-reesterilizable.
3. Technical Specifications
3.1. Technical data sheet for reference 122HS2220
| Manufacturer reference | 122HS2220 |
| Reference Cahnos (SKU) | AM122HS2220 |
| Thread material | 100 % polyglycolic acid (PGA), braided multifilament |
| Thread structure | Braided multifilament with grooved finish |
| Coating | Polycaprolactone and calcium stearate |
| Gauge | USP 2/0 (metric 3) |
| Thread length | 75 cm |
| Thread colour | Colourless |
| Needle shape | ½ circle (1/2 circular) |
| Tip type | Reverse cutting triangular (reverse cutting) |
| Needle size | 22 mm (HS 22 profile) |
| Absorption mechanism | Hydrolysis, 60-90 days |
| Sterilisation | Ethylene oxide (single-use, not re-sterilisable) |
| Tissue reaction | Minimal |
| Presentation | Box of 12 units · Minimum order 30 boxes |
3.2. Tensile strength retention profile after implantation
| Time point | Tensile strength retention |
| 2 weeks | 75 % |
| 3 weeks | 50 % |
| 4 weeks | 40 % |
| Complete absorption | 60-90 days by hydrolysis |
Complete absorption by hydrolysis between 60 and 90 days. Minimal tissue reaction, excellent tensile strength and optimal knot tying, with a fully predictable absorption rate.
3.3. Gauge equivalence according to Pharmacopoeia (indicative)
| USP size | Metric measurement (EP) | Approximate thread diameter |
| 2/0 (this product) | 3 | 0.30 – 0.349 mm |
| 3/0 | 2 | 0.20 – 0.249 mm |
| 0 | 3.5 | 0.35 – 0.399 mm |
4. Equivalences and Cross-References
Braided and coated multifilament polyglycolic acid (PGA) is the most widespread medium-duration synthetic absorbable in the surgical market, and the ½ circle configuration with a 22 mm reverse cutting triangular point is common in the approximation of firm tissues. However, for the exact combination of this data sheet (HS 22 needle of 22 mm in ½ circle with reverse cutting, USP 2/0, undyed thread of 75 cm), no official code from another manufacturer has been found in public sources that matches field by field with sufficient certainty to publish a cross-reference equivalence table.
There is no equivalence table validated by the manufacturer; consult the technical department before use in tenders.
At category level, the homologous families to compare property by property (material, gauge, curvature, needle size and family, length, colour, coating and CE/MDR classification) are: Safil® (B. Braun/Aesculap, braided coated PGA with reverse cutting needles), Surgicryl® PGA (SMI, Belgium, identical PGA polymer with polycaprolactone and calcium stearate coating, with HS 22 / USP 2/0 / 75 cm option in violet and undyed versions —exact code of the undyed variant pending confirmation with the manufacturer or its distributor—), ASSUFIL (Assut Europe) and SERAFIT (Serag Wiessner). The polyglactin 910 lines (Vicryl® from Ethicon, Polysorb® from Medtronic, Surgicryl 910 from SMI) share indication and medium absorption profile but use a different copolymer from 100% PGA, so they are not interchangeable 1:1.
5. Selection tips and variants of interest
The USP 2/0 gauge (metric 3) offers the ideal balance between initial strength and handling for tissues with moderate tension: skin, subcutaneous tissue, mucosa, fascia and soft tissue in general, as well as ligatures. The 22 mm reverse cutting triangular point needle (HS 22) is recommended when firm tissues need to be penetrated without tearing, with a clean and precise needle passage; if the technique prefers an atraumatic conical point, consult the HR cylindrical needle family from the same PGA range.
Within the Cahnos catalogue, the following variants of interest are available:
| Reference | Description |
| AMP122HS2220 | PGA Prime USP 2/0, same HS 22 needle, undyed, 75 cm (premium range, same manufacturer) |
| AM118HS2220 | PGA USP 2/0, HS 22 needle, violet, 70 cm (maximum visibility in the field) |
| AM112HR2220 | PGA USP 2/0, HR 22 mm cylindrical needle, violet, 75 cm (taperpoint conical point) |
| 1L22HS2220 | PGLA (polyglactin 910) USP 2/0, HS 22 needle, undyed, 75 cm (absorption 56-72 days) |
| AM122DS2420 | PGA USP 2/0, ⅜ circle triangular DS 24 mm needle, undyed, 75 cm (⅜ curvature) |
As this is sterile single-use material, check the integrity of the blister and the expiry date before each procedure and keep the reference together with the batch number for traceability in the operating theatre. Store in a cool, dry place, protected from sunlight. For B2B volume conditions, hospital supply agreements or validation of equivalents from other manufacturers, contact our technical-commercial team at Cahnos.