Description
1. Introduction
The absorbable multifilament polyglycolic acid (PGA) Prime suture in USP 1 gauge is the reference choice in the operating theatre for professionals seeking a synthetic absorbable thread with reliable, predictable and high-quality performance in tissue approximation, ligatures and general surgery. Its ½ circle needle with reverse cutting triangular point of 40 mm (HS 40) penetrates dense and firm tissues with controlled trauma, while the fluted violet thread of 90 cm ensures optimal visibility in the surgical field and excellent handling in every stitch. Commercial presentation in a box of 12 units, ideal for continuous supply to hospitals, clinics and purchasing centres.
2. Technical explanation
Polyglycolic acid (PGA) is a 100 % absorbable synthetic polymer presented in a braided multifilament structure with a fluted finish. Its coating of polycaprolactone and calcium stearate improves thread sliding through tissue, reduces friction and facilitates firm, secure knot tying even in medium-diameter gauges such as USP 1 (metric 4).
Absorption occurs by hydrolysis, i.e., by degradation of the polymer in contact with organic fluids, over a predictable period of 60 to 90 days, independent of the patient’s enzymatic activity. The ½ circle needle with triangular body and reverse cutting point of 40 mm is designed to penetrate firm, consistent tissues without tearing, making it especially suitable for closures of fascia, aponeurosis, muscle planes and medium-high tension ligatures. The product is supplied sterile by ethylene oxide, single-use and cannot be re-sterilised.
3. Technical Specifications
3.1. General characteristics of reference AMP113HS4001
| Reference (SKU) | AMP113HS4001 / P113HS4001 |
| Material | 100 % polyglycolic acid (PGA) absorbable by hydrolysis |
| Thread structure | Braided multifilament with grooved finish |
| Coating | Polycaprolactone and calcium stearate |
| Gauge | USP 1 (metric 4) |
| Thread colour | Violet grooved |
| Thread length | 90 cm |
| Needle shape | ½ circle |
| Tip type | Reverse cutting triangular (Reverse Cutting), body and point |
| Needle size | 40 mm (HS 40 Prime) |
| Absorption mechanism | Hydrolysis (60-90 days) |
| Sterilisation | Ethylene oxide (EO); single-use, non-reesterilizable |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures and general surgery |
| Presentation | Box of 12 units |
3.2. Tensile strength retention profile
| Time after implantation | Retained tensile strength |
| 2 weeks | 75 % |
| 3 weeks | 50 % |
| 4 weeks | 40 % |
| Complete absorption | 60-90 days (by hydrolysis) |
Minimal tissue reaction, excellent tensile strength and optimal knot tying, with a fully predictable absorption rate.
4. Equivalences and Cross-References
Polyglycolic acid (PGA) is the most widely used medium-duration braided multifilament absorbable on the market. This reference shares the absorption profile (60-90 days), coated braided structure, USP 1 gauge, HS 40 mm ½ circle reverse cutting needle and 90 cm length with the main homologous families of medium-absorption braided absorbable suture. Comparison with polymers other than PGA homopolymer is made at category level.
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence Level | Technical Differences / Remarks |
| Ethicon (Johnson & Johnson) | Coated VICRYL® Plus Antibacterial (braided polyglactin 910, violet, 90 cm, ½ circle reverse cutting needle 40 mm, USP 1) | VP2421 | Comparable by category (not interchangeable 1:1) | Copolymer composition 90/10 glycolide/L-lactide versus PGA homopolymer; declared absorption 56-70 days; same HS 40 mm reverse cutting needle, USP 1 gauge, violet and 90 cm. Code verified on the manufacturer’s official website and in authorised distributor catalogues; consult the manufacturer’s technical data sheet before substituting. |
Other commercial houses with homologous families of medium-absorption braided absorbable suture —B. Braun with its Safil line of polyglactin 910, Medtronic with the Polysorb range, SMI with Surgicryl and Spanish suture laboratories— offer products of the same clinical category, but no reference code confirmed against an official source is available for this exact configuration (USP 1 / HS 40 mm / 90 cm). There is no equivalence table validated by the manufacturer; consult the technical department before use in tenders.
5. Selection tips and variants of interest
To choose the most appropriate suture, consider the wound support time required by the tissue to be sutured and the gauge according to the anatomical area. USP 1 gauge is indicated for tissues requiring high initial strength: aponeurosis, fascia, muscle planes and moderate-high tension closures. The reverse cutting triangular point (HS) is recommended for penetrating firm, dense tissues without tearing (skin, fascia); if the tissue is soft or friable, consider the conical point variant (HR).
| Reference | Variant of interest |
| AMP113HS4001 | PGA Prime USP 1, triangular needle HS 40 mm, violet, 90 cm (current product) |
| AMP113HS4002 | PGA Prime USP 2, same HS 40 mm needle (larger gauge for tissues with more tension) |
| AMP113HS4010 | PGA Prime USP 0, same HS 40 mm needle (tissues with less tension) |
| AM113HS4001 | PGA USP 1, same HS 40 mm needle, violet, 90 cm (standard range) |
| AMP118HS4001 | PGA Prime USP 1, HS 40 mm needle, violet, 70 cm (shorter thread) |
| AMP113HS4801 | PGA Prime USP 1, triangular needle HS 48 mm (needle with larger radius) |
| P1L13HS4001 | PGLA Prime USP 1, triangular needle HS 40 mm, violet, 90 cm (alternative polymer) |
All references are supplied in a box of 12 units. As this is sterile single-use material, check the integrity of the packaging before each procedure and store it in a cool, dry place, protected from sunlight. For special B2B conditions and volume orders, contact our technical-commercial team.





Reviews
There are no reviews yet.