Description
1. Introduction
The absorbable multifilament PGA Prime suture from Cahnos is the reference choice in the operating theatre for the professional seeking a reliable, predictable, high-quality absorbable synthetic thread for the approximation of soft tissues, ligatures and general surgery. Its ½ circular needle with reverse cutting triangular point of 48 mm (HS48) guarantees firm and precise penetration even in dense tissues, while the 90 cm violet fluted thread offers optimal visibility of the surgical field and excellent handling in every stitch.
Manufactured under strict quality standards for the B2B channel, it combines minimal tissue reaction, excellent tensile strength and optimal knot security. Commercial presentation in a box of 12 units, ideal for the continuous supply of hospitals, clinics and purchasing centres.
2. Technical Explanation
Polyglycolic acid (PGA) is a synthetic absorbable polymer presented in a braided multifilament structure with a fluted finish. This configuration provides a regular surface that favours the sliding of the thread through the tissue, smooth handling and firm, secure knot tying even in medium diameter gauges such as USP 1 (metric 4). The coating of polycaprolactone and calcium stearate improves sliding and reduces friction.
Absorption occurs by hydrolysis, i.e., by degradation of the polymer in contact with organic fluids, over a predictable period of 60 to 90 days. This profile makes it suitable for tissues that require wound support during the first weeks of healing, maintaining the necessary tensile strength while the tissue recovers its own resistance. The product is supplied sterile by ethylene oxide, single-use and without the possibility of re-sterilisation.
3. Technical Specifications
3.1. General characteristics of reference AMP113HS4801
| Reference (SKU) | AMP113HS4801 / P113HS4801 |
| Material | 100 % Absorbable polyglycolic acid (PGA) |
| Thread structure | Braided multifilament with grooved finish |
| Coating | Polycaprolactone and calcium stearate |
| Gauge | USP 1 (metric 4) |
| Thread colour | Violet grooved |
| Thread length | 90 cm |
| Needle shape | ½ circle |
| Needle tip | Triangular reverse cutting (Reverse Cutting) |
| Needle size | 48 mm (HS 48 Prime) |
| Absorption mechanism | Hydrolysis (60-90 days) |
| Sterilisation | Ethylene oxide (EO), single-use / sterile |
| Tissue reaction | Minimal |
| Presentation | Box of 12 units |
3.2. Tensile strength retention profile
| Period | Remaining tensile strength |
| 2 weeks | 75% of initial tensile strength |
| 3 weeks | 50% of initial tensile strength |
| 4 weeks | 40% of initial tensile strength |
| Total absorption | 60 – 90 days, by hydrolysis |
Minimal tissue reaction, excellent tensile strength and optimal knot tying, with a predictable absorption rate. Indicated for tissue approximation, ligatures and general surgery.
4. Equivalences and Cross-References
This multifilament polyglycolic acid (PGA) suture belongs to the family of braided synthetic absorbable sutures of medium absorption and is clinically positioned alongside the reference polyglactin 910 sutures of the international market (Vicryl, Safil, Polysorb, Surgicryl). All share a braided structure, violet colour, ethylene oxide sterilisation and absorption by hydrolysis in the range of 56-90 days; the main difference is the composition (PGA homopolymer versus 90/10 glycolide-lactide copolymer), so the comparison is at category level and not interchangeable 1:1.
| Manufacturer / Brand | Family / Model | Reference Code / SKU | Equivalence Level | Technical Differences / Remarks |
| Ethicon (Johnson & Johnson) | VICRYL Plus (braided polyglactin 910, violet, 90 cm, ½ circle reverse cutting needle 48 mm, USP 1) | VCP1058H | Comparable by category (not interchangeable 1:1) | Composition 90/10 glycolide/L-lactide copolymer versus PGA homopolymer; declared absorption 56-70 days; same HS 48 mm reverse cutting needle, USP 1, violet and 90 cm. Code verified in catalogues of authorised distributors; consult the manufacturer’s technical data sheet before substituting. |
Other commercial houses with homologous families of braided absorbable suture of medium absorption —B. Braun with its Safil line of polyglactin 910, Medtronic with the Polysorb range, SMI with Surgicryl and Spanish suture laboratories— offer products of the same clinical category, but there is no equivalence table validated by the manufacturer for this specific reference in USP 1 / HS 48 mm / 90 cm configuration.
There is no equivalence table validated by the manufacturer; consult the technical department before use in tenders.
5. Selection tips and variants of interest
To choose the most suitable suture, consider the wound support time required by the tissue to be sutured and the gauge according to the anatomical area. USP 1 gauge is indicated for tissues that require high initial resistance: aponeurosis, fascia, muscle planes and closures of moderate-high tension.
- AMP113HS4801: PGA Prime USP 1, triangular needle HS 48 mm, violet, 90 cm (current reference).
- AMP113HS4810: PGA Prime USP 0, same HS 48 mm needle, violet, 90 cm (finer gauge for delicate tissues).
- AMP113HS4802: PGA Prime USP 2, same HS 48 mm needle, violet, 90 cm (greater resistance for deep planes).
- AMP113HS4001: PGA Prime USP 1, HS 40 mm needle, violet, 90 cm (shorter needle).
- AMP113HR4801: PGA Prime USP 1, cylindrical HR 48 mm needle, violet, 90 cm (atraumatic conical point for soft tissues).
- P1L13HS4801: alternative in polyglactin 910 (PGLA) with the same triangular HS 48 mm needle, USP 1, violet, 90 cm.
- AM113HS4801: same configuration in standard quality (not Prime) for general use.
All references are supplied in boxes of 12 units. As this is a sterile single-use product, check the integrity of the packaging before each procedure and comply with the traceability required by the centre. For special B2B conditions and volume orders, contact our technical-commercial team.