Description
1. Commercial Introduction
The absorbable monofilament Polydioxanone suture PDO/PDS, reference P418JRT3101 from the Prime range (PREMISUT), is the trusted choice for the professional who needs a synthetic absorbable thread with predictable behaviour, excellent tissue handling and prolonged tensile support. Its J-Hook needle, with cylindrical body and 31 mm tapercut trocar point (JRT31), guarantees clean and controlled penetration even in medium-consistency tissues, with minimal tissue damage.
Available in B2B presentation in boxes of 12 units with a minimum order of 30 boxes, this reference offers an excellent value for money for hospitals, operating theatres, clinics, purchasing centres and distributors of surgical material, with sterile, traceable and ready-to-use supply.
2. Technical Explanation
Polydioxanone is a synthetic polyester (poly(p-dioxanone)) presented as a monofilament, which reduces tissue drag, minimises inflammatory reaction and facilitates thread sliding and knotting. Its absorption occurs by hydrolysis, a uniform and predictable process independent of the patient’s enzymatic activity, with total mass loss between 180 and 210 days.
This profile maintains tensile strength during the critical phases of healing: 75% at 2 weeks, 60-70% at 4 weeks and 45% at 8 weeks. The violet thread, with a grooved finish, offers optimal visibility over the surgical field. The 31 mm JRT needle combines a hook-type curvature (J-Hook), ideal for reduced-access fields, with a tapercut trocar point: it penetrates firm tissues by separating the fibres without cutting effect, minimising trauma to delicate structures. It is supplied sterile by ethylene oxide, for single use and without re-sterilisation.
3. Technical Specifications
3.1. General characteristics
| Parameter | Specification |
| Reference (SKU) | AMP418JRT3101 |
| Factory reference | P418JRT3101 |
| Material | Polydioxanone (PDO/PDS), polyester poly(p-dioxanone) |
| Thread structure | Synthetic monofilament, fluted finish |
| Gauge | USP 1 (metric 4) |
| Thread length | 70 cm |
| Thread colour | Violet |
| Needle shape | J-Hook, J-Hook format (J-shaped) |
| Needle body | Cylindrical |
| Needle tip | Trocar (Tapercut) |
| Needle size | JRT 31 mm |
| Quality | Prime (range PREMISUT) |
| Sterilisation | Ethylene oxide (EtO), sterile, single-use |
| Presentation | Box of 12 units |
| Minimum order | 30 boxes |
3.2. Absorption profile and tensile strength retention
| Parameter | Value |
| Absorption type | Hydrolysis |
| Absorption time | 180 – 210 days |
| Tensile strength at 2 weeks | 75 % |
| Tensile strength at 4 weeks | 60 – 70 % |
| Tensile strength at 8 weeks | 45 % |
| Tissue reaction | Minimal |
| Knot tying | Excellent safety |
3.3. Size equivalence
| USP size | Metric size | Approximate diameter |
| 1 | 4 | 0.40 – 0.499 mm (indicative according to Pharmacopoeia) |
Indications: tissue approximation, ligatures, general surgery, paediatric cardiovascular tissues and ophthalmology. Price per box of 12 units; consult B2B conditions for orders in multiples of boxes.
4. Equivalences and Cross-References
Polydioxanone is the original polymer of PDS-type sutures, the clinical reference for absorbable monofilaments with prolonged absorption. There is no equivalence table validated by the manufacturer for this specific reference against third-party brands; the codes cited below have been verified in catalogues and public regulatory and commercial sources and correspond to the same clinical category, with the differences indicated:
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence level | Technical differences |
| Ethicon (Johnson & Johnson MedTech) | PDS® II | Z341H | Comparable by category (not interchangeable 1:1) | Same USP 1 gauge, violet thread of 70 cm and same absorption 180-210 days; CT-1 needle of ½ circle and 36 mm with conical point (taper point), does not match the JRT 31 mm tapercut trocar |
| Ethicon (Johnson & Johnson MedTech) | PDS® II Tapercut | Z371T | Comparable by category (not interchangeable 1:1) | Same USP 1 gauge, violet and trocar point (CTX); however the needle is 48 mm (½ circle) and the thread length is 91 cm (36″), they do not match 31 mm / 70 cm |
Other commercial houses with homologous polydioxanone monofilament lines are B. Braun (Monomax, PDO) and Dynacord/Monomax from the same house. In the range of long-lasting absorbable monofilaments, the closest competitor reference is the PDS range from Ethicon; before substituting, check on the candidate’s data sheet: USP gauge, thread length, needle configuration (1 needle / double needle), point type (conical, trocar/tapercut or reverse cutting), curvature (½, ⅜ or J-hook) and needle length.
There is no equivalence table validated by the manufacturer; consult the technical department before using in tenders.
5. Selection tips and variants of interest
USP 1 gauge (metric 4) offers high initial resistance, suitable for tissues subjected to moderate or high tension: tissue approximation, ligatures, fascia, muscular and aponeurotic planes, as well as general and digestive surgery. The J-Hook hook needle with 31 mm tapercut trocar point is especially recommended in reduced-access surgical fields where clean and controlled penetration with minimal tissue damage is required.
Within the Cahnos catalogue you can consider the following variants of the same needle configuration and gauge in other absorbable materials:
| Reference | Variant |
| 1L12JRT3101 / P1L12JRT3101 | PGLA (polyglactin 910), JRT 31 mm needle, USP 1, violet, 75 cm (absorption 56-72 days) |
| AMP112JRT3101 / P112JRT3101 | PGA (polyglycolic acid), JRT 31 mm needle, USP 1, violet, 75 cm (absorption 60-90 days) |
| AMP418HRT1740 / P418HRT1740 | PDO (polydioxanone), HRT 17 mm needle of ½ circle trocar point, USP 4/0, violet, 70 cm (fine gauges, delicate tissues) |
If your protocol requires a finer gauge in polydioxanone, consult the PDO Prime range in gauges 4/0, 3/0 and 2/0 with half-circle or hook needles; if you require faster absorption, consider the PGLA family (56-72 days) or PGA (60-90 days). As this is a sterile single-use product, check the integrity of the packaging before each procedure, respect the expiry date and store the suture in a cool, dry place, protected from sunlight. For special B2B conditions by volume, hospital supply contracts or equivalences from other commercial houses, contact the technical-commercial team at Cahnos.