Description
1. Commercial Introduction
The Prime line polydioxanone (PDO/PDS) absorbable monofilament suture is the reference choice when the operating theatre needs a thread with prolonged support, fully predictable behaviour and maximum knot security. This presentation combines a violet 70 cm thread in USP 4/0 gauge with double ⅜ circle needle, cylindrical body and conical point 17 mm (2xDR17), an atraumatic configuration designed for soft, vascular and slow-healing tissues.
Available in a box of 12 sterile units, it is the B2B supply solution for general surgery services, paediatric cardiovascular surgery operating theatres, ophthalmology units and hospital purchasing centres seeking quality homogeneity and complete traceability in each package.
2. Technical Explanation
Polydioxanone is a synthetic absorbable polymer —poly(p-dioxanone)— that degrades by hydrolysis, a mechanism independent of the patient’s enzymatic activity. This guarantees highly reproducible absorption kinetics between individuals and minimal tissue response, with complete absorption within 180 to 210 days, one of the longest support windows among absorbable sutures.
Being non-ribbed monofilament, the thread has no wicking effect or interstitial spaces: it reduces bacterial adherence, facilitates sliding through tissue and minimises the inflammatory reaction compared to braided threads. The ⅜ circle needle with cylindrical body and taperpoint conical point passes through soft tissues by separating fibres rather than cutting them, minimising bleeding and tissue damage; the double needle configuration (2xDR17) allows continuous suture techniques and approximation from both ends without re-threading. The violet colour improves thread visibility over the surgical field. Sterile single-use product, not re-sterilisable.
3. Technical Specifications
3.1. Technical data sheet for reference AMP418DR1740-2
| Parameter | Specification |
| SKU | AMP418DR1740-2 |
| Factory reference | P418DR1740-2 |
| Material | Polydioxanone (PDO/PDS), poly(p-dioxanone), synthetic non-fluted monofilament |
| Thread structure | Absorbable monofilament, non-braided |
| Quality | Prime |
| USP / metric gauge | USP 4/0 (metric 2) |
| Thread length | 70 cm |
| Thread colour | Violet (high visibility) |
| Needle shape | ⅜ circle |
| Needle body | Cylindrical |
| Tip type | Conical / Taperpoint |
| Needle size | DR 17 (17 mm), double needle (2x) |
| Swaged | Double needle, one needle at each end of the thread |
| Absorption mechanism | Hydrolysis (180-210 days) |
| Sterilisation | Ethylene oxide (EO); single-use, non-reesterilizable |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures, general surgery, paediatric cardiovascular tissues and ophthalmology |
| Presentation | Box of 12 sterile units |
3.2. Tensile strength retention profile
| Period after implantation | Retained tensile strength |
| 2 weeks | 75% of initial strength |
| 4 weeks | 60-70% of initial strength |
| 8 weeks | 45% of initial strength |
| Complete absorption | 180-210 days by hydrolysis |
This prolonged force retention profile maintains the necessary support during the critical healing phase of slow-healing tissues, which is why it is recommended for closures of planes under sustained tension and in paediatric cardiovascular surgery where tissue growth is expected.
4. Equivalences and Cross-References
Polydioxanone monofilament is the material basis of the market’s reference lines for long-lasting absorbable sutures (PDS-type family). Based on the verified configuration of this data sheet (PDO material, USP 4/0 gauge, violet 70 cm thread, 17 mm conical point needle), the following homologous references, comparable at category level and not interchangeable 1:1, have been cross-checked in the manufacturer’s official catalogue and authorised distributors:
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Ethicon (Johnson & Johnson MedTech) | PDS II (polydioxanone monofilament), BB needle 17 mm in ⅜ circle, conical point, USP 4/0, violet, 70 cm | W9077H | Comparable by category (not interchangeable 1:1) | Same polymer, USP 4/0 gauge, violet colour, 70 cm length and 17 mm taperpoint needle in ⅜ circle; the configuration differs: a single needle (1xBB) versus double needle. Code confirmed in Ethicon EMEA official catalogue and authorised distributors. |
| Ethicon (Johnson & Johnson MedTech) | PDS Plus Antibacterial (polydioxanone with triclosan), 17 mm taperpoint needle in ⅜ circle, USP 4/0, violet, 70 cm | PDP9077H | Comparable by category (not interchangeable 1:1) | Same polymer, gauge, colour, length and point; the configuration differs (single needle versus double) and it incorporates an antibacterial triclosan coating that this reference lacks. Code cross-checked with authorised distributors. |
| Ethicon (Johnson & Johnson MedTech) | PDS II (polydioxanone monofilament), RB-1 needle 17 mm in ½ circle, conical point, USP 4/0, violet, 70 cm | Z304H | Comparable by category (not interchangeable 1:1) | Same material, gauge, colour, length and 17 mm conical point; the curvature differs (½ circle versus ⅜) and the configuration (single needle versus double). Code verified in Ethicon catalogue and GUDID regulatory databases. |
| Ethicon (Johnson & Johnson MedTech) | PDS II (polydioxanone monofilament) D-special, double RB-1 needle 17 mm in ½ circle, conical point, USP 4/0, violet, 90 cm | D6261 | Comparable by category (not interchangeable 1:1) | Material, gauge, colour, 17 mm conical point and double needle configuration match; the curvature differs (½ circle versus ⅜) and the thread length (90 cm versus 70 cm). Code confirmed in Ethicon official catalogue. |
There is no equivalence table validated by the manufacturer declaring this reference as interchangeable 1:1 with other brands; consult the technical department before use in tenders. As homologous families in the sector, the PDS II and PDS Plus lines from Ethicon are cited, as well as the polydioxanone monofilament ranges from other manufacturers (e.g. European PDS/MonoPlus-type lines); in all cases, the USP gauge, curvature, body, point, needle size, configuration and thread length must be verified code by code in the official technical data sheet and IFU before any substitution. The cited brands belong to their respective manufacturers and are mentioned only as comparative reference.
5. Selection tips and variants of interest
When choosing a PDO suture, always verify the USP gauge according to the tissue plane: fine gauges (4/0 to 6/0) for skin, subcutaneous tissue, mucosa and delicate planes —like this reference—, 3/0 and 2/0 for general soft tissue and USP 1 and above for fascia and planes under tension. Needle geometry is also decisive: the taperpoint conical point, like that of this reference, is indicated for soft, vascular and gastrointestinal tissues; if you need to pass through more fibrous tissues, consider versions with trocar point (tapercut) or reverse cutting triangular point.
Within the Cahnos catalogue you will find variants of interest from the PDO/PDS family:
- AMP418DR1740-2 – analysed product: PDO Prime, double ⅜ circle needle conical point 17 mm (2xDR17), USP 4/0, violet, 70 cm.
- AMP412DR1940 – PDO Prime, ⅜ circle needle, conical point, 19 mm (DR19), USP 4/0, violet, 75 cm (single needle, longer needle).
- AMP412HR2240-2VB – PDO Prime VisiBlack, double needle ½ circle conical point 22 mm, USP 4/0, violet, 75 cm.
- AMP418HRP1740-2 – PDO Prime, double needle ½ circle trocar point Ultra Plus 17 mm, USP 4/0, violet, 70 cm (reinforced point for somewhat denser tissues).
- AM418HRT2240 – PDO standard, needle ½ circle trocar point 22 mm (HRT22), USP 4/0, violet, 70 cm.
As this is sterile single-use material, check the integrity of the packaging before each procedure, respect the expiry date and keep the reference and batch number for traceability in the operating theatre. Store in a cool, dry place, below 30 °C and protected from sunlight. For special B2B conditions by volume, hospital supply agreements or validation of equivalents from other commercial brands, contact the technical-commercial team at Cahnos.





Reviews