Description
1. Commercial Introduction
The Prime range absorbable monofilament polydioxanone (PDO/PDS) suture is the technical reference for prolonged absorption for professionals requiring sustained wound support, predictable handling and maximum knot security. This presentation combines a 75 cm violet grooved thread in USP 2/0 gauge with a ⅜ circle needle, body and reverse cutting triangular point of 24 mm (DS24), a geometry designed to pass through fibrous tissue with a clean, controlled passage, without tearing.
It is the B2B supply solution for operating theatres, general surgery departments, paediatric cardiovascular surgery units, ophthalmology services and hospital purchasing centres seeking quality consistency with CE marking. Synthetic absorbable medical device, sterile and single-use, supplied in a box of 12 units.
2. Technical Explanation
Polydioxanone is a synthetic absorbable polyester —a poly(p-dioxanone) polymer— that degrades by hydrolysis, a mechanism independent of the patient’s enzymatic activity that ensures a highly reproducible absorption kinetics between individuals, completing between 180 and 210 days. Being a monofilament, the thread has no wicking effect or capillarity: it reduces bacterial adherence, facilitates sliding through tissue and minimises the inflammatory response compared with multifilaments. Its firm feel and excellent knotting capacity make it ideal for ligatures and sutures of planes subjected to sustained tension during the critical healing phase.
The ⅜ circle needle (approximate curvature of 135°) with a triangular point of reverse cutting has the cutting edge on the outer face of the curve, which prevents tearing of superficial tissue and makes it especially suitable for skin, subcutaneous tissue, aponeurosis, fascia and other fibrous planes. With a length of 24 mm (DS24) and USP 2/0 gauge (metric 3), this configuration offers an excellent balance between strength and delicacy. The violet grooved colour improves thread visibility on the surgical field and knot holding. It is supplied sterile by ethylene oxide, single-use and cannot be re-sterilised.
3. Technical Specifications
3.1. General characteristics
| Parameter | Specification |
|---|---|
| Material | Polydioxanone (PDO/PDS), absorbable synthetic monofilament polymer |
| Thread structure | Grooved monofilament |
| Quality / range | Prime (PREMISUT) |
| Gauge / size | USP 2/0 (metric 3) |
| Approximate diameter | 0.30 – 0.349 mm (indicative according to Pharmacopoeia) |
| Thread colour | High-visibility violet, fluted |
| Thread length | 75 cm |
| Swaged | 1 needle per pack (single strand) |
| Needle shape | 3/8 circle (approx. 135°) |
| Needle body and point | Triangular, reverse cutting point, grooved |
| Needle size | DS24 Prime – 24 mm |
| Absorption mechanism | Hydrolysis (180–210 days) |
| Sterilisation | Ethylene oxide (single-use, not re-sterilisable) |
| Tissue reaction | Minimal, with excellent tensile strength and knot tying |
| Indications | Tissue approximation, ligatures, fascia closure, general surgery, paediatric cardiovascular tissue and ophthalmology |
| Presentation | Box of 12 sterile units · Minimum order of 30 boxes |
| Factory reference / SKU | P412DS2420 / AMP412DS2420 |
3.2. Tensile strength retention profile
| Period after implantation | Tensile strength retention |
|---|---|
| 2 weeks | 75 % |
| 4 weeks | 60–70 % |
| 8 weeks | 45 % |
| Complete absorption | 180–210 days by hydrolysis, with minimal tissue reaction |
This prolonged force retention profile is maintained during the critical healing phase of slow-healing tissues, so its use is recommended in closures of fascia, aponeurosis and structures subjected to sustained traction.
3.3. Size equivalence
| USP size | Metric size | Approximate diameter |
|---|---|---|
| 2/0 | 3 | 0.30 – 0.349 mm (indicative according to Pharmacopoeia) |
4. Equivalences and Cross-References
Polydioxanone monofilament is the material basis of the market reference lines for long-lasting absorbable sutures (PDS-type family). On the verified configuration of this data sheet (⅜ circular reverse cutting triangular needle of 24 mm, USP 2/0 gauge, 75 cm violet thread), the following commercial reference from another manufacturer has been compared in the official catalogue:
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| B. Braun / Aesculap | MonoPlus® (polydioxanone) uncoated monofilament, violet | C0024990 (USP 2/0 / EP 3, 70 cm, violet, DS24 ⅜ circle reverse cutting body needle 24 mm) | Comparable by category (not interchangeable 1:1) | Same PDO monofilament material, same USP 2/0 gauge (EP 3), same violet colour and same DS24 ⅜ circle reverse cutting needle family (24 mm). Differences: thread length (70 cm vs 75 cm) and commercial presentation. Confirmed in the official catalogue of B. Braun. |
There is no equivalence table validated by the manufacturer that officially cross-references this reference with the commercial series of other brands; consult the technical department before use in tenders. As homologous market families, to be verified code by code in their official technical data sheet before any substitution, the following are cited: the PDS II® and PDS Plus® lines from Ethicon (the latter with antibacterial triclosan coating, indicated only where its use is approved; the homologous ⅜ circle triangular reverse cutting needle geometry corresponds to its PC needle series), the Surgicryl® PDO line from SMI and the Maxon® line from Medtronic/Covidien (made of polyglyconate, a different polymer, so its absorption profile must be verified). All must be checked field by field in their official data sheet: USP gauge, thread length, colour, curvature, body, point and needle size. The brands cited belong to their respective manufacturers and are mentioned only as comparative reference.
5. Selection tips and variants of interest
When choosing a PDO suture, always check the USP gauge according to the tissue plane: fine gauges (4/0 to 6/0) for skin and subcutaneous tissue, 3/0 and 2/0 for general soft tissue —as is the case with this reference— and USP 0 and above for fascia and planes under tension. Needle geometry is also decisive: the reverse cutting triangular point (DS/HS) is indicated for fibrous and non-elastic tissues (skin, aponeurosis, fascia), while the conical point (HR/SH) is reserved for delicate and well-perfused tissues (vessels, parenchyma).
| Reference | Variant of interest |
|---|---|
| AMP412DS2410 | PDO Prime, same DS24 needle of 24 mm, USP 0, violet, 75 cm (higher gauge for fascia and planes under tension) |
| AMP412DS2430 | PDO Prime, DS24 needle of 24 mm, USP 3/0, violet, 75 cm (finer gauge for soft tissue) |
| AMP412DS2420.2 | PDO Prime, same DS24 needle of 24 mm, USP 2/0, violet, in double thread 2×75 cm (prolonged continuous sutures) |
| AMP412DS2620 | PDO Prime, DS26 needle of 26 mm, USP 2/0, violet, 75 cm (greater needle travel, same gauge) |
| AMP412DS3010 | PDO Prime, DS30 needle of 30 mm, USP 0, violet, 75 cm (wider access and larger gauge) |
| AM412DS2420 | Standard quality PDO, same DS24 configuration, USP 2/0, violet, 75 cm (range alternative) |
All references are supplied in a box of 12 units. As this is sterile single-use material, check the integrity of the packaging before each procedure, respect the expiry date and store the suture in a cool, dry place, protected from sunlight. For special B2B conditions, volume orders or other gauges, lengths and colours made to order, contact the technical-commercial team at Cahnos.