Description
1. Product Presentation
The Prime polydioxanone (PDO/PDS) absorbable monofilament suture is the reference technical choice when the tissue requires prolonged support during healing and fully predictable absorption. This presentation combines a 70 cm violet grooved thread in USP 4/0 gauge with a ⅜ circle, reverse cutting triangular body and point needle of 24 mm (DS24), configured to penetrate firm and fibrous tissues with a clean and secure passage. Available in a box of 12 units, it is a reliable and cost-effective solution for general surgery departments, paediatric cardiovascular surgery operating theatres, ophthalmology units and hospital purchasing centres seeking consistent quality with CE marking.
2. Technical Explanation
Polydioxanone is an absorbable synthetic polyester —poly(p-dioxanone) polymer— that degrades by hydrolysis, a mechanism independent of the patient’s enzymatic activity that ensures highly reproducible absorption kinetics between individuals. Being a monofilament, the thread has no wick effect or capillarity, which reduces bacterial adherence, facilitates sliding through tissue and minimises the inflammatory response compared to multifilaments. Its firm feel and excellent knotting ability make it ideal for ligatures and sutures of planes under sustained tension.
The ⅜ circle needle with reverse cutting triangular body and point has the cutting edge oriented towards the inside of the curvature, allowing it to penetrate fibrous and non-compliant tissues (skin, aponeurosis, mucosa) with less resistance and lower risk of tearing than a tapered point. The USP 4/0 gauge (metric 1.5) offers an optimal balance between strength and delicacy for subcutaneous tissue, skin and soft planes. The violet grooved colour improves thread visibility on the surgical field. It is supplied sterile by ethylene oxide, single-use and not resterilisable.
3. Technical Specifications
3.1. General characteristics
| Characteristic | Specification |
|---|---|
| Material | Polydioxanone (PDO/PDS), absorbable synthetic monofilament polymer |
| Thread structure | Grooved monofilament |
| Gauge | USP 4/0 (metric 1.5) |
| Thread length | 70 cm |
| Thread colour | Violet (high visibility) |
| Needle shape | ⅜ circle |
| Needle body and point | Triangular, reverse cutting point (Reverse Cutting) |
| Needle size | DS24 – 24 mm |
| Absorption mechanism | Hydrolysis (180–210 days) |
| Sterilisation | Ethylene oxide (single-use, not re-sterilisable) |
| Tissue reaction | Minimal, with excellent tensile strength and knot tying |
| Indications | Tissue approximation, ligatures, general surgery, paediatric cardiovascular tissues and ophthalmology |
| Presentation | Box of 12 sterile units |
| Reference / SKU | P418DS2440 / AMP418DS2440 |
3.2. Tensile strength retention profile
| Time | Tensile strength retention |
|---|---|
| 2 weeks | 75 % |
| 4 weeks | 60–70 % |
| 8 weeks | 45 % |
| Complete absorption | 180–210 days by hydrolysis |
This prolonged strength retention profile is maintained during the critical healing phase of slow-healing tissues, so its use is recommended in closures of fascia, aponeurosis and structures under sustained tension. It is supplied sterile, ready for use, in individual packaging.
4. Equivalences and Cross-References
Polydioxanone monofilament is the material basis of the market reference lines for long-lasting absorbable sutures (PDS type). On the verified configuration of this data sheet (⅜ circle reverse cutting triangular needle 24 mm, USP 4/0 gauge, violet thread 70 cm) the following commercial reference from another manufacturer has been compared:
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Johnson & Johnson MedTech (Ethicon) | PDS II® (polydioxanone) monofilament, FS-1 needle | Z441H | Comparable by category (not interchangeable 1:1) | Same PDO monofilament material, same USP 4/0 gauge, same ⅜ circle reverse cutting needle 24 mm (FS-1 vs DS24) and same thread length of 70 cm. Difference: undyed thread vs the violet of this reference and presentation in a box of 36 units. Confirmed in product data sheet from authorised distributors. |
There is no equivalence table validated by the manufacturer that officially cross-references this reference with the commercial series of other brands; consult the technical department before using in a tender. As homologous market families to verify code by code before a substitution: the PDS II® and PDS Plus® lines from Ethicon (the latter with triclosan antibacterial coating, indicated only where its use is approved), the MonoPlus® line from B. Braun/Aesculap (violet polydioxanone monofilament), the Surgicryl PDO line from SMI and the Maxon line from Medtronic/Covidien. All must be checked in their official technical data sheet (USP gauge, thread length, colour, curvature, body, point and needle size). The brands cited belong to their respective manufacturers and are mentioned only as comparative reference.
5. Selection tips and variants of interest
When choosing a PDO suture always verify the USP gauge according to the tissue plane: fine gauges (4/0 to 6/0) for skin and subcutaneous tissue —as is the case with this reference—, 3/0 and 2/0 for general soft tissue and USP 0 and above for fascia and planes under tension. Needle geometry is also decisive: the reverse cutting triangular (DS) is indicated for fibrous and non-compliant tissues, while the tapered point (HR) is reserved for delicate and well-perfused tissues (vessels, parenchyma) and the trocar point (HRT/DRT) for firmer tissues.
Within the Cahnos catalogue, the PDO/PDS family of ⅜ circle with reverse cutting triangular point includes interesting variants: AMP412DS1930 (DS19 needle 19 mm, USP 3/0, 75 cm) for tissues of greater consistency and wider field, and AM411DS1950 (DS19 needle 19 mm, USP 5/0, 45 cm) for finer and more superficial planes. Always verify the USP gauge, needle size and geometry against your surgical protocol before substituting one reference for another.
As it is sterile single-use material, check the integrity of the packaging before each procedure, respect the expiry date and store the suture in a cool, dry place, protected from sunlight. For special B2B conditions, volume orders or other gauges, lengths and colours on request, contact the technical-commercial team at Cahnos.





Reviews