Description
1. Product Presentation
The absorbable monofilament Polydioxanone (PDO/PDS) suture, reference 411DS1950, is the technical reference choice when tissue requires prolonged support during healing with fully predictable reabsorption. This presentation combines a 45 cm violet grooved thread in USP 5/0 gauge with a ⅜ circle needle, reverse cutting triangular body and point of 19 mm (DS19), a geometry designed to penetrate firm and fibrous tissues with minimal drag and excellent tip control.
Available in a box of 12 units, it is a reliable and cost-effective solution for general surgery departments, paediatric cardiovascular surgery operating theatres, ophthalmology units and hospital purchasing centres seeking quality homogeneity with CE marking and gauges conforming to USP/European Pharmacopoeia.
2. Technical Explanation
Polydioxanone is a synthetic absorbable polyester —a poly(p-dioxanone) polymer— that degrades by hydrolysis, a mechanism independent of the patient’s enzymatic activity that guarantees a highly reproducible absorption kinetics. Being a monofilament, the thread has no wicking effect or capillarity: it reduces bacterial adherence, facilitates sliding through tissue and minimises the inflammatory response compared with multifilaments. Its firm feel and excellent knotting capacity make it suitable for ligatures and sutures of planes subjected to sustained tension.
The ⅜ circle needle with reverse cutting triangular body and point (reverse cutting) has the cutting edge oriented towards the outer face of the curve, allowing penetration of resistant tissues —skin, dermis, mucosa and fibrous planes— with a lower risk of tearing and a stable suture channel. The USP 5/0 gauge (metric 1) provides an optimal balance between strength and delicacy for subcutaneous tissue, fine skin, paediatric surgery and slow-healing structures. The grooved violet colour improves thread visibility on the surgical field. It is supplied sterile by ethylene oxide, single-use and cannot be re-sterilised.
3. Technical Specifications
3.1. General characteristics
| Property | Specification |
| Material | Polydioxanone (PDO/PDS), absorbable synthetic monofilament polymer |
| Thread structure | Fluted monofilament (not braided) |
| Gauge / size | USP 5/0 (metric 1) · conforming to USP / European Pharmacopoeia |
| Thread length | 45 cm (1 strand per unit) |
| Thread colour | High-visibility violet, grooved |
| Needle shape | 3/8 circle (approx. 135°) |
| Needle body and point | Triangular, reverse cutting (reverse cutting), body and point |
| Needle size | DS19 – 19 mm |
| Swaged | 1 needle per suture |
| Absorption mechanism | Hydrolysis (180–210 days) |
| Sterilisation | Ethylene oxide (EO) — sterile, single-use, non-resterilisable |
| Tissue reaction | Minimal, with excellent tensile strength and knot tying |
| Indications | Tissue approximation, ligatures, general surgery, paediatric cardiovascular tissues and ophthalmology |
| Presentation | Box of 12 sterile units (individual pouches) |
| Factory reference / SKU | 411DS1950 / AM411DS1950 |
3.2. Tensile strength retention profile
| Time after implantation | Tensile strength retention |
| 2 weeks | 75 % |
| 4 weeks | 60–70 % |
| 8 weeks | 45 % |
| Complete absorption | 180–210 days by hydrolysis |
This prolonged force retention profile is maintained during the critical healing phase of slow-healing tissues, which is why its use is recommended in closures of fascia, aponeurosis, skin and structures subjected to sustained traction.
4. Equivalences and Cross-References
Monofilament polydioxanone is the material basis of the market reference lines for long-lasting absorbable sutures (PDS-type family). For the verified configuration of this data sheet (⅜ circle reverse cutting triangular needle of 19 mm, USP 5/0 gauge, 45 cm violet thread), no reference from another manufacturer has been found in public sources that matches field by field (material, gauge, ⅜ curvature, reverse cutting point family, 19 mm size and 45 cm length), so it is not possible to declare an exact verified reference equivalence.
There is no equivalence table validated by the manufacturer for this specific configuration; consult the technical department of Cahnos before use in tendering. As homologous market families, to be verified code by code in their official technical data sheet before any substitution, the following are cited: the PDS II® and PDS Plus® lines from Ethicon (the latter with antibacterial triclosan coating, indicated only where its use is approved), the MonoPlus® line from B. Braun/Aesculap (polydioxanone monofilament), the Surgicryl PDO line from SMI and the Maxon line from Medtronic/Covidien (made of polyglyconate, a different polymer, so its absorption profile must be verified). All must be checked field by field in their official data sheet: USP gauge, thread length, colour, curvature, body, point and needle size. The cited brands belong to their respective manufacturers and are mentioned solely as comparative reference.
5. Selection tips and variants of interest
When choosing a PDO suture, always verify the USP gauge according to the tissue plane: fine gauges (4/0 to 6/0) for skin, subcutaneous tissue and delicate tissues —as is the case with this reference—, 3/0 and 2/0 for general soft tissue and USP 0 and above for fascia and planes under tension. Needle geometry is also decisive: the reverse cutting triangular point (DS series) is indicated for fibrous and resistant tissues with controlled penetration, while the conical point (DR/HR) is reserved for delicate and well-perfused tissues (vessels, parenchyma) and the trocar point (DRT) for tissues of intermediate consistency.
Within the Cahnos catalogue, the PDO/PDS family of ⅜ circle with reverse cutting triangular point includes interesting variants: AMP411DS1950 (Prime quality, same DS19 needle, USP 5/0, 45 cm) as a higher-range alternative; AMP411DS1250 (DS12 needle, USP 5/0, 45 cm) for even smaller fields; AMP411DS1940FN (DS19 fine needle, USP 4/0, 45 cm) for slightly larger gauges; AMP412DS1930 (DS19 needle, USP 3/0, 75 cm) for planes of greater consistency; and AMP418DRT1350 (DRT13 trocar needle, USP 5/0, 70 cm) for firmer tissues. Always verify the USP gauge, size and needle geometry against your surgical protocol before substituting one reference for another.
As this is sterile single-use material, check the integrity of the packaging before each procedure, respect the expiry date and keep the suture in a cool, dry place, protected from sunlight. For special B2B conditions, volume orders or other gauges, lengths and colours made to order, contact the technical-commercial team at Cahnos.





Reviews