Description
1. Product Presentation
Synthetic absorbable surgical suture of polydioxanone monofilament (PDO/PDS), designed for the professional who needs a long-lasting thread, with smooth passage through tissue and minimal tissue reaction. Its ½ circle needle, cylindrical body and 17 mm trocar point (Tapercut) penetrates fibrous and resistant tissues with great ease, reducing trauma and tearing. Presentation in a box of 12 units, ideal for stock management in the operating theatre.
2. Technical detail and clinical behaviour
Polydioxanone is a synthetic absorbable polymer (poly(p-dioxanone)) that is reabsorbed by hydrolysis, without the need for enzymes, ensuring predictable loss of tensile strength and minimal tissue reaction. Its monofilament structure offers exceptional glide through tissue, reduced sawing effect and excellent knot-tying behaviour. The violet, grooved suture ensures optimal visibility throughout the procedure. The trocar point (Tapercut) combines a non-cutting cylindrical body with a triangular penetration tip, ideal for fascia and tissues requiring firm entry without lateral cutting effect.
3. Technical Specifications
| Composition | Polydioxanone (PDO/PDS), synthetic absorbable polymer |
| Structure | Monofilament |
| USP size | 5/0 |
| Thread length | 70 cm |
| Colour | Violet, grooved presentation |
| Needle shape | ½ circle |
| Needle body | Cylindrical |
| Tip type | Trocar / Tapercut |
| Needle size | 17 mm (HRT17) |
| Sterility | Ethylene oxide |
| Sales format | Box of 12 units |
| Manufacturer reference | 418HRT1750 |
Strength and absorption profile:
| Parameter | Value |
| Tensile strength at 2 weeks | 75 % |
| Tensile strength at 4 weeks | 60–70 % |
| Tensile strength at 8 weeks | 45 % |
| Total absorption by hydrolysis | 180–210 days |
4. Equivalences and Cross-References
There is no equivalence table validated by the manufacturer for this reference; consult the technical department before use in tenders. At category level, long-lasting monofilament polydioxanone is associated with the PDS II and PDS Plus lines (polydioxanone, Tapercut point, fine gauges such as 5/0), with the Monoplus polydioxanone line and with other monofilament PDO families on the European market. As a comparison reference from the manufacturer’s official catalogue, the PDS II family with ½ circle Tapercut needle and gauge 5/0 is marketed with longer needles (e.g. 36 mm, V-34 family, code W9380H for 70 cm suture); this difference in needle size (36 mm vs 17 mm) prevents it from being considered 1:1 interchangeable and the operating theatre protocol must be validated before any substitution.
5. Selection tips and variants of interest
The trocar point (Tapercut) is the recommended choice for fibrous tissues, fascia and ophthalmic surgery; for very delicate tissues, consider the conical point (Taperpoint) variants. Gauge 5/0 is indicated for fine surgery: paediatric cardiovascular, ophthalmology and approximation of delicate tissues. As a prolonged absorption suture (180–210 days), it is suitable when wound support is required during slow healing. In the Cahnos catalogue, the same HRT17 geometry is available in other gauges (4/0, 3/0 and 2/0), in the Prime quality variant (ref. P418HRT1750) and in the HR13 family with conical point for the same gauge. Consult your Cahnos advisor for minimum order conditions and B2B rates for purchasing groups and hospitals.





Reviews