Description
1. Product Presentation
The Prime quality absorbable monofilament polydioxanone (PDO/PDS) suture is the reference technical choice when tissue requires prolonged support during healing and fully predictable reabsorption. This presentation combines a 75 cm violet thread in USP gauge 4/0 with a ½ circle needle, cylindrical body and conical point (taperpoint) of 22 mm (HR22), designed to pass through soft, well-perfused tissue with the least possible trauma. Available in a box of 12 units, it is a reliable and cost-effective solution for general surgery departments, paediatric cardiovascular surgery operating theatres, ophthalmology units and hospital purchasing centres seeking quality consistency with CE marking.
2. Technical Explanation
Polydioxanone is a synthetic absorbable polyester —a poly(p-dioxanone) polymer— that degrades by hydrolysis, a mechanism independent of the patient’s enzymatic activity that ensures highly reproducible absorption kinetics between individuals. Being a monofilament, the thread has no wicking effect or capillarity, which reduces bacterial adherence, facilitates sliding through tissue and minimises the inflammatory response compared with multifilament. Its firm feel and excellent knotting capacity make it ideal for ligatures and sutures of planes under sustained tension.
The ½ circle needle with cylindrical body and conical point (taperpoint) separates and slides through soft tissue with minimal injury, indicated for tissue approximation, ligatures, general surgery, paediatric cardiovascular tissue and ophthalmology. USP gauge 4/0 (metric 1.5) offers an optimal balance between strength and delicacy for skin, subcutaneous tissue and soft planes with slow healing. The violet colour improves thread visibility on the surgical field. It is supplied sterile by ethylene oxide, single-use and with no possibility of re-sterilisation.
3. Technical Specifications
3.1. General characteristics
| Material | Polydioxanone (PDO/PDS), absorbable synthetic monofilament polymer |
| Thread structure | Monofilament |
| Gauge | USP 4/0 (metric 1.5) |
| Thread length | 75 cm |
| Thread colour | Violet (high visibility) |
| Needle shape | ½ circle (half circumference) |
| Needle body and point | Cylindrical, taperpoint conical point |
| Needle size | HR22 – 22 mm |
| Absorption mechanism | Hydrolysis (180–210 days) |
| Sterilisation | Ethylene oxide (single-use, not re-sterilisable) |
| Tissue reaction | Minimal, with excellent tensile strength and knot tying |
| Indications | Tissue approximation, ligatures, general surgery, paediatric cardiovascular tissues and ophthalmology |
| Presentation | Box of 12 sterile units |
| Factory reference / SKU | 412HR2240 / AMP412HR2240 |
3.2. Tensile strength retention profile
| Time | Tensile strength retention |
| 2 weeks | 75 % |
| 4 weeks | 60–70 % |
| 8 weeks | 45 % |
| Complete absorption | 180–210 days by hydrolysis |
This prolonged strength retention profile is maintained during the critical healing phase of slow-healing tissues, so its use is recommended in closures of fascia, aponeurosis, subcutaneous tissue and structures under sustained traction. It is supplied sterile, ready for use, in individual packaging.
4. Equivalences and Cross-References
Monofilament polydioxanone is the material basis of the market reference lines for long-lasting absorbable sutures (PDS type). Against the verified configuration of this data sheet (½ circle cylindrical needle with conical point of 22 mm, USP gauge 4/0, 75 cm violet thread), the following commercial reference from another manufacturer has been compared:
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence level | Technical differences / Remarks |
| Johnson & Johnson MedTech (Ethicon) | PDS II® (polydioxanone) violet monofilament, SH-1 needle | Z310H | Comparable by category (not interchangeable 1:1) | Same PDO monofilament material, same USP gauge 4/0, same ½ circle conical point taperpoint needle of 22 mm (SH-1) and violet colour. Difference: thread length of 70 cm versus 75 cm for this reference. Confirmed in the official Ethicon data sheet and authorised distributors. |
There is no manufacturer-validated equivalence table that officially cross-references this product with the commercial series of other brands; consult the technical department before use in tenders. As homologous market families to verify code by code before substitution: the PDS II® and PDS Plus® lines from Ethicon (the latter with antibacterial triclosan coating, indicated only where its use is approved), the MonoPlus® line from B. Braun/Aesculap (violet polydioxanone monofilament), the Surgicryl PDO line from SMI and the Maxon line from Medtronic/Covidien. All must be checked in their official technical data sheet (USP gauge, thread length, colour, curvature, body, point and needle size). The brands cited belong to their respective manufacturers and are mentioned solely as comparative reference.
5. Selection tips and variants of interest
When choosing a PDO suture, always verify the USP gauge according to the tissue plane: fine gauges (4/0 to 6/0) for skin and subcutaneous tissue, 3/0 and 2/0 for general soft tissue, and USP 0 and above for fascia and planes under tension. Needle shape is also decisive: the conical point (HR) is indicated for delicate, well-perfused tissues (vessels, parenchyma), while the trocar point (HRT) is reserved for firmer tissues and the reverse cutting triangular (HS) for fibrous tissues.
Within the Cahnos catalogue, the PDO/PDS family of ½ circle with conical point includes variants of interest: AM412HR2630 (HR26 needle of 26 mm, USP 3/0, 75 cm) for general soft tissue; AM412HR1740 (HR17 needle of 17 mm, USP 4/0, 75 cm) for finer tissues; AMP418HRP2240-2 (double HRP22 needle, USP 4/0, 70 cm) for anastomoses and deep planes; as well as the trocar point version AM418HRT2630 (HRT26, USP 3/0, 70 cm) for more resistant tissues. Always verify the USP gauge, size and geometry of the needle against your surgical protocol before substituting one reference for another.
As it is sterile single-use material, check the integrity of the packaging before each procedure, respect the expiry date and store the suture in a cool, dry place, protected from sunlight. For special B2B conditions, volume orders or other gauges, lengths and colours on request, contact the technical-commercial team at Cahnos.