Description
1. Product Presentation
The Prime line absorbable monofilament polydioxanone (PDO/PDS) suture is the reference technical choice when the tissue requires prolonged support during healing with fully predictable absorption and minimal tissue reaction. This presentation combines a 90 cm violet thread in USP 5/0 gauge with a double needle of ½ circle, cylindrical body and Ultra Plus trocar point of 17 mm (2xHRP17), a geometry that combines the atraumatic nature of the cylindrical body with a reinforced point of superior penetration.
Available in a box of 12 sterile units, it is the B2B supply solution for general surgery services, paediatric cardiovascular surgery operating theatres, ophthalmology units and hospital purchasing centres seeking quality homogeneity with CE marking and complete batch traceability in each package.
2. Technical Explanation
Polydioxanone is an absorbable synthetic polyester —poly(p-dioxanone) polymer— that degrades by hydrolysis, a mechanism independent of the patient’s enzymatic activity that guarantees a highly reproducible absorption kinetics between individuals, with complete absorption within 180 to 210 days: one of the longest support windows among synthetic absorbable sutures.
Being a non-grooved monofilament, the thread has no wicking effect or interstitial spaces: it reduces bacterial adherence, facilitates sliding through tissue and minimises the inflammatory response compared to braided threads. Its firm feel and excellent knotting capacity make it ideal for ligatures and planes subjected to sustained tension.
The ½ circle needle with cylindrical body and trocar point (commercial designation Ultra Plus, HRP profile) combines a body that separates fibres without cutting effect with a sharp trocar geometry point that improves initial penetration in intermediate consistency tissues, reducing surgeon effort and damage at the passage point. The double needle armament (2xHRP17) allows continuous suture techniques and approximation from both ends without rethreading, optimising performance in the operating theatre. The violet colour guarantees immediate contrast on the surgical field. Product sterile by ethylene oxide, single-use and non-resterilisable.
3. Technical Specifications
3.1. General characteristics of reference P413HRP1750-2
| SKU Cahnos | AMP413HRP1750-2 |
| Factory reference | P413HRP1750-2 (kept unchanged) |
| Thread material | Polydioxanone (PDO/PDS), poly(p-dioxanone) polymer, synthetic absorbable monofilament |
| Thread structure | Non-grooved monofilament, uncoated (Prime finish) |
| Quality / range | Prime (premium range) |
| Gauge / size | USP 5/0 (metric EP 1.0), conforming to USP / European Pharmacopoeia |
| Approximate diameter | 0.100–0.149 mm (USP/EP standard) |
| Thread length | 90 cm (1 thread · 2 needles per unit) |
| Thread colour | Violet (high visibility) |
| Needle shape | ½ circle (half circumference) |
| Needle body and point | Cylindrical body, Ultra Plus trocar point (HRP profile), with greater penetration firmness |
| Needle size | HRP17 – 17 mm, double needle (2x) |
| Swaged | Double needle, one needle at each end of the thread |
| Absorption mechanism | Hydrolysis (180–210 days) |
| Sterilisation | Ethylene oxide (EO); sterile, single-use, non-resterilisable |
| Tissue reaction | Minimal, with excellent tensile strength and knot tying |
| Indications | Tissue approximation, ligatures, general surgery, paediatric cardiovascular tissues and ophthalmology |
| Presentation | Box of 12 sterile units |
3.2. Tensile strength retention profile
| Period after implantation | Retained tensile strength |
| 2 weeks | 75% of initial strength |
| 4 weeks | 60–70% of the initial strength |
| 8 weeks | 45% of initial strength |
| Complete absorption | 180–210 days by hydrolysis, with minimal tissue reaction |
This prolonged force retention profile maintains the necessary support during the critical healing phase of slow-healing tissues, which is why it is recommended in closures of planes subjected to sustained traction and in paediatric cardiovascular surgery, where tissue growth is expected.
4. Equivalences and Cross-References
Polydioxanone monofilament is the material basis of the market reference lines for long-lasting absorbable sutures (PDS-type family). On the verified configuration of this data sheet —PDO material, USP 5/0 gauge, 90 cm violet thread, ½ circle 17 mm needle with double armament— the following homologous references have been contrasted in primary sources (official catalogue of manufacturers and authorised distributors):
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Ethicon (Johnson & Johnson MedTech) | PDS II (polydioxanone monofilament) violet, double needle RB-1 17 mm in ½ circle, conical point, USP 5/0, 90 cm | Z9108H | Comparable by category (not interchangeable 1:1) | Matches material, USP 5/0 gauge, violet colour, length 90 cm, ½ circle curvature, needle size 17 mm and double needle armament. Differs in point: conical taper point (RB-1) versus the Ultra Plus trocar point (HRP17) of this reference. Code confirmed in official Ethicon catalogue and authorised distributors. |
| Ethicon (Johnson & Johnson MedTech) | PDS Plus Antibacterial (polydioxanone with triclosan), needle RB-1 17 mm in ½ circle, conical point, USP 5/0, 90 cm, double armament | PDP9108H | Comparable by category (not interchangeable 1:1) | Same polymer, gauge, colour, length, curvature, needle size and double armament. Differences: conical point versus trocar, and antibacterial coating of triclosan which this reference lacks (indicated only where its use is approved). Code verified in authorised distributors. |
| B. Braun / Aesculap | MonoPlus (violet polydioxanone monofilament), needle HR17 17 mm in ½ circle, conical point, USP 5/0, 70 cm | C0024013 | Comparable by category (not interchangeable 1:1) | Matches material, USP 5/0 gauge, violet colour, 17 mm point and ½ circle curvature. Differences: conical point (Easyslide) versus trocar, single needle armament versus double needle and length 70 cm versus 90 cm. Code confirmed in official B. Braun catalogue. |
There is no equivalence table validated by the manufacturer that declares this reference as interchangeable 1:1 with other brands; consult the technical department before using in a tender. As homologous market families, to be verified code by code in their official technical data sheet before any substitution, the following are cited: the PDS II and PDS Plus lines from Ethicon, the MonoPlus line from B. Braun/Aesculap (violet polydioxanone monofilament), the Surgicryl PDO line from SMI, the PDO lines from DemeTECH and the Maxon line from Medtronic/Covidien (polyglyconate, different polymer, whose absorption profile must be verified). In all cases, the USP gauge, thread length, colour, curvature, body, point and needle size must be checked, as well as the armament. The brands cited belong to their respective manufacturers and are mentioned only as comparative reference.
5. Selection tips and variants of interest
When choosing a PDO suture, always check the USP gauge according to the tissue plane: fine gauges (4/0 to 6/0) for skin, subcutaneous tissue, mucosa and delicate planes —such as this reference—, 3/0 and 2/0 for general soft tissue and USP 0 and above for fascia and planes under tension. Needle geometry is also decisive: the taper point (HR/DR) is reserved for delicate and well-perfused tissues (vessels, parenchyma), the Ultra Plus trocar point (HRP) offers greater penetration firmness in tissues of intermediate consistency while maintaining an atraumatic body, and the reverse cutting triangular point (HS/DS) is used for fibrous and inelastic tissues. The double-armed configuration of this reference facilitates continuous sutures and bidirectional approximations without rethreading.
Within the Cahnos catalogue, the PDO/PDS family of ½ circle with trocar point and taper point includes variants of interest:
| Reference | Variant of interest |
| AMP412HR1710 | PDO Prime, ½ circle HR17 taper point needle, USP 0, violet, 75 cm (larger gauge, same needle) |
| AM412HR1740 | PDO standard, ½ circle HR17 taper point needle, USP 4/0, violet, 75 cm (single needle, intermediate gauge) |
| AMP413DR1750-2 | PDO Prime, double needle ⅜ circle DR17 taper point, USP 5/0, violet, 90 cm (same gauge and length, different curvature) |
| AMP418DRT1350 | PDO Prime, single needle ⅜ trocar point Tapercut DRT 13 mm, USP 5/0, violet, 70 cm (same point family, shorter needle) |
As this is sterile single-use material, check the integrity of the packaging before each procedure, respect the expiry date and keep the reference P413HRP1750-2 together with the batch number for traceability in the operating theatre. Store in a cool, dry place, below 30 °C and protected from sunlight. For special B2B conditions by volume, hospital supply agreements or validation of equivalents from other commercial houses, contact the technical-commercial team of Cahnos.