Description
1. Product Presentation
The Prime line absorbable monofilament Polydioxanone (PDO/PDS) suture is the reference technical choice when prolonged tissue support is required with fully predictable resorption and minimal tissue reaction. This presentation combines a 70 cm violet grooved thread with a ½ circle needle, reverse cutting triangular point, 26 mm (HS26), designed to penetrate firm tissues such as fascia with the least possible trauma. Available in a box of 12 units, it is a reliable and cost-effective solution for general surgery departments, paediatric cardiovascular surgery operating theatres, ophthalmology units and hospital purchasing centres seeking consistent quality with CE marking.
2. Technical explanation of the product and procedure
Polydioxanone is a synthetic absorbable polyester that degrades by hydrolysis, a mechanism independent of the patient’s enzymatic activity that ensures a highly reproducible absorption kinetics between individuals. Being a monofilament, the thread has no wicking effect or capillary edges, which reduces bacterial adherence and facilitates sliding through tissue; its firm feel and excellent knotting capacity make it ideal for ligatures and sutures of planes under sustained tension. The ½ circle needle (half-circumference curvature) with reverse cutting triangular point (reverse cutting) penetrates tissue with a cutting edge that avoids tearing in fibrous walls, and its triangular body provides maximum stability in the needle holder. The USP 0 gauge (metric 3.5) is versatile: suitable for subcutaneous tissue, superficial muscle planes, soft tissue approximation, fascia closure and ligatures. The grooved violet colour improves thread visibility on the surgical field. It is supplied sterile by ethylene oxide, single-use and cannot be re-sterilised.
3. Technical Specifications
3.1. General characteristics of reference P418HS2610
| Characteristic | Specification |
| Material | Polydioxanone (PDO/PDS), absorbable synthetic monofilament polymer |
| Thread structure | Grooved monofilament |
| Quality | Prime (range PREMISUT) |
| Gauge | USP 0 (metric 3.5) |
| Thread colour | Violet |
| Needle shape | ½ circle (half circumference) |
| Tip type | Reverse cutting triangular (reverse cutting) |
| Needle size | HS 26 mm |
| Thread length | 70 cm |
| Absorption mechanism | Hydrolysis (180-210 days) |
| Sterilisation | Ethylene oxide (single-use, not re-sterilisable) |
| Tissue reaction | Minimal, with excellent tensile strength and knot tying |
| Indications | Tissue approximation, ligatures, general surgery, fascia, paediatric cardiovascular tissues and ophthalmology |
| Presentation | Box of 12 units · Minimum order 30 boxes |
| Manufacturer reference | P418HS2610 |
| SKU Cahnos | AMP418HS2610 |
3.2. Tensile strength retention profile
| Time | Tensile strength retention |
| 2 weeks | 75 % |
| 4 weeks | 60-70 % |
| 8 weeks | 45 % |
| Complete absorption | 180-210 days by hydrolysis |
This strength retention profile is maintained during the critical healing phase of slow-healing tissues, which recommends its use in closures of fascia, aponeurosis and structures under sustained tension.
4. Equivalences and Cross-References
Monofilament polydioxanone is the material basis of the market reference lines for long-lasting absorbable sutures. Based on the verified configuration of this technical data sheet (½ circle reverse cutting triangular needle, 26 mm, USP 0 gauge, 70 cm violet thread), the following commercial reference from another manufacturer has been cross-checked against primary sources:
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence Level | Technical Differences / Remarks |
| Ethicon (Johnson & Johnson MedTech) | PDS II, CP-2 needle | Z970H | Comparable by category (not interchangeable 1:1) | Same PDO monofilament violet material without coating, same USP 0 gauge, same ½ circle curvature and reverse cutting point family (CP-2) of 26 mm, 70 cm thread (27″). Confirmed on the GUDID/FDA regulatory basis (ETHICON, LLC, class II, catalogue Z970H) and in authorised distributors. |
As homologous market families to be verified code by code before substitution: Ethicon’s PDS Plus line (same PDO base with antibacterial triclosan coating), B. Braun’s MonoPlus line (violet polydioxanone monofilament) and SMI’s Surgicryl PDO line, which must be checked in their official technical data sheet (gauge, geometry and needle length). There is no equivalence table validated by the manufacturer for this reference; consult the technical department before use in tenders.
5. Selection tips and variants of interest
When choosing a PDO suture, always check the USP gauge according to the tissue plane: fine gauges (4/0 to 6/0) for skin and subcutaneous tissue, 3/0 and 2/0 for general soft tissue, and USP 0 and above for fascia and planes under tension. The needle shape is also decisive: the ½ circle reverse cutting triangular (HS) is designed for fibrous and somewhat firm tissues, while the taper point (HR) is reserved for delicate tissues such as vessels.
Within the Cahnos catalogue, the PDO/PDS family of ½ circle reverse cutting triangular includes several configurations: the same HS26 needle in standard quality (AM418HS2610), the HS36 needle in USP gauges 2/0, 0 and 1 in standard and Prime versions (AM412HS3620, AM412HS3610, AMP412HS3610, AM412HS3601), the HS40 and HS48 needles with 90 cm thread (AM413HS4001, AM413HS4010, AM413HS4801, AM413HS4810) and HR taper point variants for delicate tissues (AM412HR2630). As this is sterile single-use material, check the integrity of the packaging before each procedure and store it in a cool, dry place, protected from sunlight. For special B2B conditions, volume orders or other gauges, lengths and colours on request, contact the technical-commercial team at Cahnos.