Description
1. Product Presentation
The Helix Test PCD Kit + steam strips KH2X15-35 from the CHEMDYE® range (Pergut Medical) is the complete solution that every sterilisation centre needs to validate the correct removal of air and steam penetration in vacuum-assisted autoclaves, both large and small. Each kit includes the Helix-PCD process challenge device with reusable metal capsule, together with 250 simulation strips self-adhesive with Type 2 chemical indicator for the Bowie & Dick cycle. A system ready for daily use in hospitals, clinics, dental practices, pharmaceutical industry and medical material processing companies, which guarantees the safe release of each load.
2. Technical explanation of the procedure
The Bowie & Dick test is a steam penetration test that checks the air removal efficiency of steam sterilisers with dynamic air extraction (pre-vacuum and vacuum pump). If residual air pockets remain in the chamber, the steam does not reach the load under saturation conditions and the process may not be effective, even if the recorded time and temperature parameters are correct.
The Helix-type design reproduces the worst-case hollow load scenario: the steam must travel through a high-resistance plastic tube of 1.5 m length with only 2 mm internal diameter before reaching the chemical indicator housed in the metal capsule. The special silicone connector, covered by an external connector, guarantees a hermetic seal and a secure connection to the sterilisation equipment. The indicator strip PCD-A-3.5BD (Type 2 in accordance with ISO 11140-1), printed with steam-sensitive ink, changes colour uniformly only when the air has been removed and saturated steam penetrates under correct conditions, allowing the load to be released safely.
3. Technical Specifications
Technical data sheet for the KH2X15-35 kit:
| Characteristic | Specification |
|---|---|
| Reference | KH2X15-35 (CHEMDYE® range) |
| Product type | Helix-PCD kit + steam simulation strips |
| Processing | Steam in large and small vacuum-assisted sterilisers |
| Application | Evaluation of air removal and steam penetration (Bowie & Dick test) |
| Capsule | Metal, highly durable, designed to house the chemical indicator of the Helix-PCD system |
| Connection tubing | High-resistance plastic, 1.5 m length and 2 mm internal diameter; other sizes available on request |
| Connector | Special silicone connector covered by an external connector, with hermetic seal |
| Chemical indicator | PCD-A-3.5BD, Type 2 (ISO 11140-1), certified for the Bowie & Dick cycle |
| Cycle conditions | Bowie & Dick 3.5 min at 134 ºC; alternative cycle 132 ºC for 4 min |
| Simulation strips | Self-adhesive printed with indicator ink, 77 x 6 mm (10 strips per sheet) |
| Large format | 80 x 100 mm strips |
| Presentation | 1 Helix-PCD device + 250 strips per box |
| Helix-PCD service life | Reusable up to 250 cycles |
| Regulatory standard | ISO 11140-1, ISO 11140-4 and EN 867-5 |
| Manufacturer | Pergut Medical, S.L.U. (Spain) |
4. Equivalences and Cross-References
This kit belongs to the Helix process challenge device family from Pergut Medical (CHEMDYE® range). The following references of interest are located in the same Helix-PCD system:
| Reference | Description | Presentation |
|---|---|---|
| KH2X15-35 | Helix Test PCD kit + steam simulation strips (current product) | 250 strips/box |
| KH2X15-35B | Helix-type Bowie & Dick test for steam (metal device, ISO 11140-4) | On request |
| BD125X/20 | Helix-type Bowie-Dick test for steam (1.5 m x 2 mm), single-use | 20 uds/caja |
| BD125X/1 | Bowie-Dick test package with pre-notice sheet | 20 uds/caja |
| 24102 | Bowie-Dick test package for daily control (120 x 120 x 17 mm) | 20 uds/caja |
| KH2X15F1P | Helix Test PCD kit + strips for vaporised formaldehyde (VFBT / LTSF) | 250 strips/box |
| KH2X12-P1 | Helix Test PCD kit + strips for plasma (hydrogen peroxide) | 100 strips/box |
As complementary consumables for steam processes, Pergut Medical offers the CD30 multivariable chemical indicator (Type 4), indicator adhesive tapes and biological steam controls, all available in the Cahnos catalogue. Internationally, functionally equivalent kits are marketed with identical conformity to ISO 11140-1 Type 2 and EN 867-5.
5. Selection tips and variants of interest
Hollow load control: if your autoclave processes long-lumen instruments (endoscopes, laparoscopes, cannulas, etc.), the Helix-type test is the most demanding and the recommended one for validating steam penetration at the worst point of the load.
Other processes: for validating hydrogen peroxide cycles, reference KH2X12-P1 is available, and for vaporised formaldehyde, KH2X15F1P. For daily control of large autoclaves with vacuum pump, the classic reference is the Bowie-Dick pack 24102.
Frequency of use: perform the test periodically following the equipment manufacturer’s protocol and standard ISO 11140-1. The Helix-PCD supports up to 250 cycles; replace the PCD-A-3.5BD strips when they run out and always check that the strip’s colour change is uniform and complete before releasing the load.
Storage and handling: keep the product in its original packaging, in a cool, dry place, protected from light and away from sterilising agents. Check the integrity of the packaging before each cycle.
Professional B2B purchase: for sterilisation centres, hospitals, pharmaceutical industry and medical device processing companies, consult our technical-commercial team about volume conditions, continuous supply contracts and dose customisation.