Description
1. Equipment presentation
The 29-litre EXL autoclave is the high-capacity sterilisation solution designed for healthcare centres with high patient turnover: dental clinics with several surgeries, medical practices, podiatry and aesthetic medicine centres, laboratories and veterinary clinics. Its extra-wide chamber reduces the number of daily cycles, optimises instrument management and keeps the safety of the class B sterilisation process intact. In addition to its large useful volume, the EXL stands out for its low water and electricity consumption, its colour touchscreen and a complete and expandable digital traceability system, making it one of the most cost-effective options on the market for intensive working environments. With 5 XL trays of 55 cm, up to 7 kg of load and Ethernet connectivity as standard, it covers the needs of services that demand maximum capacity without giving up the documentary control of each cycle.
2. Technical explanation of the equipment
The EXL is a steam steriliser of class B compliant with the standard EN 13060, which allows it to process, by means of fractionated pre-vacuum, bagged solid loads, hollow instruments, porous materials and unwrapped loads. The weld-free stainless steel chamber, certified according to the PED 2014/68/EC directive, offers with its 29 litres 45% more useful space than a 24-litre autoclave and 275 cm of linear load length, accommodating up to 5 XL trays of 55 cm (17 cm longer than those of a 24 L unit) or, by turning the tray holder, 3 rows of bagged cassettes.
The unit incorporates an instantaneous steam generator that produces steam at the moment it enters the chamber, reducing heating times and energy consumption. The Pro System control systems —integrated conductivity meter, air-water separator, Inspection System and Dirt Control System— guarantee water quality, chamber cleanliness and long-term process reliability. The colour touchscreen allows selection of 5 type B cycles (121 °C, 134 °C, 134 °C Prion, 134 °C rapid and 134 °C Prion rapid) and type S and N cycles protected by password, in addition to the verification tests Vacuum, Helix and Bowie & Dick programmable with reminder. The E-Timer function allows programming the start of cycles and tests before the clinic opens. Regarding traceability, the SD card and Ethernet connection come as standard, with E-Memory and E-Data software included, and can be expanded on paper with three optional printing kits.
3. Technical Specifications
Technical data sheet of the 29-litre EXL autoclave (SKU Cahnos 115760):
| Parameter | Specification |
|---|---|
| Reference / SKU Cahnos | 115760 |
| Brand / Range | Pro System range, model EXL (CE Type Ref: EXL) |
| Set type | Class B benchtop steam autoclave |
| Compliance | Class B compliant with EN 13060; chamber certified PED 2014/68/EC |
| Regulatory classification | Medical device class IIb; CE Ref. SSUX30L2019 (according to the manufacturer’s official data sheet) |
| Chamber capacity | 29 litres |
| Chamber material | Stainless steel without welds |
| Maximum load | 7 kg (bagged solid products) |
| Trays | 5 XL trays included (18.5 x 2 x 55 cm); standard tray holder 19.5 x 19.5 x 57 cm |
| Linear capacity | 275 cm; 45% more useful space than a 24 L autoclave |
| Sterilisation cycles | 5 type B cycles: 121 °C, 134 °C, 134 °C Prion, 134 °C fast and 134 °C fast Prion |
| Additional cycles | Type S (S134, S121) and type N (N134, N121), password-activated |
| Verification tests | Vacuum, Helix and Bowie & Dick; programmable with reminder |
| Display | Colour touchscreen, intuitive interface with clear icons |
| Steam generator | Instant; parameters adapted to the effective load |
| Connectivity | Ethernet and SD card as standard; optional WiFi (E-WiFi kit) |
| Management software | E-Memory and E-Data included |
| External dimensions | 46 x 45.5 x 85.5 cm (length x height x width) |
| Weight | 62 kg (empty machine with tray holders and trays) |
| Power supply | 230 V, 50/60 Hz |
| Power consumption | 3200 W (14 A) |
| Water consumption | 0.7–0.9 litres per cycle according to the manufacturer’s documentation version |
| Clean water tank | 5 litres |
| Noise | 60 dB |
| Operating temperature | +5 °C to +40 °C |
Traceability options and compatible accessories from the same range:
| Accessory | Function | Reference |
|---|---|---|
| Print Set 1 | Internal thermal paper printer: printed cycle record for filing | 120010 |
| Print Set 2 | Internal adhesive label printer with cycle data and barcodes | 120011 |
| Print Set 3 | External label printer for continuous production and large quantities | 120012 |
| E-WiFi kit | Wireless connectivity of the autoclave with the clinic network | Optional |
| E-Cart | Multifunction trolley compatible with EXL for space organisation | Optional |
| Aquabox / Aquafilter / Aquaosmo | Feed water treatment and control systems | Optional |
4. Equivalences and Cross-References
The Cahnos SKU for this equipment is 115760. Cahnos uses its own code for its catalogue; the manufacturer’s official documentation identifies the equipment as EXL, with CE Type Ref: EXL and CE Ref. SSUX30L2019 (according to the manufacturer’s official data sheet). Before using any of these codes in a tender or technical specification, it is recommended to confirm the current reference directly with the manufacturer or with the Cahnos technical department, as the range has evolved through successive versions.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| MELAG | Vacuklav 24 BL+ | Vacuklav 24 BL+ | Comparable by category (not interchangeable 1:1) | Class B according to EN 13060 with a deep chamber of 60 cm and 29 litres; load of 7 kg of instruments and 2.5 kg of textiles; differs in mandatory fixed water connection, power consumption (2100 W) and proprietary documentation systems (MELAprint / MELAnet). Verify the technical data sheet and IFU field by field. |
| W&H | Lisa 22 | Lisa 22 | Comparable by category (not interchangeable 1:1) | Class B benchtop autoclave with fractionated pre-vacuum, EliSense screen and EliTrace traceability; differs in capacity (22 L vs 29 L), shorter trays and Eco Dry+ drying technology specific to W&H. |
| Tuttnauer | Elara 11 | Elara 11 | Comparable by category (not interchangeable 1:1) | Class B benchtop autoclave with pre and post vacuum; chamber of 28.5 litres with 5 trays (data from Tuttnauer’s official brochure); differs in power consumption (approx. 2300 W), external dimensions and proprietary control and connectivity platform. |
| IcanClave | STE-29D | STE-29D | Comparable by category (not interchangeable 1:1) | Class B autoclave of 29 litres with extra-long chamber (~600 mm), printer and USB; differs in power (2500 W), brand of another origin and absence of the Pro System connectivity and traceability platform. |
There is no equivalence table validated by the manufacturer that cross-references reference 115760 with codes from other brands; consult the technical department before using in a tender. All class B autoclave manufacturers use proprietary platforms, control systems and software, so equipment from other brands is not interchangeable 1:1 with this autoclave without prior technical validation. Within the manufacturer’s own range, the homologous variants available in the Cahnos catalogue are the E8 24 L (SKU 115716), entry-level model; the E9 24 L (SKU 115712); and the E10 24 L (SKU 115750), a higher generation of 24 litres. They share platform, accessories and control systems, but differ in features, connectivity and price.
5. Selection tips and variants of interest
- Adjust the capacity to your activity: the 29-litre EXL is the recommended option for clinics with high patient volume, long instruments (55 cm trays) or continuous sterilisation of cassettes. If your volume is medium and you are looking for a balanced unit, consider the E10 or the E9 of 24 litres; if you need an economical entry point, the E8 of 24 litres (115716).
- Plan traceability from the start: the SD card and Ethernet connection come as standard with the E-Memory and E-Data software, sufficient to archive, search, print and send the records of each cycle. If you need physical proof on the load, install Print Set 1 (thermal paper, ref. 120010), Print Set 2 (adhesive labels with barcode, ref. 120011) or Print Set 3 for large volumes (ref. 120012). The kits can be installed at any time, even on autoclaves already in service.
- Complete the connectivity: if you wish to view the data from mobile devices or integrate the unit into the clinic network without cabling, add the optional E-WiFi WiFi kit.
- Water quality: use distilled or demineralised water of controlled quality; the integrated conductivity meter monitors the water continuously and, together with the air-water separator and the Pro System control systems, protects the hydraulic circuit and extends the service life of the chamber.
- Maintenance and checks: carry out the periodic tests (Vacuum, Helix, Bowie & Dick) using the equipment’s programming and reminder, and take advantage of the easy access to the tank and the removable side panels for quick and effective cleaning.
- For volume purchases, supply agreements or validation of equivalents from other brands, contact the technical-commercial team at Cahnos.