Description
1. Product Presentation
The Bowie Dick Helix Test Pack (ref. 35951) is the process control system that every sterilisation centre needs to reliably verify air removal and steam penetration in steam autoclaves, both large units with vacuum pump and bench-top class B sterilizers. Each box contains 100 tests ready to be incorporated into the daily validation routine of hospitals, clinics, dental practices and medical device processing companies. Thanks to its helix design, it detects cycle failures much earlier than other chemical indicators or integrators, ensuring the safe release of each load.
2. Technical explanation of the procedure
The Bowie & Dick test is a steam penetration test that checks the air removal efficiency of steam sterilizers with dynamic air extraction (pre-vacuum and vacuum pump). If residual air pockets remain in the chamber, the steam does not reach the load under saturation conditions and the process may not be effective, even if the recorded time and temperature parameters are correct.
The helix design reproduces the worst-case hollow load scenario: the steam must travel through a 1.5 m tube before reaching the chemical indicator housed in the device. The indicator strip, sensitive to water vapour, changes colour clearly and uniformly to pink only when the air has been removed and saturated steam penetrates under correct conditions. Standard EN 285:2009 specifies the Helix test (Hollow A) as a Bowie & Dick test and for batch monitoring when sterilising hollow loads; furthermore, this helix test is mandatory for class B (bench-top) steam sterilizers.
3. Technical Specifications
The set includes the cotton bag, the helix device with 1.5 m tube, 100 tests and the user manual in GB, FR, IT and ES.
| Characteristic | Specification |
|---|---|
| Manufacturer | GIMA |
| Manufacturer reference | 35951 |
| Sales unit | Box of 100 units |
| Kit contents | Cotton bag, helix device (1.5 m tube), 100 tests and user manual (GB, FR, IT, ES) |
| Device type | No medical device |
| EAN | 8023279359510 |
| Indicator classification | ISO 11140-1 Type 2 – EN 867 Part 5 Class B |
| Cycle conditions | PCD guaranteed for 500 cycles (134 °C – 3.5 min) |
| Colour change | Clear and uniform colour change to pink |
| Interpretation | Easy to interpret and document |
| Safety | Non-toxic / lead-free |
4. Equivalences and Cross-References
There is no equivalence table validated by the manufacturer that cross-references code 35951 with codes from other brands on a 1:1 basis. Functionally, the Cahnos catalogue includes other helix-type Bowie & Dick test devices with identical compliance to ISO 11140-1 Type 2 and EN 867-5, such as the Helix Test PCD Kit for steam CHEMDYE® (KH2X15-35), the 1.5 m x 2 mm helix Bowie-Dick test (BD125X/20) or the Bowie-Dick pack for daily control of large autoclaves (24102). Before proposing a substitution in a specification or tender, it is recommended to compare the technical data sheet and cycle conditions with the technical department of Cahnos.
5. Selection tips and variants of interest
Hollow load validation: if your autoclave processes long-lumen instruments (endoscopes, laparoscopes, cannulas), the helix test is the most demanding and the recommended one for verifying steam penetration at the worst point of the load. Frequency of use: perform the test following the equipment manufacturer’s protocol and standard ISO 11140-1, and always check that the colour change to pink is uniform and complete before releasing the load. Other processes: for daily validation of large autoclaves with vacuum pump, the Bowie-Dick pack 24102 can be used, and for low-temperature cycles (hydrogen peroxide or formaldehyde) the Helix-PCD family from CHEMDYE® (refs. KH2X12-P1 and KH2X15F1P). Store the product in its original packaging, in a cool, dry place away from light. For B2B volume or continuous supply conditions, consult the technical-commercial team of Cahnos.