Description
1. Commercial Introduction
The TheranoJet®ARA is the reference shielded injection system from the French manufacturer Lemer Pax for Nuclear Medicine and Radiology departments performing radiopharmaceutical therapies. Developed in collaboration with the Beaujon AP-HP Hospital in Paris, this injection pump shield allows the intravenous administration of radiopharmaceutical products for vectorised internal radiotherapy (RIV), also known as radioligand therapy (RLT) or radiopharmaceutical therapy (RPT), particularly labelled with lutetium-177 (177Lu). It is a single-use per patient, lightweight, mobile and versatile solution, designed to protect healthcare staff throughout the entire injection procedure without compromising patient comfort.
2. Technical explanation of the equipment
The TheranoJet®ARA acts as a mobile biological barrier between the radioactive source and the operator, integrating into a single stainless steel 304L structure all the elements necessary for the safe administration of the radiopharmaceutical. The vial is placed in a shielded vial shield (30 ml) mounted on a rotating support that allows the vial to be inverted, ensuring complete administration of the contents and facilitating rinsing of the vial by inversion, which recovers residual activity and transfers it to the patient.
The bidirectional infusion pump incorporates a high-pressure detector and an air bubble detector, avoiding the risk of extravasation or air injection; its configurable screen allows the flow rate to be adapted to each clinical protocol. The connection to the vial is made via a sterile needle-free transfer system, which minimises the risk of contamination and accidental needlestick injuries. The equipment has a removable containment tray that retains the radiopharmaceutical in the event of any incident and facilitates microbiological and radioactive decontamination without altering the components of the unit, as well as a transparent organic protective screen that allows the line and retention area to be visualised at all times.
3. Technical Specifications
General characteristics
| Parameter | Specification |
| Product | Shielded injection pump shield for theranostic products |
| Manufacturer | Lemer Pax (France) |
| Manufacturer reference | 00050009 |
| External dimensions (with drip stand) | L 712 x D 759 x H 1,760 mm |
| Material | Frame and drip stand in stainless steel 304L |
| Weight | 86 kg |
| Mobility | 4 twin castors (2 with brake), 2 side handles with height-adjustable grip |
| Protection principle | ALARA (as low as reasonably achievable) |
Shielding and radiological protection
| Item | Specification |
| Transparent organic screen | 0.5 mm lead eq. (dimensions L 220 x H 170 mm) |
| Lower protective housing | Lead, eq. 2 mm |
| Shielded vial shield (ref. 00050036) | Lead glass + lead; vial volume 30 ml; weight with lid 3.1-3.9 kg |
| Maximum manageable activity | 7,400 MBq of 177Lu for a dose rate below 100 µSv/h at 5 cm from the walls |
Compatible radionuclides
| Therapy type | Radionuclides |
| Targeted internal radiotherapy (RIV/RPT) | 177Lu, 161Tb, 223Ra, 225Ac, 212Pb, 211At |
Integrated equipment and options
| Component | Reference |
| Bidirectional injection pump | Integrated |
| Patient injection kit | 00055414 |
| Shielded vial shield | 00050036 |
| Shielded connection system for the vial shield | 00051943 |
| Forceps for removing the vial shield connection | 00055988 |
| Shielded decay cover | 00054547 |
| Shielded container for transporting the vial shield (option) | 00051923 |
4. Equivalences and Cross-References
The TheranoJet®ARA corresponds to the factory reference 00050009 from Lemer Pax, marketed by Cahnos under the code CAH-00050009. Its associated shielded vial shield (ref. 00050036) is a specific component of this system, although functionally it can be compared with the medium-energy radiopharmaceutical vial shields available in the Cahnos catalogue, such as the VialShield Pro (CAH-VSP), the 20 mm leaded glass container protector (CAH-LP0016) or the mixed PMMA/lead/leaded glass vial shields (CAH-BVS-MIX-BME).
For the protection of syringes in the preliminary phases of dose preparation and measurement, it is recommended to complete the workflow with tungsten syringe shields from the same Lemer Pax family: Mediclic LME (CAH-LP11453), Mediclic HE (CAH-LP11653) or the Easyview range (CAH-LEM21896 and CAH-LEM24946). For the transport of loaded vials and syringes between the radiopharmacy and the injection room, the catalogue offers syringe carriers and shielded cases such as the SC10 (CAH-LPSC10), the EasyBOX Carrier (CAH-LEM13222) and the Crocobox case (CAH-LM00038013).
5. Selection tips and variants of interest
Before integrating the TheranoJet®ARA into your department, verify the radionuclide and activity of your protocols: the equipment is optimised for RIV therapy labelled with 177Lu (up to 7,400 MBq) and supports other emitters such as 161Tb, 223Ra, 225Ac, 212Pb and 211At, as well as some diagnostic protocols. Confirm the compatibility of the vial shield with the 30 ml vials used in your radiopharmacy and consider acquiring the shielded transport container (option ref. 00051923) to move the vial safely from the hot cell to the injection room.
Plan the replenishment of consumables, especially the patient injection kit (ref. 00055414) and the needle-free sterile connection systems, as well as periodic decontamination using the removable tray. If your department combines therapy and diagnostics, complete your comprehensive radiation protection strategy with syringe shields, vial shields, transport cases and waste containers from the Cahnos range. Our technical-commercial team can advise you on the optimal configuration for your centre and process customised quotes for comprehensive equipment projects in Nuclear Medicine and Radiology.







Reviews