Description
The Easyview LME range from Lemer Pax represents the complete solution for radiation protection in Nuclear Medicine with low and medium energy radiopharmaceuticals. These tungsten syringe shields, distributed by Cahnos, are specifically designed for the daily tasks of preparing, measuring and injecting doses, ensuring maximum safety for the operator, the patient and the entire department. Its combination of effective shielding, careful ergonomics and low maintenance makes it the ideal choice for radiopharmacies, injection rooms and Nuclear Medicine departments seeking to optimise their processes while complying with the ALARA principle (keeping doses as low as reasonably possible).
Technical explanation
The Easyview LME acts as a biological barrier between the radioactive source and the operator. Its high-density tungsten body, with a minimum thickness of 2 mm, effectively attenuates the gamma radiation emitted by the most common low and medium energy radionuclides in SPECT scans, such as technetium-99m, indium-111, iodine-123, thallium-201 or lutetium-177, reducing the dose received by the hands and torso throughout all syringe handling.
The large viewing window incorporates 8 mm lead glass (density 5.2) with a zoom effect, allowing the contained volume and the syringe graduations to be observed in full, facilitating precise dose measurement and minimising errors. The clamping system securely holds the syringe during direct injection into the patient, connection to a catheter or a three-way stopcock, preventing unwanted movement. In addition, the lead glass can be replaced by the user by removing a single screw, simplifying cleaning, decontamination and periodic maintenance of the equipment.
Technical specifications
| Parameter | Specification |
|---|---|
| Product | Shielded syringe protector for low and medium energy |
| Manufacturer | Lemer Pax (France) |
| Range | Easyview LME (Low & Medium Energy) |
| Shielding material | High-density tungsten |
| Shielding thickness | Minimum 2 mm tungsten |
| Viewing window | Lead glass 8 mm (density 5.2) with zoom effect |
| Attenuation | Greater than 99% |
| Energy range | Low and medium energy (SPECT gamma emitters) |
| Compatibility | Compatible with all syringes on the market |
| Volume range | 1 to 10 ml |
| Ergonomics | Easy grip and clear reading of the graduations with zoom effect |
| Maintenance | Replaceable lead glass by removing a single screw |
| Protection principle | ALARA |
| Application | Preparation, measurement and injection of radiopharmaceuticals in Nuclear Medicine |
| Reference (SKU Cahnos) | CAH-LEM21896 |
Compatible low and medium energy radiopharmaceuticals:
| Radionuclide | Main energy | Typical application |
|---|---|---|
| Technetium-99m (99mTc) | 140 keV | SPECT diagnostics |
| Indium-111 (111In) | 171/245 keV | SPECT and cell labelling |
| Iodine-123 (123I) | 159 keV | SPECT |
| Thallium-201 (201Tl) | 135/167 keV | Perfusion SPECT |
| Lutetium-177 (177Lu) | 208 keV | Theranostics |
Equivalences and cross-references
The Easyview LME is functionally equivalent to the low and medium energy syringe shields from other radiation protection manufacturers. Within the Cahnos catalogue, its direct equivalent is the Mediclic LME (CAH-LP11453), also from Lemer Pax, which shares the same design philosophy with 2 mm tungsten shielding and a lead glass window. For high energy radiopharmaceuticals, the same family offers the Easyview HE (CAH-LEM24946) and the Mediclic HE (CAH-LP11653), while the shielded SyringeShield Pro (CAH-SSP) and the low energy beta emitter shield (CAH-BSS-MIX-BE) complete the range of syringe protection in radiopharmacy.
| Related product | Manufacturer | Reference Cahnos | Notes |
|---|---|---|---|
| Mediclic LME | Lemer Pax | CAH-LP11453 | Direct equivalent for low and medium energy |
| Easyview HE | Lemer Pax | CAH-LEM24946 | High energy version (SPECT and PET), non-magnetic |
| SyringeShield Pro | Cahnos | CAH-SSP | Low and medium energy with storage support (up to 8) |
| Beta emitter shield | Cahnos | CAH-BSS-MIX-BE | Mixed shielding of tungsten, PMMA and lead glass |
For the transport of already loaded syringes, the complementary systems in the range are recommended: shielded syringe holder SC10 (CAH-LPSC10), EasyBOX Carrier case (CAH-LEM13222) and Crocobox briefcase (CAH-LM00038013).
Selection tips and variants of interest
To choose the most suitable configuration of the Easyview LME range, check the radionuclide being used and its energy spectrum (the usual low and medium energy ones: 99mTc, 111In, 123I, 201Tl and 177Lu), as well as the volume, from 1 to 10 ml, and the length of the syringes used in your protocols; the range is compatible with all syringes on the market. If your department also handles high energy PET emitters (18F, 68Ga), consider incorporating the high energy versions of the family: Easyview HE (CAH-LEM24946) or Mediclic HE (CAH-LP11653), to adapt the protection to the energy spectrum of each isotope.
Plan the periodic replacement of the lead glass according to frequency of use and take advantage of rapid access via a single screw for thorough decontamination. Complete your comprehensive radiation protection strategy with vial shields, table and injection shields, PRA3/PRA10 needle containers and shielded transport systems. Our technical-commercial team can advise you on the optimal configuration for your facility and process customised quotations for comprehensive equipment projects.








Reviews