Description
1. Presentation
The NYLON Suture PREMISUT Prime 1/2 Triangular HS40 is the reference in the Prime range of non-absorbable sutures from Cahnos for procedures requiring a high-performance permanent thread mounted on a 40 mm reverse cutting needle. It combines a nylon monofilament (polyamide 6 or 6.6) in identifying black, USP 1 gauge and 75 cm of grooved uncoated thread, with a ½ circle needle with reverse cutting triangular point (HS40 profile).
It is a common choice in ophthalmology, neurosurgery and general surgery departments, as well as in purchasing centres and tenders that value in the Prime range superior sharpening precision, batch-to-batch consistency and full traceability. Sterile, ready-to-use product, presented in a box of 12 units in individual sachets with batch traceability.
2. Technical explanation
Nylon (polyamide 6 or 6.6) is a non-absorbable synthetic polymer that remains in the tissue indefinitely, gradually encapsulating and maintaining high tensile strength for more than three months, comfortably covering the critical healing phase. Being an uncoated monofilament, the thread slides easily through tissues, reduces the wicking effect and bacterial drag compared with multifilament, and minimises the inflammatory reaction. The grooved finish of the strand improves knot retention and handling.
The 40 mm HS40 needle features a ½ circle curvature with a reverse cutting triangular point: the cutting edge cuts in the opposite direction to the wound, which preserves the skin edge and prevents the cut from widening in the puncture bed, facilitating passage through firm, non-distensible tissues. The USP 1 gauge (metric EP 4) provides high strength for tissues under medium tension, and the 75 cm thread offers sufficient length for continuous sutures or multiple stitches without re-threading. Sterilisation by ethylene oxide (EO); single-use product, not re-sterilisable.
3. Technical Specifications
3.1. General characteristics
| Parameter | Specification |
| Brand / quality | PREMISUT Prime (premium range) |
| Thread material | Nylon monofilament (polyamide 6 or 6.6), uncoated |
| Thread structure | Monofilament, grooved surface |
| Absorption | Non-absorbable; gradual encapsulation, wound support for more than 3 months |
| Gauge / size | USP 1 (EP metric 4), conforming to USP and European Pharmacopoeia gauges |
| Approximate thread diameter | 0.40 – 0.499 mm |
| Thread length | 75 cm (free end, without loop) |
| Thread colour | Black identifier (high visibility in surgical field) |
| Needle shape | 1/2 circle (½ circle) |
| Tip type | Triangular reverse cutting (Reverse Cutting) |
| Needle size | HS 40 (40 mm) |
| Configuration | 1 needle per unit, grooved thread |
| Sterilisation | Ethylene oxide (EO) – sterile |
| Use | Single-use, sterile. Do not re-sterilise |
| Indications | Approximation of soft tissues, ligatures, ophthalmology and neurosurgery |
| Commercial presentation | Box of 12 units (sterile individual sachets); batch traceability |
| Factory reference (Prime range nomenclature) | P642HS4001 |
| Reference Cahnos (SKU) | AMP642HS4001 |
3.2. Gauge equivalence according to USP/EP Pharmacopoeia (synthetic non-absorbable suture)
| USP size | Metric measurement (EP) | Approximate thread diameter |
| 2/0 | 3 | 0.30 – 0.339 mm |
| 0 | 3.5 | 0.35 – 0.399 mm |
| 1 (this product) | 4 | 0.40 – 0.499 mm |
| 2 | 5 | 0.50 – 0.599 mm |
3.3. PREMISUT HS40 family (40 mm ½ circle needle, reverse cutting triangular point, black thread)
| Needle profile | USP size | Thread | Reference |
| HS 40 Prime | 1 | 1 × 75 cm (black) | P642HS4001 / AMP642HS4001 (this product) |
| HS 40 Prime | 1 | 1 × 100 cm (black) | P647HS4001 / AMP647HS4001 |
| HS 40 Prime | 0 | 1 × 75 cm (black) | P642HS4010 / AMP642HS4010 |
| HS 40 | 1 | 1 × 100 cm (black, standard range) | 647HS4001 / AM647HS4001 |
The same HS needle is available in other sizes (HS36, HS26, HS48, etc.) and in other thread length series for procedures with shorter or longer reach.
4. Equivalences and Cross-References
The reference P642HS4001 corresponds to the official nomenclature of the PREMISUT Prime range (HS40 profile: ½ circle needle with 40 mm reverse cutting triangular point, USP 1 gauge, 75 cm grooved black thread) and Cahnos markets it without altering characters: no digit should be added, removed or reordered when ordering or tendering.
Black non-absorbable nylon monofilament is one of the most standardised permanent suture families on the market (commercial type ‘Ethilon’). There is no manufacturer-validated equivalence table cross-referencing this reference with other brands for the exact combination of USP 1 + ½ circle reverse cutting needle of 40 mm + 75 cm black strand; the following comparisons are functional, at category level, and must be confirmed against the technical data sheet or instructions for use (IFU) of each manufacturer before use in tendering or direct substitution.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| B. Braun (Aesculap) | DAFILON® Blue – uncoated polyamide monofilament, USP 1 (EP 4), 90 cm, 40 mm HS40 needle (½ circle, reverse cutting) | C0932787 (verified in official catalogue B. Braun) | Comparable by category (not interchangeable 1:1) | They match in material (polyamide), USP 1 gauge (EP 4), ½ circle curvature, HS40 needle profile (40 mm, reverse cutting) and 1-needle configuration; they differ in thread colour (blue vs. black) and thread length (90 cm vs. 75 cm). Confirm IFU and CE/MDR classification before substituting. |
Other homologous families on the market that should be reviewed property by property (material, gauge, curvature, point family, needle and thread length, colour and CE/MDR regulatory classification) before any substitution: ETHILON® (Ethicon – J&J MedTech), MONOSOF™ / DERMALON™ (Medtronic/Covidien), DemeLON™ (DemeTECH), LUXAMID® (LuxSutures) and SUPRAMID® (SMI / B. Braun). No codes from these manufacturers have been verified in public sources for the exact HS40 40 mm, USP 1 and 75 cm black strand configuration; they must be confirmed with the manufacturer before use in substitution or tendering.
There is no manufacturer-validated equivalence table; consult the technical department before use in tendering. This needle geometry (HS40, USP 1, 75 cm) is only fully safely interchangeable within the PREMISUT system itself (P642HS4001 / AMP642HS4001).
5. Selection tips and variants of interest
- Tissue to suture: USP 1 (EP 4) gauge offers high strength for subcutaneous tissue, skin under moderate tension, fascia and deep planes. The reverse cutting needle passes through dense tissues without tearing the puncture bed. For finer planes, consider USP 0, USP 2/0 or USP 3/0 from the same family.
- Prime quality: if the procedure requires maximum sharpening precision and batch-to-batch consistency, the same geometry is available in the Prime range (this reference, P642HS4001 / AMP642HS4001) versus the standard range (e.g., 647HS4001 / AM647HS4001 in 100 cm thread).
- Needle length: the 40 mm HS40 profile covers medium-depth access. For smaller fields, the same nylon range offers the HS36 or HS26 profiles; for wide access or deep planes, the HS48 profile (Ref. P647HS4801, SKU AMP647HS4801).
- Thread length: the 75 cm covers continuous sutures or several stitches without re-threading; if you need more length, the same HS40 Prime configuration is available in 100 cm (P647HS4001 / AMP647HS4001).
- Presentation: box of 12 units in individual sterile sachets with batch traceability; intended for operating theatre stock and replenishment in ophthalmology and neurosurgery departments.
- Sterility verification: sterile product by ethylene oxide and single-use; check the integrity of the blister and the expiry date before each use and keep the reference P642HS4001 together with the batch number for traceability. Store below 30 °C, away from moisture and direct sunlight.





Reviews