Description
1. Commercial Introduction
The Fiocchetti MEDIKA 200 refrigerator with glass door is the professional positive-temperature solution for storing medicines and vaccines in pharmacies, hospital departments, clinical, aesthetic, dental and ophthalmological practices, laboratories and the pharmaceutical industry. Its tempered glass door allows the contents to be viewed without opening the chamber, reducing cold loss and facilitating visual stock control: a key advantage in dispensing points and vaccination rooms. Made in Italy, it complies with current international standards and can be customised to order on quotation.
2. Technical explanation of the equipment
The MEDIKA® series combines a clean, functional design with the flexibility of a customisable internal configuration using adjustable shelves and drawers. The door is of tempered glass according to IEC 61010-1, anti-fog and low-emissivity in an aluminium frame, with magnetic gasket for an airtight seal, automatic closing when the opening angle is less than 90° and reversible opening. The chamber incorporates professional ventilated refrigeration (uniform temperature distribution), adaptive automatic defrosting with condensate evaporation, full-height LED lighting and high-density injected polyurethane foam insulation (43 kg/m³), CFC-free and with ecological water injection (70 mm at the rear and 50 mm at the sides). The structure is made of steel with white antibacterial plasticised PVC, with an option for an interior chamber in 18/10 AISI 304 stainless steel, rounded corners and chamber back; it includes 4 unidirectional wheels and two front stability feet. It uses ecological R290 refrigerant gas with a consumption of only 0.6 kWh/24h and a noise level below 45 dBA.
The unit is available with three electronic controllers to adapt the level of traceability and monitoring to each facility: ECT-F (basic storage with three types of temperature alarms and 12-event logging), ECT-F PLUS (adds 48 h backup battery for power cuts, minimum/maximum temperature logging and clean contact for remote alarm) and ECT-F TOUCH (7″ touchscreen with real-time graph, SD card, USB port, SIM port, integrated Web Server LAN RJ45 and MODBUS TCP/IP, and integration with local and cloud monitoring systems). All offer 0.1 °C resolution and ±0.3 °C accuracy.
3. Technical Specifications
Structure and refrigeration
| Parameter | Specification |
|---|---|
| Temperature range | +2 °C / +15 °C |
| Nominal capacity | 200 litres (7.06 cu ft) |
| Net capacity | 221 litres (7.80 cu ft) |
| Internal dimensions | 50 x 48.5 x 90 cm |
| External dimensions (HxDxW) | 60 x 66 x 146 cm |
| Insulation type | High-density injected polyurethane foam (43 kg/m³), CFC-free and with ecological water injection |
| Insulation thickness | 70 mm rear / 50 mm on the sides |
| Structure | Steel with white antibacterial plasticised PVC (interior/exterior) or optional interior in 18/10 AISI 304 steel; rounded corners and chamber with back |
| Mobility | 4 unidirectional wheels + 2 front stability feet |
| Door | Tempered glass IEC 61010-1, anti-fog and low-emissivity in aluminium frame; magnetic gasket; automatic closing <90°; reversible opening |
| Internal lighting | Full LED along the entire height of the chamber |
| Cooling type | Professional ventilated |
| Defrosting type | Adaptive automatic with condensate evaporation |
Electronic controller
| Parameter | Specification |
|---|---|
| Available controllers | ECT-F / ECT-F PLUS / ECT-F TOUCH |
| Display | Backlit alphanumeric LCD display with 16 characters, customisable in 5 languages (ECT-F / ECT-F PLUS); 7″ touchscreen (ECT-F TOUCH) |
| Resolution / accuracy | 0.1 °C / ±0.3 °C |
| Keyboard | 4 membrane keys with password protection (on-off and setpoint) |
| Alarms (ECT-F) | Three types of temperature alarms and 12-event recording |
| ECT-F PLUS | Backup battery for power failure alarm (up to 48 h), min/max recording and clean contact for remote alarm |
| ECT-F TOUCH | SD card with history, USB port, SIM port (optional GSM module), integrated Web Server (LAN RJ45) and MODBUS TCP/IP; dry contact for remote alarm; silencable audible and visual alarms with automatic logging |
Technical data of the refrigerator
| Parameter | Value |
|---|---|
| Refrigerant gas type | R290 |
| Energy consumption | 0.6 kWh/24h |
| Absorbed power | kW 0.17 / Amp 1 |
| Starting absorption | Amp 3 |
| Noise level | <45 dBA |
| Heat emission EN 12900 | W 25 |
| Maximum number of shelves | 5 (dimensions cm 46×45) |
| Maximum number of drawers | 8 (with dividers: cm 11.8x40x7.5, 3 spaces; without dividers: cm 39x40x7.5) |
| Maximum number of stainless steel wire drawers | 8 (internal dimensions cm 43.5×40.5×6.2) |
Available references
| Model | Fiocchetti code | Controller | Temperature range | Capacity | External dimensions (HxDxW) |
|---|---|---|---|---|---|
| MEDIKA 200 ECT-F | MED0201 | ECT-F | +2 / +15°C | 221 lt (7.80 cu ft) | 60 x 66 x 146 cm |
| MEDIKA 200 ECT-F PLUS | MED0203 | ECT-F PLUS | +2 / +15°C | 221 lt (7.80 cu ft) | 60 x 66 x 146 cm |
| MEDIKA 200 TOUCH | MED0202 | ECT-F TOUCH | +2 / +15°C | 221 lt (7.80 cu ft) | 60 x 66 x 146 cm |
4. Equivalences and Cross-References
Within the manufacturer’s own range, the three versions on the datasheet (MED0201 ECT-F, MED0203 ECT-F PLUS and MED0202 TOUCH) share the cabinet, a net capacity of 221 litres, dimensions (60 x 66 x 146 cm) and temperature range (+2/+15 °C); they differ exclusively in the electronic controller and its recording, alarm and monitoring features. The LABOR 200 series (LAB0201/LAB0203/LAB0202) offers the same cabinet with a solid door, and the EMOTECA 200 series (DME0202) the same platform configured as a blood bank at +4 °C.
There is no equivalence table validated by the manufacturer against other brands; consult the technical department before use in tenders. At category level, in the segment of positive-temperature pharmacy and vaccine refrigerators with glass door and electronic control with alarms, the homologous families of Liebherr (MediLine range), Kirsch, Gram Commercial and Dometic Medical stand out. Before any substitution between brands, the following must be verified field by field: net useful capacity, thermal range, door type (glass vs. solid), climate class, refrigerant gas, installation dimensions, alarm and recording features, and certification as a medical device.
5. Selection tips and variants of interest
Choose the controller according to the required traceability level: the ECT-F covers basic storage with three types of temperature alarm and logging of 12 events; the ECT-F PLUS adds a 48 h backup battery for power cuts, minimum/maximum logging and a clean contact for remote alarm —key in pharmacies and vaccination centres without continuous supervision—; the ECT-F TOUCH is the hospital option for integration with local and cloud monitoring systems, real-time graphical logging and data export via SD/USB/LAN (MODBUS TCP/IP).
Check the installation space: the unit occupies 60 x 66 x 146 cm (width x depth x height). If you need a smaller volume with a glass door, consider the MEDIKA 140 series (MED0141/MED0143/MED0142); for greater capacity, the MEDIKA series is extended in larger volumes within the same family. If you prefer a solid door for maximum thermal and light protection, the LABOR 200 series (LAB0201/LAB0203/LAB0202) is the direct equivalent from the same manufacturer. You can complete the unit with additional shelves, pull-out drawers, stainless steel wire drawers, a lockable narcotics drawer, a wheel kit with brakes and monitoring accessories (GSM/Wi-Fi modules, DMLP TOUCH module, local and cloud monitoring systems), all available on request. Sale is made on quotation: the manufacturer configures each unit to order.