Description
1. Introduction
The biopsy needle Complete-Chiba® from Cahnos is the choice of interventional radiology and diagnostic imaging departments that need to obtain reliable cytological samples without leaving the MRI room. Designed and manufactured by the Italian specialist Sterylab S.r.l., it combines the most proven Chiba tip geometry in aspiration biopsy with a fully non-magnetic construction, making it safe for use in magnetic resonance imaging. A professional, sterile, single-use solution designed for hospitals, private clinics and diagnostic centres that demand maximum safety and diagnostic quality in every puncture.
2. Technical explanation of the equipment and procedure
The Chiba-type needle is the international standard for fine needle aspiration (PAAF/BAAF). Its long, sharp bevel allows it to pass through tissues with less insertion force and a more predictable trajectory, preserving the cellular integrity of the sample. In the Complete-Chiba® version, the atraumatic Quincke tip improves penetration and makes it less invasive, while the ultra-thin wall cannula facilitates easier and more abundant aspiration. The centimetric marking with sliding stop allows exact depth control during the procedure. Its construction with non-magnetic materials ensures that the device does not generate artefacts in the MRI image or suffer traction forces within the magnetic field, making it a safe tool for MRI-guided punctures. In addition, the transparent handle incorporates a double magnifying lens and an adjustment tooth for the introducer needle, and the colour code instantly identifies the gauge. Sterile single-use presentation, which minimises the risk of nosocomial infections.
3. Technical specifications and tables
Made from non-magnetic medical-grade steel, the Complete-Chiba® is supplied sterile and disposable, individually packaged and presented in boxes of 30 units.
Summary technical data sheet
| Parameter | Specification |
| Reference Cahnos | CAHCHIBA |
| Needle type | Disposable Chiba needle for cytological biopsy |
| Tip | Atraumatic Quincke type |
| Cannula | Ultra-thin wall |
| Material | Non-magnetic medical-grade steel |
| MRI safety | Non-magnetic, safe for MRI |
| Marking | Centimetric with sliding stop |
| Identification | Gauge colour code |
| Handle | Transparent with double magnifying lens and adjustment tooth for introducer needle |
| Sterility | Sterile, disposable, single-use |
| Packaging | Box of 30 units |
| Manufacturer | Sterylab S.r.l. (Rho, Milan, Italy) |
Available models in the range
| Gauge (G / mm) | Length (cm) | Colour code |
| 18G (1.20 mm) | 7 – 9 – 11 – 15 – 20 – 25 – 35 | Pink |
| 20G (0.90 mm) | 7 – 9 – 11 – 15 – 20 – 25 – 35 | Yellow |
| 21G (0.80 mm) | 7 – 9 – 11 – 15 – 20 – 25 | Green |
| 22G (0.70 mm) | 7 – 9 – 11 – 15 – 20 – 25 | Black |
| 23G (0.65 mm) | 7 – 9 – 11 – 15 – 20 | Blue |
| 25G (0.50 mm) | 5 – 7 – 9 – 11 | Orange |
The lengths indicated in the manufacturer’s documentation are also available in a specific version for nuclear magnetic resonance, adding the suffix «MR» to the order code.
Order code composition
| Range | Example code |
| Complete-Chiba® standard (20G – 15 cm) | CCH2015 |
| Complete-Chiba® MRI version (20G – 15 cm) | CCH2015 MR |
4. Equivalences and Cross-References
The Complete-Chiba® belongs to Sterylab’s range of biopsy and localisation needles, a leading European manufacturer in this segment. Its Chiba tip geometry makes it clinically comparable to the standard Chiba cytological biopsy needles on the European market, with the key difference of its non-magnetic construction and atraumatic Quincke tip. Within the Cahnos catalogue, for MRI-guided biopsies it also offers the BIO-CUT® needle (ref. CAHBIOCUT, non-magnetic TruCut type) and the semi-automatic SPRING-CUT® needle (ref. CAHSPRING-CUT). For cytological punctures outside MRI, the disposable Chiba family from Cahnos includes references BIOCHIBACPL001 to BIOCHIBACPL006 in gauges 18G to 25G, the echo-reflective series BIOCHIBAECHO and the echogenic CHGA-type needles. If the protocol also requires a histological sample, the modified Menghini cutting series with total recovery (FULLOPTY) can be considered. Please consult our technical-commercial department to confirm compatibility with your usual supply brand.
5. Selection tips and variants of interest
Select the gauge according to the target organ and the required sample volume: 18G and 20G gauges provide more abundant cellular samples in deep tissues; 21G to 23G gauges offer an optimal balance between sample and comfort; 25G is the most delicate option for fine accesses and sensitive areas. The length should be chosen according to the depth of the lesion and the approach route: short lengths (7-11 cm) for superficial lesions and long lengths (20-35 cm) for deep accesses or patients with greater adipose panniculus. For MRI-guided punctures, verify that the requested reference includes the suffix «MR» in the manufacturer’s order code (e.g., CCH2015 MR). Contact our technical-commercial team to confirm availability, quotation conditions, volume orders and adaptation to public tender specifications.