Description
1. Complete and sterile kit for percutaneous airway access
The percutaneous dilatation kit Ref. TP02-66 is the comprehensive single-use solution for establishing a safe airway quickly, minimally invasively and in a controlled manner in adult patients. It incorporates a tracheostomy tube of 9 mm internal diameter with fixed cervical flange and dedicated subglottic suction lumen, a feature especially valued in intensive care units for its contribution to the prevention of ventilator-associated pneumonia (VAP), together with Griggs-type dilatation forceps, the hallmark of this technical variant.
All the necessary material is presented in a single sterile kit, ready to use: one order, one code and the assurance that the team has all the instruments with full traceability. Preparation time is reduced, the risk of forgetting material is minimised and standardisation of the procedure is facilitated in ICU, emergency, anaesthesia and airway units.
2. Technical explanation of the equipment and procedure
Percutaneous dilatational tracheostomy (PDT) using the Griggs technique is an evolution of the Ciaglia technique: it is performed at the bedside, usually under bronchoscopic control, without the need to transfer the critically ill patient to the operating theatre. The procedure begins with tracheal puncture using the 16 G introducer needle, continues with the passage of the J-tip guidewire (atraumatic tip that minimises tissue damage) and ends with progressive dilatation of the stoma using Howard-Kelly type dilatation forceps, which advance over the guide opening the orifice, followed by advancement of the pre-mounted cannula over the introducer with locking system and conical tip, which avoids steps and maintains the position of the tube during insertion.
The cannula incorporates an additional subglottic suction lumen with outlet above the balloon (cuff), which allows aspiration of secretions accumulated in the subglottic space and reduces the risk of contaminated micro-aspirations into the lower airway. The fixed cervical flange, flexible and in transparent plastic, facilitates stable fixation and correct visualisation of the stoma; the balloon is low-pressure, the connector is universal 15 mm (15M), compatible with the ventilator circuit, and the tube incorporates a radiopaque line for radiological control. All components are single-use, latex-free and phthalate-free.
3. Technical specifications and kit contents
Kit contents (Ref. TP02-66)
| Component | Specification |
|---|---|
| Tracheostomy tube | 9 mm tube with fixed flange and subglottic suction lumen |
| Dilatation forceps | Griggs type (Howard-Kelly), dilatation over guidewire |
| Tracheostomy tube introducer | Locking system and conical tip for safe and progressive insertion |
| Dilator | 14 Fr for progressive dilatation of the stoma |
| Guidewire | J-tip 0.05\”, NiTiNol core resistant to kinking |
| Guide catheter | For guiding the dilator and the tube |
| Scalpel | Safety scalpel No. 15 with retractable protection and fixed blade |
| Syringe | 10 ml |
| Introducer needle | 16 G |
| Fixation band | Skin colour with Velcro-type closure |
| Instrument table cover | Sterile field cover |
| Tracheal compress | 9 × 9 cm, three-layer foam |
| Gauze swabs | 10 × 10 cm |
| Sterile lubricating gel | Individual sachet |
Technical features of the tube and kit
| Parameter | Value |
|---|---|
| Technique | Percutaneous dilatational tracheostomy, Griggs technique with forceps |
| Internal tube diameter | 9.0 mm; exact outer diameter and length according to the tube’s technical data sheet |
| Flange type | Fixed, cervical, flexible, transparent plastic |
| Subglottic suction | Yes, dedicated lumen with outlet above the balloon |
| Balloon (cuff) | Low pressure |
| Connector | Universal 15 mm (15M) |
| Radiological control | Radiopaque line |
| Use | Single-use (disposable) |
| Sterilisation | Ethylene oxide (ETO), process validated according to EN 11135-1; desorption in accordance with ISO 10993-7 |
| Presentation | Sterile, individually packaged in TYVEK bag, in accordance with EN ISO 11607-1/2 |
| Materials | Latex-free and phthalate-free in all components |
| Regulatory classification | CE medical device class IIa (Annex IX, Directive 93/42/EEC) |
| Indication | Percutaneous tracheostomy, cricothyrotomy, coniotomy or laryngotomy |
Family of Griggs kits with fixed flange at Cahnos
| Reference | Diameter | Subglottic suction | Forceps |
|---|---|---|---|
| TP02-64 | 7 mm | Yes | Yes |
| TP02-65 | 8 mm | Yes | Yes |
| TP02-66 (this product) | 9 mm | Yes | Yes |
| TP02-67 | 7 mm | No | Yes |
| TP02-68 | 8 mm | No | Yes |
| TP02-69 | 9 mm | No | Yes |
| TP02-74 | 7 mm | Yes | No |
| TP02-75 | 8 mm | Yes | No |
| TP02-76 | 9 mm | Yes | No |
| TP02-77 | 7 mm | No | No |
| TP02-78 | 8 mm | No | No |
| TP02-79 | 9 mm | No | No |
4. Equivalences and Cross-References
The Griggs technique with dilatation forceps over a guidewire is a widely established standard in percutaneous tracheostomy. The best-known references are listed below, classified according to the level of verification achieved; none should be treated as an exact 1:1 equivalence.
- Comparable at category level: Portex percutaneous kits (ICU Medical / Smiths Medical) from the GRIGGS family with Blue Line Ultra tube with suction lumen, such as the 100/893 series (reference 100/893/070 corresponds to the 7.0 mm size) and the 100/891 series with guidewire dilatation forceps. Same functional family (Griggs technique, tube with subglottic suction and introducer), but different manufacturer, materials and component geometry: not interchangeable 1:1. The equivalent code for the 9.0 mm size is pending confirmation with the manufacturer.
- Comparable at category level: TRACOE experc kit (REF 520, TRACOE / Atos Medical) completed with P cannula with subglottic suction device. Same clinical indication and suction function, but different component configuration and dilatation technique: not interchangeable 1:1.
- Comparable at category level: Griggs technique kits from DEAS/ACCEX (Deaflux line) with fixed-flange tracheostomy tube or for subglottic suction. Same procedural approach, different presentation and component geometry: not interchangeable 1:1.
- Within the Cahnos catalogue: references TP02-64 (7 mm) and TP02-65 (8 mm) share the exact configuration (fixed flange, subglottic suction and forceps) and differ only in the cannula diameter: direct range equivalents, not interchangeable 1:1 depending on the patient.
There is no equivalence table validated by the manufacturer that allows this reference to be cross-referenced with specific codes from other suppliers; it is recommended to consult the technical department of Cahnos before using this data sheet in a tendering process or stock replacement.
5. Selection tips and variants of interest
The cannula diameter should be chosen according to the patient’s anatomy and ventilation requirements: 7 mm (TP02-64) for adults of smaller build or more limited anatomies, 8 mm (TP02-65) as the usual standard size and 9 mm (this kit, TP02-66) for patients with a wide airway or need for greater ventilatory flow.
If the patient is likely to require prolonged mechanical ventilation or presents with copious secretions, the subglottic suction lumen of this series is particularly recommended for its role in preventing ventilator-associated pneumonia (NAVM). If the service prefers the dilation sequence with a single dilator without forceps, versions TP02-74, TP02-75 and TP02-76 are available (fixed flange with suction); and if a tube without a suction lumen is preferred, references TP02-67, TP02-68 and TP02-69 are available. For variable cervical anatomies, consult the Griggs series with adjustable flange (for example, TP02-09 in 9 mm size). All kits in the family are single-use: check the integrity of the sterile packaging and the expiry date before use, and dispose of them after the procedure as clinical waste in accordance with current local regulations.