Description
1. Continuous, safe infusion without electronic equipment
The ELASPUMP® 100 ml Elastomeric Infusion Pump with UV protection (Ref. 093002/4/B) is the ideal solution for the constant and controlled administration of medications in hospitals, clinics, home hospitalisation units and outpatient care centres. It is a single-use, sterile device free from DEHP, PVC and latex, which allows drugs to be infused at a calibrated flow rate of 2 ml/hour without relying on electronic pumps, batteries or maintenance, with a very competitive cost per treatment and no risk of cross-contamination between patients. Its UV protection design makes it especially recommended for photosensitive medications, common in oncology treatments, and its ease of use makes it a practical option for both healthcare professionals and patients themselves.
2. Technical description and operating principle
The ELASPUMP® pump operates through the elastic recovery pressure of a silicone elastomer contained within a hermetic, rigid and transparent housing, which prevents accidental leaks and allows visual inspection of the drug level from the outside without having to open the device. This pressure drives the medication through the tubular system at a constant flow rate calibrated at the factory of 2 ml/hour, resulting in an approximate autonomy of 2 days + 2 hours (approx. 50 hours) with the full 100 ml volume. The integrated UV protection preserves the stability of the drug throughout the infusion period.
- Silicone elastomer encapsulated in a hermetic, rigid and transparent housing with UV protection, leak-free and with visual level inspection.
- One-way filling valve with closed circuit and flat bottom, compatible with a standard syringe.
- Hydrophobic filter on the male luer-lock connector cap: allows priming of the line without removing the cap, preserving internal sterility until actual use.
- In-line particle filter, which retains impurities and guarantees the purity of the administered drug.
- Removable sliding clamp, easy to remove when necessary.
- Colour-coded label with flow rate printing for quick and immediate identification of the flow.
- Patient label, designed for easy personalisation of the treatment.
- Full traceability: variable data and product code printed on the external housing and on the pump body.
3. Technical Specifications
General device characteristics
| Characteristic | Specification |
|---|---|
| Brand | ELASPUMP® |
| Reference | 093002/4/B |
| Capacity | 100 ml |
| Flow type | Fixed continuous (factory-calibrated) |
| Infusion flow rate | 2 ml/hora |
| Approximate duration | 2 days + 2 hours (approx. 50 hours at full volume) |
| Drive mechanism | Silicone elastomer in rigid and transparent casing |
| Filling system | Closed-circuit one-way valve, compatible with standard syringe |
| Connection | Male Luer-lock with cap and hydrophobic filter |
| Filtration | Hydrophobic filter + in-line particle filter |
| Protection | UV protection for photosensitive drugs |
| Routes of administration | Intravenous, arterial, epidural and subcutaneous |
| Clinical indications | Chemotherapy, pain therapy, antiviral and antibiotic treatments |
| Power supply | Does not require battery or electrical power source |
| Use | Single-use, sterile |
| Composition | DEHP-free, PVC-free, latex-free |
Indicative infusion duration by flow rate (100 ml volume)
| Flow rate | Approximate duration |
|---|---|
| 1.5 ml/hora | 2 days + 19 hours (approx. 67 hours) |
| 2 ml/hour (current product) | 2 days + 2 hours (approx. 50 hours) |
| 2.5 ml/hora | 1 day + 16 hours (approx. 40 hours) |
| 4 ml/hora | 1 day + 1 hour (approx. 25 hours) |
| 5 ml/hora | 0 days + 20 hours (approx. 20 hours) |
| 8 ml/hora | 0 days + 12 hours |
| 10 ml/hora | 0 days + 10 hours |
Indicative values calculated at full volume; actual duration may vary slightly depending on usage conditions and drug characteristics.
ELASPUMP® continuous flow range with UV protection available at Cahnos
| Reference | Capacity | Flow rate | Approximate duration |
|---|---|---|---|
| 093002/3/B | 100 ml | 1.5 ml/hora | 2 days + 19 hours |
| 093002/4/B (current product) | 100 ml | 2 ml/hora | 2 days + 2 hours |
| 093002/41/B | 100 ml | 2.5 ml/hora | 1 day + 16 hours |
| 093002/5/B | 100 ml | 4 ml/hora | 1 day + 1 hour |
| 093002/6/B | 100 ml | 5 ml/hora | 0 days + 20 hours |
| 093002/7/B | 100 ml | 8 ml/hora | 0 days + 12 hours |
| 093002/8/B | 100 ml | 10 ml/hora | 0 days + 10 hours |
| 093004/4/B | 200 ml | 2 ml/hora | 4 days + 4 hours |
| 093004/41/B | 200 ml | 2.5 ml/hora | 3 days + 8 hours |
| 093004/5/B | 200 ml | 4 ml/hora | 2 days + 2 hours |
| 093004/6/B | 200 ml | 5 ml/hora | 1 day + 16 hours |
| 093004/7/B | 200 ml | 8 ml/hora | 1 day + 1 hour |
| 093004/8/B | 200 ml | 10 ml/hora | 0 days + 20 hours |
| 093005/8/B | 275 ml | 10 ml/hora | 1 day + 3 hours |
4. Equivalences and Cross-References
The ELASPUMP® 100 ml with UV protection (Ref. 093002/4/B) belongs to the range of continuous flow elastomeric pumps of the ELASPUMP® family and is complementary to the following references available at Cahnos:
| Reference | Description |
|---|---|
| 093002/3/B | ELASPUMP® 100 ml with UV protection, continuous flow of 1.5 ml/hour (2 days + 19 hours), same capacity with lower flow rate |
| 093002/41/B | ELASPUMP® 100 ml with UV protection, continuous flow of 2.5 ml/hour (1 day + 16 hours), same capacity with slightly higher flow rate |
| 093002/5/B | ELASPUMP® 100 ml with UV protection, continuous flow of 4 ml/hour (1 day + 1 hour), same capacity with higher flow rate |
| 093004/4/B | ELASPUMP® 200 ml with UV protection, continuous flow of 2 ml/hour (4 days + 4 hours), same flow rate with larger volume |
| 093001/4 | ELASPUMP® 60 ml continuous flow (2 ml/hour = 1 day + 6 hours), for smaller volumes |
| 093002/M1 · M2 · M3 | ELASPUMP® 100 ml variable flow (4 and 8 speeds), DEHP-free, for protocols with adjustable flow rate |
Due to its 100 ml capacity, its pressurised elastomer system with luer-lock connector, hydrophobic filter, fixed calibrated flow rate and UV protection, this device is comparable to disposable continuous flow elastomeric pumps from other manufacturers used in ambulatory infusion, controlled analgesia and home intravenous therapy, being interchangeable in most clinical protocols for continuous infusion and patient-controlled analgesia (PCA). Consult our technical-commercial team to confirm exact compatibility with your usual reference supplier.
5. Selection tips and variants of interest
- Volume to infuse: the 100 ml version is ideal for medium-duration therapies of approximately 50 hours (2 days + 2 hours) at 2 ml/hour. For smaller volumes, the 60 ml version (Ref. 093001/4) is available, and for longer therapies, the 200 ml and 275 ml versions with UV protection.
- Required flow rate: if your protocol is fixed, this continuous flow version at 2 ml/hour offers the best cost-benefit ratio. If you need to adjust the speed during treatment, consider the variable flow versions of 100 ml (Refs. 093002/M1, M2 and M3).
- Photosensitive drugs: for light-sensitive medications, always choose the ELASPUMP® versions with UV protection (Refs. 093002/3/B, 093002/4/B, 093002/41/B, 093002/5/B, 093002/6/B, 093002/7/B, 093002/8/B, 093004/4/B, 093004/41/B, 093004/5/B, 093004/6/B, 093004/7/B, 093004/8/B and 093005/8/B).
- Recommended accessories: combine the pump with 1-way or 3-way infusion extension lines with stopcock to adapt the system to the patient’s needs.
- Purchase conditions: contact our technical-commercial department to request samples, technical data sheets or special conditions for tenders and hospital purchases.




Reviews