Description
1. Commercial Introduction
The neonatal radiant warmer TECH-MED R10MB with phototherapy (Ref. TMD2912) is a transportable infrared radiator from the MB series, designed to provide controlled radiant heat to the newborn in delivery rooms, neonatology units and NICUs. Its design on a wheeled trolley allows it to be easily moved to the cot or incubator, functioning as a substitute for the open incubator. It also integrates an LED phototherapy lamp for bilirubin that allows treating neonatal jaundice at the same point of care, without the need for additional equipment. It is the technical solution for services that need mobility, versatility and safety in the care of the smallest patient.
2. Technical explanation of the equipment
The radiant warmer works as a heat source by infrared radiation: the 700 W quartz lamp emits heat directed onto the neonate, providing direct and controllable thermal input without contact. The MB series incorporates a control system with automatic air temperature regulation, range 34 to 38 °C with precision of ±0.2 °C and digital measurement with resolution of 0.1 °C, so that the equipment adjusts the power autonomously. It also allows manual adjustment of the heating power from 0 to 700 W in increments of 5% of the maximum range.
Safety is a cornerstone of the equipment: it has visual and acoustic alarms for low temperature, over-temperature (alarm threshold ±0.8 °C) and sensor failure (ERR). It incorporates a clock for the APGAR score, with audible and visual signal at 1, 3, 5 and 10 minutes, and a 20 W LED lighting system. The LED phototherapy lamp for bilirubin emits blue light with a maximum of 465 nm, maximum irradiance of 60 µW/cm²/nm and a maximum irradiation area of 55×40 cm at a distance of 90 cm, with smooth adjustment of irradiation, silent operation (0 dB) and a service life of the phototherapy elements exceeding 50,000 hours. Classified as a class II b medical device (MDD), class I protection, type BF.
3. Technical Specifications
| Reference (SKU) | TMD2912 |
| Brand | TECH-MED (ZTM TECH-MED Sp. z o.o.) |
| Model / family | R10MBF – MB series (IR-10 MB radiant heater with phototherapy) |
| Set type | Transportable neonatal radiant warmer (infrared radiator) |
| Temperature control | Automatic air, range 34–38 °C with precision ±0.2 °C |
| Temperature measurement | Digital, resolution of 0.1 °C |
| Heating power | Manual adjustment 0–700 W in increments of 5% of the maximum range |
| Infrared lamp | 700 W quartz lamp |
| Lighting | 20 W LED |
| Alarms | Low temperature and over-temperature (visual and acoustic), threshold ±0.8 °C; sensor failure ERR |
| APGAR clock | Audible and visual signal at 1, 3, 5 and 10 min |
| Phototherapy | LED phototherapy lamp for bilirubin, blue light max. 465 nm |
| Irradiation | Max. 60 µW/cm²/nm; max. irradiation area 55×40 cm at 90 cm distance; smooth irradiation adjustment |
| Noise / phototherapy service life | 0 dB / > 50,000 hours |
| Protection class | Class I, type BF |
| Article classification | MDD, class II b |
Dimensions
| Width | 700 mm |
| Height | 1900 mm |
| Weight | Approx. 35 kg |
| Wheel diameter | 100 mm |
| Bassinet dimensions | 720 mm × 470 mm |
Available references
| Reference | Description |
| TMD2912 | Neonatal radiant warmer R10MB with LED phototherapy (model R10MBF) |
4. Equivalences and Cross-References
The manufacturer TECH-MED does not publish a validated equivalence table that allows this reference to be substituted by the specific code of another manufacturer on a 1:1 interchangeable basis. Any comparison between brands must be understood as comparable by category and validated field by field before use in tendering or stock substitution.
| Manufacturer / Brand | Family / Model | Reference | Equivalence level | Observations |
| TECH-MED | Radiant warmer R10MB/W (MB series without phototherapy) | R10MB/W | Same family (manufacturer variant) | It incorporates 100 W fluorescent lighting instead of the 20 W LED phototherapy lamp. |
| TECH-MED | Neonatal resuscitation station | SRN-10 | Same neonatal range from the manufacturer | Higher-specification resuscitation equipment (92 kg) with 700 W quartz heater, table and accessories; not interchangeable 1:1. |
| GIMA | Infant LED phototherapy lamp on trolley | GM31192 | Comparable by category (neonatal phototherapy) | Independent phototherapy source on trolley; does not provide radiant heat. |
| Dixion | BabyGuard radiant warmers with integrated phototherapy | W-1140 / W-1145 | Comparable by category (not interchangeable 1:1) | Radiant heat systems with integrated LED phototherapy; different configurations and specifications. |
5. Selection tips and variants of interest
Select this R10MB radiant warmer with phototherapy when your unit needs a mobile radiant heat source that can be brought to the cot as an alternative to the open incubator, combining in a single device the controlled thermal input and the treatment of neonatal jaundice with blue light (465 nm). The 100 mm wheel and the weight of approximately 35 kg facilitate movement between stations in the room.
If you only require radiant heat without phototherapy treatment, you may consider the manufacturer’s R10MB/W variant from the same MB series, equipped with 100 W fluorescent lighting; for a more complete resuscitation setup (table, drawer unit and accessories), consider the SRN-10 resuscitation station from TECH-MED. Always verify the exact model reference, dimensions (700 × 1900 mm), the available height in the room and the electrical compatibility of the equipment before placing the order. For quotations, supply lead times and tender requirements, contact the technical-commercial department at Cahnos.