Description
1. Introduction
The Chlamydia Rapid Test cassette (ref. ICH-502) is an in vitro diagnostic device designed for the qualitative detection of the Chlamydia trachomatis antigen in cervical swab (women), urethral swab (men) and urine (men) samples. Manufactured by ALLTEST Biotech and marketed within its professional range, it provides a reliable result in just 10 minutes, with direct visual reading and no laboratory equipment required. It is the ideal solution for STI clinics, health centres, gynaecology and urology services, clinical laboratories, emergency units and sexually transmitted infection screening programmes.
2. Technical explanation and procedure
The test is a lateral flow chromatographic immunoassay based on the double antibody sandwich technique. A specific antibody against the Chlamydia trachomatis antigen is immobilised on the cassette membrane; during the assay, the extracted sample solution reacts with the conjugated antibody and, if the antigen is present, generates a visible coloured line in the test zone (T). The appearance of the control line (C) validates that the correct sample volume has been added and that migration has occurred correctly.
The procedure is simple: the antigen is extracted with the included reactive buffer solutions 1 (0.2M NaOH) and 2 (0.2M HCl), 3 drops of the extracted solution (approx. 100 µl) are placed in the sample well of the cassette and migration is awaited. Reading is performed at 10 minutes and must not be interpreted after 20 minutes. No additional equipment is required, which facilitates its use at the point of care.
3. Technical Specifications
3.1. General characteristics
| Parameter | Value |
|---|---|
| Product reference | ICH-502 |
| Manufacturer | ALLTEST Biotech (Hangzhou ALLTEST Biotech Co., Ltd.) – JusChek® range |
| Method | Lateral flow chromatographic immunoassay (double antibody sandwich technique) |
| Analyte detected | Chlamydia trachomatis antigen |
| Sample type | Cervical swab (woman), urethral swab (man) and urine (man) |
| Format | Visual reading cassette |
| Time to result | 10 minutes (do not interpret after 20 minutes) |
| Certification | CE0123 – Directive 98/79/EC (Annex II, other device); evaluation route according to Annex III; CE-IVDR 2017/746 Class C (certificate available on request) |
| Storage | 2–30 °C, in sealed packaging; do not freeze |
| Shelf life | 24 months (2 years) |
3.2. Clinical data by sampling type
| Sample type | Sensitivity | Specificity | Accuracy |
|---|---|---|---|
| Cervical sampling in women | 93.3 % | 97.5 % | 96.6 % |
| Urethral sampling in men | 86.2 % | 95.8 % | 92.7 % |
| Urine samples in men | 94.6 % | >99.9 % | 97.9 % |
3.3. Box contents
| Component | Quantity |
|---|---|
| Test cassettes | 20 |
| Dropper bottles with cap | 20 |
| Reactive buffer solution 1 (0.2M NaOH) | 1 |
| Reactive buffer solution 2 (0.2M HCl) | 1 |
| Rack (workstation) | 1 |
| Sterile swabs for cervical collection in women | 20 |
| Instructions for use | 1 |
3.4. Packaging
| Detail | Value |
|---|---|
| Sales presentation | 20 tests per box |
| Boxes per carton | 36 boxes per carton (720 tests per carton) |
| Carton dimensions | 55 × 39.5 × 46 cm |
| Weight per carton | 13 kg |
4. Equivalences and Cross-References
This test belongs to ALLTEST Biotech’s family of rapid antigen tests for STIs, all in lateral flow cassette format and professional presentation. Within the Cahnos catalogue, the Chlamytop Ag (ref. V07-1030003), also in a box of 20 cassettes for rapid detection of Chlamydia antigen, can be considered the direct equivalent of this test. It is complemented by the rest of the same manufacturer’s STI screening line: the Gonorrhoea Rapid Detection Test (ref. IGO-502), the HIV 1/2 Rapid Test Kit (ref. IHI-402) and the JusChek® Syphilis Rapid Test IgG/IgM (ref. ISY-N402), allowing a complete diagnostic battery for sexually transmitted diseases to be configured at a single point of care.
5. Selection tips and variants of interest
- Professional use: is indicated exclusively for healthcare facilities, clinical staff and laboratories; it is not a self-test for home use.
- Sample type: accepts cervical swab (female), urethral swab (male) and urine (male). For urine, the first morning void is recommended and prior processing with a centrifuge and sedimentation tube (not included).
- Sampling quality: antigen detection depends on vigorous cell collection; avoid excess blood on the swab, which may cause false positives.
- Quality control: check the control line (C) in each test; if it does not appear, the result is invalid and the test must be repeated.
- Storage: keep the cassettes in their original sealed packaging, between 2 and 30 °C, protected from humidity and direct sunlight; do not freeze.
- Usage volume: the box of 20 tests is the standard presentation; for high-volume screening programmes, the full carton (720 tests) offers the best cost per test ratio. Combine this reference with IGO-502, IHI-402 and ISY-N402 for a complete STI panel.





Reviews
There are no reviews yet.