We use our own and third-party cookies to analyse the use of the website and show you advertising related to your preferences, based on a profile built from your browsing habits (for example, pages visited). More information

Cahnos.com Compare invoice

Reusable electrosurgical pencil (handle) with push-buttons – Valleylab-type connector, 320 cm cable and 2.38 mm (3/32″) shank, with F4060 blade – Ref. F4141 (30 sterilisation cycles, box of 5 units)

20,00 €

Reusable electrosurgical pencil (handle) with dual push-buttons (cut and coagulation) for monopolar electrosurgery, with universal tripolar Valleylab-type connector and highly flexible 320 cm silicone cable. Includes the reusable F4060 blade with 2.38 mm (3/32″) shank, the industry standard, and is guaranteed for 30 sterilisation cycles in autoclave or ethylene oxide. Ergonomic high-resistance plastic body, latex-free and sealed. Supplied in a box of 5 units (non-sterile, individual packaging). Compatible with most electrosurgical generators with Valleylab socket (Covidien/Medtronic and similar). Reference F4141.

International shipping. Import duties and taxes are payable by the recipient.

Datasheet (opens in a new tab)

Reference
0565099

Description

1. Reusable electrosurgical pencil handle for the operating theatre

We place at your disposal the reusable electrosurgical unit handle (pencil) Ref. F4141, a professional reference from the European line of high-frequency monopolar electrosurgery. Designed for continuous use in the operating theatre, major ambulatory surgery, minor surgery consultations and general surgery services, it combines an ergonomic body of high-resistance plastic (latex-free), dual cut and coagulation button, universal Valleylab-type connector and 320 cm cable.

It is supplied in a box of 5 handles together with the reusable blade electrode F4060, ready to be integrated into the service’s reusable instrument inventory. Its warranty of 30 sterilisations offers a very competitive cost per procedure compared with disposable pencils, with consistent cut and coagulation performance in each cycle.

2. Technical explanation of the equipment and procedure

In a monopolar electrosurgery system, the electrosurgical generator (ESU) sends high-frequency current through the pencil to the active electrode —the F4060 blade mounted on the handle—, which concentrates the energy at the point of contact with the tissue; the circuit is closed through the neutral (dispersive) plate placed on the patient. Depending on the waveform selected on the generator, the effect is cut, coagulation or mixed.

This handle incorporates manual control with dual button (cut and coagulation) whose activation click allows immediate recognition of activation. Its watertightness is guaranteed by a silicone O-ring that prevents liquid infiltration and protects the operator against electric shocks, and the electrode fixation system prevents its rotation during use. The universal tripolar Valleylab-type connector allows its use with the vast majority of electrosurgical generators on the market without the need for adapters, and the shaft diameter of 2.38 mm (3/32\”) is the industry standard for electrode holders, which facilitates replacement with other tips from the reusable family.

3. Technical Specifications

Characteristic Specification
Reference F4141 (SKU Cahnos: 0565099)
Product type Reusable electrosurgical pencil (handle) with push buttons
System Monopolar, cut and coagulation
Activation Dual manual button (cut / coagulation)
Cable length 320 cm
Electrode shaft diameter 2.38 mm (3/32″) – universal fit
Connector Universal tripolar, Valleylab type
Included electrode Reusable blade F4060 (shaft 2.38 mm and 69 mm long)
Body material Sterilisable high-resistance plastic, latex-free; highly flexible silicone cable
Number of sterilisations 30 guaranteed cycles (handle)
Presentation Box of 5 pieces, individually packed in heat-sealed polyethylene bag – non-sterile
Electrical safety EN 60601-1 · EN 60601-2-2 · EN 60601-1-2
Regulatory classification Medical device class IIb (Directive 93/42/EEC and amendments)

Technical note: the value of 2.38 mm corresponds to the diameter of the electrode shaft (3/32\”), according to the manufacturer’s official technical data sheet; the handle housing is universal for reusable electrodes of this size.

Parameter Specification
Handle sterilisation Ethylene oxide or autoclave: up to 30 cycles
Autoclave – pre-vacuum 134 °C / 2.05 bar, 12 min exposure (+ pre-vacuum and drying)
Autoclave – alternative 121 °C / 1.05 bar, 20 min exposure
F4060 electrode sterilisation Up to 20 cycles
Storage 0 to +50 °C · relative humidity 20–80 %

4. Equivalences and Cross-References

No official equivalence table validated by the manufacturer cross-referencing F4141 with codes from other brands has been found in public sources. The following row is a functional comparison verified against the cited manufacturer’s official catalogue and should not be treated as 1:1 interchangeable without validation from the technical department.

Manufacturer / Brand Family / Model Reference Code / SKU Equivalence Level Technical Differences / Remarks
Gima (Italy) ORION PLUS – autoclavable handle with fixing screw (screw clamp), 3 m silicone cable and blade electrode 30539 Comparable by category (not interchangeable 1:1) Reusable dual-button handle with Valleylab-type connection, autoclavable up to 100 cycles (electrode up to 20) and compatible with Ø 2.4 mm shaft electrodes. It differs in the number of guaranteed sterilisations (100 vs 30), cable length (3 m vs 320 cm) and electrode fixation system (screw vs hexagonal locking). Code and attributes verified in the official catalogue of Gima and authorised distribution.

Other homologous families in the sector that must be evaluated with technical data sheet or IFU in hand before any substitution: reusable button pencils from Medtronic/Valleylab (Covidien), autoclavable handles from Aspen Surgical (Bovie) and reusable pencils with Valleylab-type connection of the same category present in the European market. In all cases, each field must be checked: material, shaft diameter (2.38 mm / 3/32\”), electrode fixation system, connector type (Valleylab), number of authorised sterilisation cycles and regulatory classification of the actual product. There is no equivalence table validated by the manufacturer; consult the technical department of Cahnos before use in tenders.

5. Selection tips and variants of interest

Depending on cable length: the F4141 (320 cm) is the balanced option for standard operating theatres. If your team works in fields where the generator is far from the table, consider the variant of the same family with 500 cm cable (ref. F4151, 30 sterilisations). If your priority is to extend the useful life per handle, reference F4243 (320 cm) offers the same technical configuration with 100 guaranteed sterilisations; the family is completed with 300 cm versions (F4242/F4244) and 500 cm versions (F4252/F4254).

Generator compatibility: the tripolar Valleylab-type connector of this pencil is compatible with most electrosurgical generators on the market (Covidien/Medtronic and similar). For equipment with 4 mm, 5 mm sockets or other configurations, consult the reusable adapters in the range (F4831–F4833) or the version with 5 mm Erbe connector of the same family (F4141/ERB).

Electrode spares: the reusable family offers F4060 blades (shank 2.38 mm, 69 mm) and, for 4 mm electrode holders, the F4060/4 with F4246/F4256 handles. Always check the shank diameter against the handle housing. The same family includes ball tips (F4068) and needle tips (F4084) for contact coagulation and fine dissection.

Operating theatre maintenance: use the sterile Tip Cleaner F7520 tip cleaner (box of 100 units) to remove eschar from the blade during the procedure and extend the electrode’s service life, and the sterile F4545 sheath to store the handle between uses. Record the number of sterilisation cycles (handle up to 30, electrode up to 20) and discard the handle at any sign of deterioration of the body or cable. For institutional volumes, hospital supply agreements or validation of equivalents from other manufacturers, contact the technical-commercial team at Cahnos.

Additional information

Product type

Reusable electrosurgical pencil (handle) with push buttons

System

Monopolar, cut and coagulation

Activation

Dual manual button (cut / coagulation)

Cable length

320 cm

Shaft diameter

2.38 mm (3/32") – universal fit

Connector

Universal tripolar, Valleylab type

Included electrode

Reusable blade F4060 (shaft 2.38 mm, 69 mm)

Material

Sterilisable high-resistance plastic, latex-free; silicone cable

Number of sterilisations

30 guaranteed cycles (handle); up to 20 cycles for the electrode

Sterilisation

Autoclave (134 °C / 2.05 bar, 12 min or 121 °C / 1.05 bar, 20 min) or ethylene oxide

Presentation

Box of 5 pieces, individual packaging – non-sterile

Compatibility

Electrosurgical generators with Valleylab-type connector (Covidien/Medtronic and similar)

Electrical safety

EN 60601-1 · EN 60601-2-2 · EN 60601-1-2

Regulatory classification

Medical device class IIb (Directive 93/42/EEC and amendments)

Reviews

There are no reviews yet.

Be the first to review “Reusable electrosurgical pencil (handle) with push-buttons – Valleylab-type connector, 320 cm cable and 2.38 mm (3/32″) shank, with F4060 blade – Ref. F4141 (30 sterilisation cycles, box of 5 units)”

Your email address will not be published. Required fields are marked *