Description
1. Commercial Introduction
We offer you the reusable adapter FIAB F4832, the connection accessory that allows you to use your electrosurgery pencils with manual switching (handswitch) and Valleylab-type connector on Erbe electrosurgical equipment from the ACC and ICC series. A small, robust and reusable piece that avoids having to duplicate instruments when the operating theatre combines generators from different manufacturers.
Manufactured by FIAB SpA (Vicchio, Florence – Italy), a prestigious reference in European electrosurgery, this adapter is supplied as a single unit individually packaged, ready to be integrated into the reusable equipment of the operating theatre service without replacement costs per procedure.
2. Technical explanation
In monopolar electrosurgery, the generator’s output sends the high-frequency current to the active electrode through the pencil. For manual activation (cut and coagulation) to reach the generator correctly, the pencil connector must be electrically compatible with the equipment socket: this piece acts as a transition element between the Valleylab-type connector of the push-button pencil and the 5 mm plug of the Erbe ACC/ICC generators.
It is a reusable adapter, designed for continuous use in the operating theatre, which maintains the activation and high-frequency signal throughout the procedure and is sterilised together with the rest of the instruments.
3. Technical Specifications
Data verified against the official FIAB Electrosurgery catalogue (section Pencils > Accessories > Adapters For Handswitch Pencils).
Manufacturer’s specifications
| Characteristic | Value |
|---|---|
| Reference FIAB | F4832 |
| Packaging | Single unit |
| Units per box (web) | 1 |
| Compatible with | Erbe ACC/ICC |
Technical features of the adapter
| Characteristic | Specification |
|---|---|
| Brand / Manufacturer | FIAB SpA (Vicchio, Florence – Italy) |
| Product type | Reusable adapter for electrosurgery pencils with manual switching (handswitch) |
| Pencil connection | Valleylab type |
| Connection to the generator | 5 mm plug |
| Compatibility | Erbe ACC/ICC electrosurgical generators |
| Use | Reusable (clinical/hospital environment) |
| Presentation | Single unit |
4. Equivalences and Cross-References
Reference F4832 is the official manufacturer code FIAB, verified in the FIAB Electrosurgery catalogue (Adapters For Handswitch Pencils), which describes it as a reusable adapter for pencils with manual switching, from Valleylab type to 5 mm plug, compatible with Erbe ACC/ICC equipment. Cahnos markets it without altering characters (FIAB-F4832 / F4832).
No equivalence table validated by the manufacturer that cross-references this code with codes from other brands has been found in public sources; no 1:1 equivalences are indicated. Within the same FIAB family there are other reusable adapters for pencils with different connector configurations (references F4830, F4834, F4835, F4836, F4837 and F4838), used mainly on foot-switch activation handles: before selecting one, the type of pencil (handswitch versus foot switch) and the diameter of the plug on your generator must be verified. Consult the technical department before use in tenders.
5. Selection tips and variants of interest
Check the standard of your equipment’s output: this reference is designed for pencils with manual switching (push buttons) and Valleylab-type connector, converting the output to a 5 mm plug for Erbe ACC/ICC generators. If your fleet of Erbe generators or other equipment uses 4 mm, 6 mm or 8 mm sockets, the range of reusable adapters FIAB F4830–F4838 covers different conversions and must be chosen according to the exact diameter of the socket.
To complete the reusable equipment, combine this adapter with the electrosurgical pencils with push buttons from the FIAB family itself (F4243, F4246 or F4256) or with the foot-switch control handles (F4814) that already incorporate the compatible adapter box described in their datasheet. In all cases, follow the internal cleaning and sterilisation protocols for the accessory and remove the piece if there is any deterioration of the body or the contacts.
For institutional volumes, operating theatre equipment or tenders with special B2B conditions, contact our technical-commercial department at Cahnos. Product without published price: sale on quotation.