Description
1. Product Presentation
The disposable electrosurgical unit Ref. F4797/PL is a single-use monopolar electrosurgical pencil with dual push-button manual control, designed for open surgery requiring reliable coagulation and haemostasis, sterile and ready to use in each procedure. It incorporates the F4067 ball electrode (Ø 4 mm, length 156 mm), 2.38 mm (3/32″) shaft, 320 cm cable and universal 3-pin Valleylab-type connector, compatible with virtually all electrosurgical generators (ESU) found in operating theatres, hospitals, clinics and ambulatory surgery centres. Its presentation in a box of 50 units with individual sterile packaging optimises logistics, stock control and cost per procedure for institutional purchasing.
2. Technical explanation and operation
The electrosurgical unit acts as the active electrode of a monopolar electrosurgery system: the pencil connects to the generator (ESU) via the 3-pin Valleylab-type connector and the circuit is completed through the dispersive (neutral) plate placed on the patient. The two buttons on the handle allow the cutting (CUT) and coagulation (COAG) functions to be activated without the need for a foot pedal, giving the surgeon ergonomic, direct control during dissection and haemostasis.
The F4067 ball-type electrode (Ø 4 mm, 156 mm long) concentrates the high-frequency energy on a spherical surface, making it particularly effective for coagulation and haemostasis of small vessels and soft tissue, with less carbonisation of adjacent tissue than other tip formats. Its long tip length (156 mm) provides excellent reach in deep surgical fields. The 2.38 mm shaft ensures rigidity and stability during handling, while the 320 cm cable provides ample freedom of movement around the operating table. As a sterile single-use device, it eliminates the risk of cross-contamination and the costs of cleaning and re-sterilisation.
3. Technical Specifications
| Specification | Value |
| Product type | Disposable electrosurgical unit (electrosurgical pencil) with push-button |
| Manufacturer reference | F4797/PL |
| Included electrode | Sphere F4067 (Ø 4 mm, length 156 mm) |
| Shaft diameter | 2.38 mm (3/32″) |
| Cable length | 320 cm |
| Connector | Valleylab tripolar (3 pins) |
| Control | Double push button: cut (CUT) and coagulation (COAG) |
| Material | High-strength plastic · latex-free · stainless steel electrode |
| Sterility | Sterile, single-use |
| Presentation | Box of 50 units · sterile individual packaging |
| Electrical safety | IEC EN 60601-1 / IEC EN 60601-2-2 |
Variants of the same family (same button pencil, 2.38 mm shaft and Valleylab connector):
| Reference | Electrode | Cable | Connector |
| F4797 | Blade F4046 | 320 cm | Valleylab |
| F4797/5 | Blade F4046 | 500 cm | Valleylab |
| F4797/A | Needle F4048 | 320 cm | Valleylab |
| F4797/AL | F4052 needle electrode | 320 cm | Valleylab |
| F4797/L | Blade F4050 | 320 cm | Valleylab |
| F4797/P | Sphere F4044 | 320 cm | Valleylab |
| F4797/PL | Sphere F4067 | 320 cm | Valleylab |
| F4797/SE | Blade F4046/F4050 with smoke evacuator | 300 cm | Valleylab |
4. Equivalences and Cross-References
The F4067 ball configuration (Ø 4 mm) on a 2.38 mm shaft is a specific coagulation tip; no official manufacturer cross-reference table validating an exact 1:1 equivalence with pencils from other brands has been found in public sources. The following references have been checked against primary sources (GUDID/FDA, official manufacturer catalogue and authorised distribution) and should be treated as comparable at category level, not interchangeable 1:1.
| Manufacturer / Brand | Family / Model | Reference Code / SKU | Equivalence Level | Technical Differences / Remarks |
| Medtronic / Valleylab (Covidien) | Disposable push-button pencil with stainless steel blade, 10 ft cable (≈3 m) | E2516 (box of 50) | Comparable by category (not interchangeable 1:1) | Disposable push-button pencil with standard 3-pin connector; accepts 3/32″ (≈2.4 mm) shaft electrodes. Supplied with blade, not with F4067 ball. Code and attributes verified in GUDID (Covidien LP) and authorised distribution. |
| Erbe | Disposable 2-button electrosurgical pencil, 3 m cable, spatula electrode and sheath | 20190-107 (box of 10) | Comparable by category (not interchangeable 1:1) | Disposable 2-button pencil for VIO/ICC/ACC units and non-Erbe equipment; international 3-pin connector. Spatula-type tip, not F4067 ball; presentation in box of 10. Verified in official Erbe catalogue. |
| ConMed | GoldLine® Push Button, 1″ (2.54 cm) blade, 10 ft cable and sheath | 130309A (box of 40) | Comparable by category (not interchangeable 1:1) | Disposable push-button pencil with preloaded blade and standard 3-pin connector; the tip geometry (blade versus ball) and the presentation format (40 units) differ. Code and attributes verified in ConMed catalogue and authorised distribution. |
| Bovie / Aspen Surgical | Disposable push-button pencil with blade, standard 3-pin connector | ESP1 (box of 50) | Comparable by category (not interchangeable 1:1) | Disposable push-button pencil with blade, individual sterile packaging, standard 3-pin connector; the tip geometry differs (blade versus F4067 ball). Verified in authorised Bovie/Aspen distribution. |
There is no equivalence table validated by the manufacturer; consult the technical department before use in tenders. Before any substitution, check field by field: electrode geometry and length (Ø 4 mm ball versus blade, needle or spatula), shaft diameter (3/32″ ≈ 2.4 mm), connector type (3-pin Valleylab versus other systems), cable length, presentation and regulatory classification of the actual product.
5. Selection tips and variants of interest
The choice of electrode determines the effect on tissue: the F4067 ball of this pencil (156 mm) is designed for coagulation and haemostasis of large surfaces and soft tissue with long tip reach; the F4044 ball (F4797/P) offers the same philosophy with a shorter tip (70 mm), ideal for areas with limited access; the F4046/F4050 blades (F4797, F4797/L) are the versatile option for cutting; and the F4048/F4052 needles (F4797/A, F4797/AL) provide maximum precision in pinpoint incisions. If more reach around the operating table is needed, request the F4797/5 variant (500 cm cable) and, for procedures with surgical smoke, the F4797/SE version with integrated evacuator. Always verify that your generator has a 3-pin Valleylab-type socket: if you use an Erbe generator, request the direct connection variant (/ERB) or the F4831, F4832 and F4833 adapters for Berchtold, Martin and Erbe generators. As complementary accessories from the electrosurgery catalogue, the F4545 sterile sheath and the F7520 sterile tip cleaner are recommended to maintain electrode performance during the procedure.