Description
Paediatric return plate with REM system and TAB connection
The paediatric foam electrosurgical plate is the essential return electrode for completing the monopolar circuit of any high-frequency electrosurgery unit when operating on paediatric patients. This reference, with dimensions 148 x 90 mm and 72 cm² of active surface area, incorporates a REM system and cable-free tab (TAB) connection, making it compatible with generators that have their own plate connection cable or with the reusable cables in our range. It is supplied in a box of 25 units with individual heat-sealed packaging that protects the hydrogel until the moment of use: the option chosen by hospitals, clinics and ambulatory surgery centres that need reliable, consistent consumption and permanent stock for their paediatric surgical schedule.
How the REM return plate works and why it is important
In any procedure using a monopolar electrosurgical unit, the high-frequency current must return to the generator through a low current-density path. The neutral plate collects this current in a controlled manner: if contact is not correct or the surface area is insufficient, the current density at the return point increases and, with it, the risk of thermal burn to the patient.
This plate is constructed with a polyethylene foam backing and conductive hydrogel adhesive, which promotes adaptation to the body contour of the paediatric patient and uniform contact with the skin. Its active surface is divided into two sections, allowing operation with generators equipped with a REM (Return Electrode Monitoring) system: the unit continuously measures contact impedance and, if it detects a loss of contact quality or a fault, it interrupts energy delivery before injury occurs. As this is a cable-free plate with TAB connection, the professional must use a reusable plate cable (F7902 universal 6.3 mm, F7903 Valleylab type or F7904 for Erbe) that connects to the plate tab, a configuration that reduces the cost per procedure and facilitates stock management.
Technical specifications
| Reference / SKU | F7820P |
| Product type | Disposable return plate (electrode) for monopolar electrosurgery |
| Population | Paediatric (patient weight < 15 kg) |
| Measurements | 148 x 90 mm |
| Active surface | 72 cm² |
| Support | Polyethylene foam |
| Conductive adhesive | Hydrogel |
| Safety system | REM (Return Electrode Monitoring), plate divided into two sections |
| Cable | Not included (tab connection) |
| Connection | TAB (tabs) |
| Use | Single-use (disposable, non-sterile) |
| Latex-free | Yes |
| Presentation | Box of 25 units with individual packaging |
| Regulatory standard | Directive 93/42/EEC; EN 60601-2-2; EN ISO 10993-1 |
| Regulatory classification | CE-marked medical device, class IIb (family F7820) |
Paediatric range of the foam family (148 x 90 mm): all references share the same foam backing and hydrogel adhesive; the difference lies in the REM system and the method of connection to the generator.
| Reference | Measurements | Active surface | REM | Cable | Connection |
|---|---|---|---|---|---|
| F7805P | 148 x 90 mm | 78 cm² | No | No | TAB (tabs) |
| F7805PW/V | 148 x 90 mm | 78 cm² | No | 300 cm | Valleylab |
| F7820P | 148 x 90 mm | 72 cm² | Yes | No | TAB (tabs) |
| F7820PW/V | 148 x 90 mm | 72 cm² | Yes | 300 cm | Valleylab |
Equivalences and cross-references
There is no official equivalence table validated by the manufacturer that cross-references this reference with codes from other brands. The comparisons shown below are functional comparisons verified against official documentation from each cited manufacturer (technical catalogue and GUDID/FDA regulatory database) and should not be treated as 1:1 interchangeable without validation by the technical department.
| Manufacturer / Brand | Family / Model | Reference Code / SKU | Equivalence Level | Technical Differences / Remarks |
|---|---|---|---|---|
| Nissha Medical Technologies (SWAROPLATE) | Split Diathermy Pad, Paediatric, REM connector | 1253H / 6000834N (1253H-SWARO-V1-8-96) | Comparable by category (not interchangeable 1:1) | Very similar option: paediatric split plate with REM control, polyethylene foam backing (PE foam) and conductive hydrogel adhesive, for infants and premature babies < 5 kg. Differences: dimensions (165 x 74 mm versus 148 x 90 mm), active surface area (≈66 cm² versus 72 cm²) and connector (Valleylab-REM with cable versus cable-free TAB connection). Code and attributes verified in the manufacturer’s official catalogue. |
| Medtronic / Covidien (Valleylab) | Valleylab™ REM PolyHesive™ II Infant/Paediatric Return Electrode with cable | E7510-25 (box of 25) | Comparable by category (not interchangeable 1:1) | Same REM return function for paediatric/infant patients (< 13.6 kg / 30 lb), single-use, non-sterile and latex-free. They differ in the substrate (PolyHesive™ II hydrogel adhesive versus foam backing), exact dimensions, cable length (2.7 m / 9 ft versus cable-free) and connector (Valleylab-type REM versus TAB). Code verified on Medtronic’s official website and in the GUDID/FDA database. |
The cited references have been cross-checked against official sources from each manufacturer (technical catalogue, official website and GUDID/FDA regulatory databases). Although they coincide in function (return with REM control, paediatric population, single-use), they differ in substrate, exact dimensions and connection system, and are therefore not 1:1 interchangeable. Other equivalent families in the sector that should be evaluated with their technical data sheet in hand before any substitution: paediatric split neutral plates from B. Braun / Aesculap, the Skintact® series from Leonhard Lang, BOWA Medical, EMED and REM split plates of Italian manufacture. In all cases, check field by field: exact dimensions, active surface area in cm², backing and hydrogel type, REM system, physical TAB connector and compatibility with the installed electrosurgical generator. Before any substitution, check these same parameters field by field. There is no equivalence table validated by the manufacturer; consult the technical department before using in tenders.
Selection tips and variants of interest
Check your generator’s connection system: this cable-free F7820P plate with TAB connection requires a reusable plate cable (F7902 universal 6.3 mm, F7903 Valleylab type or F7904 for Erbe) that is reused between patients and reduces the cost per procedure. If you prefer a plate with integrated cable, reference F7820PW/V offers the same plate body (148 x 90 mm, 72 cm², REM) with a 300 cm cable and Valleylab connection, for direct use without an additional cable.
For paediatric patients who do not require REM control, the equivalent alternative in the same size is the F7805P (148 x 90 mm, 78 cm², without REM, TAB connection); for adults, the F7820 (202 x 101 mm, 118 cm², REM, TAB) and for neonates the F7820N (88 x 74 mm, 33-36 cm², cable-free, TAB). To extend compatibility with other generators, adapters F4828 (from Valleylab-type connection to 4 mm plug) and F4829 (to 6.3 mm plug) can be used. Always apply to dry, intact, hair-free skin, on a well-vascularised muscle area and away from prominent bony areas, pacemakers and metal implants. Check the integrity of the individual packaging and the expiry date before use: it is a single-use, non-reusable and non-sterile product.