Description
Neonatal foam electrosurgical plate with REM system and TAB connection
The neonatal foam electrosurgical plate Ref. F7820N is the neutral return electrode specifically designed to complete the monopolar circuit in neonatal or very low birth weight patients. With a compact 88 x 74 mm backing and 36 cm² of active surface area, it offers the optimal balance between contact area and adaptation to the newborn’s anatomy, minimising current density in the return zone and, with it, the risk of thermal burn.
It is supplied in a box of 25 units with individual heat-sealed packaging that protects the conductive hydrogel until the moment of use: the consumable format chosen by hospitals, clinics and neonatology units that require permanent stock, batch traceability and predictable cost per procedure.
Technical explanation: how the return plate with REM system works
In any procedure using a monopolar electrosurgical unit, the high-frequency current leaving the active electrode (handpiece or pencil) passes through the tissue and must return to the generator via a low current density path. The neutral plate collects this current in a controlled manner: if contact is not correct or the surface is insufficient, the current density at the return point increases and, with it, the risk of burn to the patient.
This plate incorporates a REM system (Return Electrode Monitoring): its conductive surface is divided into two sectors, allowing the electrosurgical generator to continuously measure the skin-plate contact impedance. If the system detects partial detachment, dry gel or a failure in contact quality, it interrupts energy emission before injury occurs. The polyethylene foam backing with conductive hydrogel adhesive promotes adaptation to the neonate’s body contour and uniform contact with the skin.
As this is a cable-free plate with TAB connection (tabs), the professional must use a reusable plate cable (F7902 universal 6.3 mm, F7903 Valleylab type or F7904 for Erbe) that connects to the plate tab and is reused between patients: a configuration that reduces the cost per procedure and facilitates warehouse management.
Technical specifications
| Characteristic | Value |
| Reference / SKU | F7820N |
| Product type | Disposable return plate (electrode) for monopolar electrosurgery |
| Population | Neonatal |
| Backing measurements | 88 x 74 mm |
| Active surface | 36 cm² |
| Support | Polyethylene foam |
| Conductive adhesive | Hydrogel |
| Safety system | REM (Return Electrode Monitoring), plate divided into two sections |
| Cable | Not included (tab connection) |
| Connection | TAB (tabs) |
| Use | Single-use (disposable, non-sterile) |
| Latex-free | Yes |
| Presentation | Box of 25 units with individual packaging |
| Regulatory classification | CE-marked medical device, class IIb (family F7820) |
Neonatal range of the foam family (88 x 74 mm): all references share the same foam backing and hydrogel adhesive; the difference lies in the REM system and the method of connection to the generator.
| Reference | Measurements | Active surface | REM | Cable | Connection |
| F7805N | 88 x 74 mm | 36 cm² | No | No | TAB (tabs) |
| F7820N | 88 x 74 mm | 36 cm² | Yes | No | TAB (tabs) |
| F7820NW/V | 88 x 74 mm | 36 cm² | Yes | 300 cm | Valleylab |
Equivalences and cross-references
There is no official equivalence table validated by the manufacturer that cross-references this reference with codes from other brands. The comparisons shown below are functional comparisons verified against public documentation from each cited manufacturer (GUDID/FDA regulatory databases and official catalogues) and should not be treated as 1:1 interchangeable without validation by the technical department.
| Manufacturer / Brand | Family / Model | Reference Code / SKU | Equivalence Level | Technical Differences / Remarks |
| Medtronic / Covidien (Valleylab) | Valleylab™ REM PolyHesive™ II Infant Patient Return Electrode | E7510-25 (UDI 10884524001043) | Comparable by category (not interchangeable 1:1) | Same single-use REM return function for paediatric patients from 2.7 to 13.6 kg. They differ in substrate (PolyHesive™ II hydrogel vs. foam backing), exact dimensions and connection (integrated 2.7 m Valleylab cable vs. cable-free TAB). Code and attributes verified in GUDID/FDA and on the manufacturer’s official website. |
| Nissha Medical Technologies (Graphic Controls) | SWAROPLATE Split Diathermy Pad Infant/Pediatric REM | 1253H (PMF 6000834N) | Comparable by category (not interchangeable 1:1) | Very similar option: split plate with REM control, polyethylene foam backing (PE foam) and conductive hydrogel adhesive, for infants and premature babies < 5 kg. Differences: dimensions (165 x 74 mm vs. 88 x 74 mm), active surface area (≈66 cm² vs. 36 cm²) and connector (Valleylab-REM with cable vs. cable-free TAB). Code and attributes verified in the manufacturer’s official catalogue. |
| BOENMED | Grounding Pad with Cable, Bipolar, Valleylab REM Plug, Infant | 512412 | Comparable by category (not interchangeable 1:1) | Paediatric return plate with foam backing, conductive hydrogel and Valleylab-type REM plug with cable. They differ in dimensions and active surface area (not officially published), connection (Valleylab with cable vs. cable-free TAB) and presentation (box of 500 units vs. box of 25 units). Code and attributes verified in the manufacturer’s official catalogue. |
The cited references have been cross-checked against regulatory databases (GUDID/FDA) and official catalogues of each cited manufacturer. Although they match in function (return with REM control, neonatal/paediatric population and single-use), they differ in substrate, exact dimensions, active surface area and connection system, so they are not 1:1 interchangeable. There is no equivalence table validated by the manufacturer; consult the technical department before use in tenders. Before any substitution, check field by field: exact dimensions, active surface area in cm², backing and hydrogel type, REM system, physical TAB connector and compatibility with the installed electrosurgical generator.
Other homologous families in the sector that should be evaluated with the technical data sheet in hand: split paediatric/neonatal neutral plates from B. Braun / Aesculap, disposable REM bipartite plates from GIMA and paediatric neutral electrodes from manufacturers present on technical portals such as EMED, Promed or Upspring Medical.
Selection tips and variants of interest
- Check your generator connection: the cable-free F7820N plate with TAB connection requires a reusable plate cable (F7902 universal 6.3 mm, F7903 Valleylab type or F7904 for Erbe). Choose the cable according to the plug on your electrosurgical unit.
- Prefer a plate with integrated cable? Reference F7820NW/V offers the same plate body (88 x 74 mm, 36 cm², REM) with a 300 cm cable and Valleylab-type connection, for direct use without an additional cable.
- Does your generator not have REM control? The equivalent alternative in the same size with TAB connection is the F7805N (88 x 74 mm, 36 cm², without REM).
- Other populations: paediatric F7820P / F7820PW/V (148 x 90 mm, 72 cm²) and adult F7820 / F7820W/V / F7820W/6.3 (202 x 101 mm, 118 cm²) with the same REM technology.
- Extended compatibility: adapters F4828 (from Valleylab-type connection to 4 mm plug) and F4829 (to 6.3 mm plug) for other generators.
- Application: always place on dry, intact, hair-free skin, on a well-perfused muscle area and away from prominent bony areas. In neonatal patients, especially verify the available contact surface and do not cut or reuse the plate. Check the integrity of the individual packaging and the expiry date before each use: it is a single-use and non-sterile product.