Description
1. Introduction
The reusable monopolar wire and ribbon loop electrodes with a 2.4 mm Ø shaft from REGER Medizintechnik GmbH are active electrodes designed for excision, resection and coagulation of soft tissue using high-frequency current. Thanks to their reusable nature and sterilisation in an autoclave at 134 °C, they help optimise the cost per procedure in operating theatres, minor surgery clinics, gynaecology and colposcopy, while maintaining stable cutting performance and excellent specimen preservation.
2. Technical explanation
In a monopolar electrosurgery system, the active electrode concentrates the current emitted by the electrosurgical generator at the point of contact with the tissue, with the circuit returning through the neutral plate. The geometry of the loop delimits the resection area and allows tissue samples to be taken with defined margins, simultaneously controlling cutting and haemostasis. The 2.4 mm Ø shaft incorporates hexagon anti-twist protection, which secures the electrode to the electrode holder handle and prevents accidental rotation during use. They are suitable for high-frequency (HF) and radiofrequency (RF) generators, with a maximum permissible voltage that, depending on the reference, reaches 500 Vp or 4.3 kVp. The wire loops of references 306-1 and 306-2 are made of tungsten (wolfram), ensuring clean, stable cutting with minimal carbonisation.
3. Technical Specifications
| Characteristic | Value |
|---|---|
| Electrode type | Reusable active monopolar loop electrode |
| Models | Wire loop and ribbon loop |
| Shaft diameter | Ø 2.4 mm |
| Use | Reusable |
| Sterilisation | Autoclavable at 134 °C |
| Protection | Hexagon anti-twist protection |
| Compatibility | HF and RF generators |
| Maximum voltage (Umax) | 500 Vp (ref. 353-0) · 4.3 kVp (ref. 306-1) |
| Loop material | Tungsten (wolfram) in wire loops 306-1 and 306-2 |
| Units per pack | 5 (series 306 and 307) · 1 (series 346 to 361) |
Available references
| REGER reference | Size | Units per pack | Compatible with |
|---|---|---|---|
| 306-1 | Wire loop · Dimensions Ø 5 mm | 5 | HF and RF generators |
| 306-2 | Wire loop · Dimensions Ø 10 mm | 5 | HF and RF generators |
| 306-3 | Wire loop · Diameter Ø 14 mm | 5 | |
| 307-2 | Ribbon loop · Diameter Ø 10 mm | 5 | |
| 307-3 | Ribbon loop · Diameter Ø 14 mm | 5 | |
| 353-0 | Length 110 mm · Loop Ø 5 mm | 1 | |
| 346-0 | Length 110 mm · Loop 10 mm x 10 mm | 1 | |
| 350-1 | Length 110 mm · Loop 15 mm x 15 mm | 1 | |
| 350-0 | Length 110 mm · Loop 20 mm x 15 mm | 1 | |
| 351-0 | Length 110 mm · Loop width 20 mm · Loop height 20 mm | 1 | |
| 352-0 | Length 110 mm · Loop width 20 mm · Loop height 25 mm | 1 | |
| 358-0 | Diameter Ø 10 mm · Length 110 mm | 1 | HF and RF generators |
| 359-0 | Diameter Ø 15 mm · Length 110 mm | 1 | HF and RF generators |
| 360-0 | Diameter Ø 20 mm · Length 110 mm | 1 | HF and RF generators |
| 361-0 | Diameter Ø 25 mm · Length 110 mm | 1 | HF and RF generators |
4. Equivalences and Cross-References
There is no manufacturer-validated equivalence table cross-referencing the references in this family (306-1 to 361-0) with codes from other brands on a 1:1 interchangeable basis. In the sector, there are homologous families of reusable loop electrodes with a 2.4 mm Ø shaft from manufacturers such as Erbe, KLS Martin, Bovie (Aspen Surgical) and Söring, as well as sterile disposable LEEP/LLETZ ranges from CooperSurgical, Megadyne and Medline. These alternatives are offered solely as a functional category comparison, never as direct substitutes. Within REGER’s own catalogue, the same HF electrosurgery system offers the family of wire and ribbon loops with a 4 mm Ø shaft (refs. 212-1 to 261-1, SKU RGR-HF-ASA-4) and the conisation families with 2.4 mm Ø shafts (refs. 364-0, 362-0 and 363-0) and 4 mm Ø shafts (refs. 264-0, 262-0 and 263-0). Before any substitution in tenders or surgical protocols, check shaft diameter, connector, fixation system and dimensions with the technical department at Cahnos.
5. Selection tips and variants of interest
Select the loop type and size according to the lesion and technique: wire loops of Ø 5 mm, Ø 10 mm and Ø 14 mm (refs. 306-1, 306-2 and 306-3, pack of 5 units) are the usual options for small resections and cervical excisions; ribbon loops of Ø 10 mm and Ø 14 mm (refs. 307-2 and 307-3) offer a larger contact surface for wider resections. For access with a fixed handle, series 353-0 to 352-0 (110 mm length) allow precise adjustment of the loop size: Ø 5 mm loop, 10×10 mm, 15×15 mm, 20×15 mm and 20×20/25 mm formats. Refs. 358-0 to 361-0 cover diameters from Ø 10 mm to Ø 25 mm with a 110 mm length. Always verify that your electrode holder handle accepts a 2.4 mm Ø shaft with hexagonal anti-twist fixation. As they are reusable, sterilise them in an autoclave at 134 °C after each use and periodically inspect the condition of the loop, removing the electrode if any deformation or deterioration is observed. For hospital purchases, tenders or volume conditions, consult your technical-commercial advisor at Cahnos.





Reviews