Description
1. Product Presentation
The disposable electrosurgery electrode Ref. F4050S/TEF is the reference consumable when a long blade with non-stick performance and maximum protection of healthy tissue is required in open surgery. Its 150 mm Teflon-coated blade tip slides over tissue without dragging carbonised eschar, and its 145 mm insulating sleeve concentrates energy at the active tip, protecting adjacent structures in deep access sites. Each box includes 24 sterile single-use units, ready to open at the time of the procedure: a safe solution that eliminates the risk of cross-contamination and the costs of re-sterilisation. Available with clinical stock for hospitals, clinics, ambulatory surgery centres and minor surgery clinics.
2. Technical Explanation
This electrode acts as the active electrode of a monopolar electrosurgery system: the radiofrequency current generated by the electrosurgical unit (ESU) travels to the blade tip, where it concentrates at the cutting edge to cut tissue or coagulate small vessels with precise control, returning through the patient’s neutral (dispersive) plate.
- 150 mm blade: long reach for cutting and dissection in deep surgical fields, with stability and precision.
- Teflon (PTFE) coating: non-stick effect that prevents tissue from adhering to the tip, reduces carbonisation and eschar formation, maintains conductivity and avoids interruptions to clean the tip during the procedure.
- 145 mm insulating sleeve: insulates virtually the entire shaft, concentrating energy at the active end and protecting adjacent tissues from accidental contact in delicate areas.
- 2.38 mm (3/32″) shaft with hexagonal anti-rotation locking: universal fixation standard, compatible with virtually all pencils and electrosurgical units on the market (Valleylab type and compatible), without the need for adapters.
- Sterile single-use: sterilised with ethylene oxide (EO), latex-free, ready to use in each procedure.
3. Technical Specifications
| Specification | Detail |
|---|---|
| Reference / SKU | F4050S/TEF |
| Product type | Disposable monopolar electrosurgery electrode, blade tip with insulating sleeve |
| Blade length | 150 mm |
| Insulating sleeve | 145 mm |
| Coating | Non-stick Teflon (PTFE) |
| Electrode material | Surgical grade stainless steel |
| Tip shape | Integrated blade |
| Shaft diameter | 2.38 mm (3/32″) — universal standard |
| Fixation system | Hexagonal anti-rotation lock |
| Compatibility | Pencils and electrosurgical units with 2.38 mm (3/32″) receptacle, Valleylab type and compatible (ConMed, Bovie, Erbe via adapter) |
| Use | Disposable, single-use |
| Sterility | Sterilised with ethylene oxide (EO), individually heat-sealed packaging |
| Latex | Latex-free |
| Presentation | Box of 24 units |
| Classification | Medical device class IIb · CE marking |
Variants of the same family (Teflon coating, 2.38 mm shaft, box of 24 units):
| Reference | Length | Tip shape |
|---|---|---|
| F4044/TEF | 70 mm | Sphere Ø 4 mm |
| F4046/TEF | 70 mm | Blade |
| F4048/TEF | 70 mm | Needle |
| F4050/TEF | 150 mm | Blade |
| F4050S/TEF | 150 mm (sleeve 145 mm) | Blade with sleeve (this product) |
| F4050/16TEF | 165 mm | Blade |
| F4052/TEF | 150 mm | Needle |
| F4053/TEF | 100 mm | Blade |
| F4053S/TEF | 100 mm (95 mm sleeve) | Blade with sleeve |
| F4067/TEF | 150 mm | Sphere Ø 4 mm |
4. Equivalences and Cross-References
Reference F4050S/TEF belongs to the family of disposable electrodes with Teflon coating, 2.38 mm shaft and insulating sleeve configuration. No official manufacturer cross-reference table has been found in public sources to validate an exact 1:1 equivalence with other brands. The following rows are functional comparisons verified against official manufacturer catalogues, authorised distribution and the GUDID/FDA regulatory database, and should be treated as comparable at category level, not 1:1 interchangeable without validation by the technical department.
| Manufacturer / Brand | Family / Model | Reference Code / SKU | Equivalence Level | Technical Differences / Remarks |
|---|---|---|---|---|
| Avicenna Surgical (United Kingdom) | Single Use Blade Electrode, Non-Stick Teflon Coated, 152 mm, Extended Insulation (3 mm active tip), hexagonal lock | AVBE012 (pack of 10) | Comparable by category (not interchangeable 1:1) | Same functional configuration: 152 mm Teflon-coated blade (vs 150 mm) with extended shaft insulation equivalent to the 145 mm sleeve and hexagonal locking. The exact length, presentation (pack of 10 vs 24) and manufacturer differ. Code and attributes verified on the manufacturer’s official website. |
| Avicenna Surgical (United Kingdom) | Single Use Blade Electrode, Non-Stick Teflon Coated, 152 mm, hexagonal lock | AVBE010 (pack of 10) | Comparable by category (not interchangeable 1:1) | Teflon (PTFE)-coated blade with almost identical length (152 mm vs 150 mm) and equivalent hexagonal fixation; does not incorporate extended insulation/sleeve. The presentation (pack of 10 vs 24) and manufacturer differ. Code and attributes verified on the manufacturer’s official website. |
| Prima Medical (United Kingdom) | NonStick™ blade electrode, PTFE-coated, 165 mm, shielded tip | PMS 2933 (pack of 10) | Comparable by category (not interchangeable 1:1) | Same non-stick Teflon coating with shielded tip, function equivalent to the insulating sleeve; the length (165 mm vs 150 mm) and presentation (pack of 10 vs 24) differ. Code and attributes verified in the brand’s official catalogue and authorised distribution (Eurosurgical). |
| Prima Medical (United Kingdom) | NonStick™ blade electrode, PTFE-coated, 100 mm, shielded tip | PMS 2926 (pack of 10) | Comparable by category (not interchangeable 1:1) | Same concept of non-stick PTFE blade with shielded tip equivalent to the sleeve; the length (100 mm vs 150 mm), presentation (pack of 10) and manufacturer differ. Code and attributes verified in the brand’s official catalogue and authorised distribution. |
| Bovie / Aspen Surgical (USA) | Resistick II™ Coated Extended Blade Electrode 6″ (15.24 cm), PTFE-coated | ES04T (box of 12) | Comparable by category (not interchangeable 1:1) | Disposable sterile extended blade electrode with non-stick PTFE coating; almost identical length (152.4 mm vs 150 mm) and standard 3/32″ shaft. The presentation (box of 12 vs 24), shaft insulation and manufacturer differ. Code and attributes verified in GUDID/FDA (class II, sterile) and the manufacturer’s catalogue. |
There is no manufacturer-validated equivalence table; consult the technical department before use in tenders. Before any substitution, check field by field: blade length (150 mm), insulating sleeve (145 mm), coating (Teflon/PTFE), material (AISI stainless steel), shaft diameter (2.38 mm / 3/32″), fixation system (hexagonal locking), sterility and presentation. The third-party codes cited have been verified in public sources and correspond to electrodes of the same category, not certified 1:1 equivalents.
5. Selection tips and variants of interest
The insulating sleeve configuration of the F4050S/TEF is indicated when a long blade with protection of adjacent tissues is required: cutting and dissection in open surgery, general surgery, dermatology and minor surgery, with long reach in deep fields. If shaft insulation is not required, reference F4050/TEF offers the same 150 mm blade with Teflon coating without sleeve, at a lower cost per unit; the F4053S/TEF maintains the sleeve configuration (95 mm) in a shorter 100 mm format. For even deeper access, the F4050/16TEF extends the blade to 165 mm.
The non-stick coating range is completed with needles (F4052/TEF of 150 mm and F4048/TEF of 70 mm) for precision dissection and pinpoint haemostasis, and with spheres (F4067/TEF of 150 mm and F4044/TEF of 70 mm) for contact coagulation of large surfaces. If cost per procedure is prioritised, the stainless steel electrode F4050 (153 mm uncoated) and its sleeved version F4050/S are the alternatives in the same series; the titanium-coated version F4050T (153 mm) provides an alternative non-stick behaviour. Always verify that your pencil accepts a 2.38 mm (3/32″) shaft with hexagonal locking: Valleylab, ConMed, Bovie, Erbe and compatible electrosurgical pencils accept it. This is a sterile single-use product: do not re-sterilise or reuse. For institutional volumes or tenders, consult our technical-commercial department.





Reviews