Description
Complete vital signs monitoring for adult, paediatric and neonatal
The K12 multiparameter monitor (reference 35308) is a multifunction bedside device with a 12-inch TFT touchscreen and 10-lead ECG, designed for real-time recording and display of the main physiological parameters: ECG, heart rate (HR), non-invasive blood pressure (NIBP), oxygen saturation (SpO2), respiration (RESP) and temperature (TEMP). Thanks to its specific modes for adult, paediatric and neonatal patients, it enables comprehensive analysis of the patient’s physiological condition in hospitalisation, ICU, emergency, resuscitation and operating theatre, as well as intra-hospital transport, with an excellent balance between performance and cost for hospital procurement and tenders.
Technical explanation of the equipment
The K12 integrates all vital signs monitoring functions into a single device and processes them using real-time monitoring algorithms. The ECG signal is captured via 10 leads and features pacemaker pulse detection and inhibition, protection against defibrillator discharge and resistance to interference from the electrosurgical unit, ensuring a stable signal even in surgical environments. Blood pressure is measured by the oscillometric method (NIBP), oxygen saturation by the dual-wavelength optical method (SpO2, compatible with original Nellcor sensors) and temperature via a skin probe. The device operates with an integrated rechargeable lithium battery, allowing continuous use during transfers or power cuts, and connects to the hospital network via Ethernet.
- ergonomic design with intuitive, easy-to-use user interface
- touchscreen with up to 13 simultaneous waveform traces
- USB port for data transfer and software updates
- visual and audible alarm for physiological and technical alarms
- large data storage capacity
- suitable for use with original Nellcor sensors
- protection against defibrillator discharge, resistance to electrosurgical unit interference and detection and inhibition of cardiac pacemaker pulses
- internal software in Spanish (available languages: GB, DE, FR, PT, ES, IT, PL, TR, RU) and PC software in Spanish (also GB and IT)
- compatible with Windows XP, 7, 8, 10 and 11
Technical specifications
Product data
| Manufacturer | GIMA |
|---|---|
| Manufacturer reference | 35308 (SKU GM35308) |
| Sales unit | 1 unit |
| Device type | Medical device (medical device) |
| Device class | II B |
| CND | Z1203020202 |
| GMDN | 33586 |
| EAN | 8023279353082 |
Electrocardiogram (ECG – 10 leads)
| Parameter | Specification |
|---|---|
| Dynamic input range | ±(0.5~5 mVp) |
| Differential input impedance | =10 MΩ |
| Band width | 0.05~150 Hz (diagnostics); 0.5~40 Hz (monitoring); 1~20 Hz (operating theatre) |
| CMRR | =90 dB (diagnostics) / =105 dB (monitoring and operating theatre) |
| Sensitivity selection | ×1/4, ×1/2, ×1, ×2, ×4 and automatic |
| Sweep speed | 6.25 / 12.5 / 25 / 50 mm/s |
| HR measurement range | 15~350 lpm ±1% or ±2 lpm (whichever is greater) |
| Pacemaker | Pacemaker pulse detection and rejection function |
Respiration (RESP)
| Parameter | Specification |
|---|---|
| Measurement range | 0~120 rpm ±5% or ±2 rpm (whichever is greater) |
Temperature (TEMP)
| Parameter | Specification |
|---|---|
| Measurement range | 21.0~50.0 °C ±0.2 °C (25 to 45 °C) |
Non-invasive blood pressure (NIBP)
| Parameter | Specification |
|---|---|
| Technique | Oscillometric method |
| Typical measurement time | <30 seconds (adult cuff) |
| SYS measurement range | 40~275 mmHg (adult) / 40~200 mmHg (paediatric) / 40~135 mmHg (neonate) |
| DIA measurement range | 10~210 mmHg (adult) / 10~150 mmHg (paediatric) / 10~95 mmHg (neonate) |
| MAP measurement range | 20~230 mmHg (adult) / 20~165 mmHg (paediatric) / 20~110 mmHg (neonate) |
| Measurement mode | Manual, automatic, STAT and multi-cycle |
| Measurement accuracy | Mean difference: ±5 mmHg; standard deviation: 8 mmHg |
| Automatic measurement intervals | 1-480 min |
Oxygen saturation (SpO2)
| Parameter | Specification |
|---|---|
| Technique | Dual-wavelength optical method |
| Measurement range | 0%~100% ±2% (for SpO2 range 70~100%) |
| Pulse measurement range | 30~250 lpm ±2 lpm or ±2% (whichever is greater) |
| Low perfusion performance | Up to 0.3% |
Other specifications
| Parameter | Specification |
|---|---|
| Power supply | AC 100-240 V, 50/60 Hz, 60 VA |
| Integrated lithium battery | 11.1 V / 4400 mAh |
| Display | 12 or 15-inch TFT |
| Alarm type | 3-level audible and visible alarm |
| Red | Ethernet |
| Safety standard | IEC 60601-1 |
Included accessories
- Adult SpO2 catheter
- Adult NIBP cuff
- SpO2 extension cable
- ECG cable
- Skin temperature probe
- Power cable
- Rechargeable lithium battery
- Dust cover
- Disposable electrodes
- Manual: GB, FR, IT, ES
- PC software: GB, IT, ES
Equivalences and cross-references
There is no official equivalence table published by the manufacturer relating reference 35308 to codes from other manufacturers; for use as a replacement for other equipment or in tenders, consult the technical department at Cahnos before any substitution. The factory reference Ref. 35308 is distributed as SKU GM35308 with EAN 8023279353082 and is classified as a class II B medical device (CND Z1203020202, GMDN 33586). As a notable compatibility reference, the device is suitable for use with original Nellcor SpO2 sensors, which facilitates integration with the existing sensor fleet at the facility without the need for additional adapters.
Selection tips and variants of interest
Consider the type of patients and the level of care in your unit: the K12 covers monitoring of adult, paediatric and neonatal patients, making it suitable for ICU, emergency, resuscitation, operating theatres, recovery rooms and general hospitalisation areas. If your facility prioritises mobility, note that the device incorporates an integrated rechargeable lithium battery, ideal for intra-hospital transfers and power cuts. Review the accessories included as standard (adult SpO2 probe, adult NIBP cuff, ECG cable, temperature probe and disposable electrodes) and, if your patient mix requires it, plan for paediatric or neonatal cuffs and additional SpO2 probes. Also check the screen version (12 or 15 inches) and Ethernet connectivity for integration into the hospital network. For personalised advice, accessory configuration and a quote, contact the technical-commercial team at Cahnos.





Reviews
There are no reviews yet.