Description
1. Introduction
The MS400 Multiparameter Simulator (factory reference 54602) is the medical electronics tool your maintenance department needs to check, adjust and verify multiparameter patient monitors. In a compact, portable unit weighing just 1.4 kg, the MS400 integrates 12-lead ECG simulation, respiration, temperature and 4-channel invasive blood pressure (IBP), enabling complete functional testing of bedside, ICU, operating theatre and medical transport equipment without relying on real patients.
2. Technical explanation
The MS400 generates reference electrical and physiological signals with adjustable, stable values, allowing the technician to compare the monitor reading against a known standard. In the ECG section it reproduces the standard 12-lead trace with rates from 0 to 350 bpm, simulates 35 types of arrhythmia (atrial, ventricular, extrasystoles and conduction defects) and pacemaker waveforms (asynchronous, demand, sequential A/V, no capture and no function). It also generates a respiratory signal with adjustable rate, impedance and base impedance, 3 apnoea settings, temperature data in 4 types and 4-channel IBP recordings with independent pressure scales.
The unit can simulate square, triangular, pulse and sine waveforms with adjustable frequency, commonly used to verify the dynamic response of monitors. It can also display and export to the unit under test data from the AHA and MIT ECG databases, as well as user-defined waveforms stored on a TF card. The ECG waveform complies with standard ANSI/AAMI EC13-2002.
3. Technical Specifications
| Parameter | Specification |
|---|---|
| Manufacturer | GIMA |
| Manufacturer reference | 54602 |
| SKU Cahnos | GM54602 |
| Device type | Not a medical device |
| ECG | 12 leads |
| ECG rate | 0, 15, 20, 25, 30, 40, 60, 80, 100, 120, 140, 160, 180, 200, 220, 240, 260, 280, 300, 320, 340 and 350 bpm |
| ECG amplitude | 0.5 mV, 1.0 mV, 1.5 mV and 2.0 mV; ±2% |
| Arrhythmias | 35 types (atrial, ventricular, extrasystoles and conduction defects) |
| Pacemaker | Asynchronous, Demand Occ Sinus, Demand Freq Sinus, Sequential A/V, No capture, No function |
| Respiratory rate | 0 (apnoea), 15, 20, 30, 40, 60, 80, 100, 120 and 150 r/min; ±1 r/min |
| Apnoea | 12 s, 22 s, 32 s |
| Impedance | 0.2 Ω, 0.5 Ω, 1.0 Ω and 3.0 Ω; ±3% |
| Temperature range | 0-40 °C; ±0.2 °C |
| IBP | 4 channels (see range table) |
| Display | 3.5” colour TFT touchscreen |
| Keyboard | 10 numeric keys and 7 function keys |
| Power supply | Integrated rechargeable lithium-polymer battery 3,700 mAh / 3.7 V or 9 V DC power supply, 1,000 mA |
| Dimensions | 188.6 × 125 × 43.2 mm |
| Weight | 1.4 kg |
| Regulatory standard | ECG waveform compliant with ANSI/AAMI EC13-2002 |
| Sales unit | 1 unit |
| EAN | 6945040101031 |
Invasive blood pressure (IBP) ranges
| Channel | Simulated pressures (mmHg) |
|---|---|
| IBP 1 | -100, 0, 80, 160, 240, 320 and 400 |
| IBP 2 | -100, 0, 50, 100, 150, 200 and 250 |
| IBP 3 | -100, 0, 20, 40, 60, 80 and 100 |
| IBP 4 | P4 follows the static pressure levels of P3 |
Contents and languages
| Item | Detail |
|---|---|
| Supplied with | Integrated lithium battery, power adapter and manual (GB, IT, DE) |
| Firmware languages | GB, IT, DE, PT, RU |
4. Equivalences and Cross-References
Reference 54602 is the manufacturer’s official reference for the MS400 simulator (SKU Cahnos GM54602), verified in the Product_DB catalogue. No manufacturer-validated equivalence table is available to cross-reference this simulator 1:1 with models from other brands: multiparameter simulators from other manufacturers (simulation ranges of medical electronics equipment such as Fluke Biomedical, Datrend, Rigel or BC Biomedical) are considered comparable by category, not interchangeable 1:1. Consult our technical department before using any cross-reference in a tender.
5. Selection tips and variants of interest
The MS400 is the ideal choice for medical electronics and technical maintenance services that want to cover ECG, respiration, temperature and IBP verification in a single unit. Before requesting a quote, confirm the model and reference printed on your unit (54602 / GM54602) and verify the compatibility of connectors and patient cables with the monitors in your fleet. For periodic calibration plans, record the simulated values of each channel and use the reference waveforms (square, triangular, pulse, sine) as a frequency response standard. This product has no catalogue price: contact our technical-commercial team for terms, lead times and supply options for institutional purchases or maintenance contracts.