Description
1. Introduction
The NEO ECG T180 Tablet (ref. 54231) is a portable electrocardiograph with a compact tablet design, designed for the acquisition, analysis and printing of 9/12-lead electrocardiographic traces. Its high-resolution 10.1″ colour touchscreen, its lightness and its base with integrated thermal printer make it a versatile solution for cardiology practices, hospital services, primary care and community medicine, including telemedicine. A professional device for clinical use that combines the ergonomics of a tablet with the reliability of a 12-channel electrocardiograph.
2. Technical explanation of the equipment
The NEO ECG T180 performs synchronous acquisition of 9/12 leads with 24-bit A/D conversion and a sampling rate of 32,000 samples/second, ensuring excellent signal fidelity. It incorporates the Glasgow interpretation algorithm and offers 4 sampling modes: pre-sampling, real-time sampling, periodic sampling and triggered sampling. The system includes automatic detection and marking of pacemaker stimulation, automatic RR interval analysis, heart rate variability analysis (HRV/VFC), drug test and ECG event mode. Patient information can be entered via virtual keyboard or barcode scanner, and the device stores up to 10,000 ECGs. Wireless data transmission is carried out via WiFi and mobile networks, and records can be exported to USB memory in multiple formats (Carewell ECG, PDF, BMP, HL7, DICOM, SCP). The trace can be frozen on screen for review, and the base integrates a thermal printer, as well as support for an external laser printer. Power is supplied via an external DC power supply or an integrated rechargeable lithium battery.
3. Technical Specifications
| ECG leads | Synchronous acquisition of 9/12 leads · A/D resolution: 24 bits · Heart rate range: 30–300 bpm ± 1 · Sensitivity: automatic, 2.5, 5, 10, 20, 40 mm/mV (error less than ±5%) · Print speed: 5, 6.25, 10, 12.5, 25, 50 mm/s (error less than ±3%) · Sampling rate: 32,000 samples/s |
|---|---|
| Filters | AC (50–60 Hz, off) · EMG (25–35–45 Hz, off) · ADS (0.01–0.05–0.32–0.67 Hz) · Low-pass filter (75–100–150–300 Hz, off) |
| Frequency response | 0.01–350 Hz (+0.4 / −3.0 dB) · Minimum detectable signal: 20 µVp-p |
| Electrical specifications | Internal noise: ≤ 12.5 µV · Input impedance: ≥ 100 MΩ (10 Hz) · Circuit current: ≤ 10 nA · Calibration voltage: 1 mV ± 2% · Depolarization voltage: ± 900 mV, ± 5% · Common mode rejection: ≥ 140 dB (with AC filter) / ≥ 120 dB (without AC filter) |
| Amplitude quantification | 0.95 µV/LSB · Recovery time after discharge: < 10 s · Display: 10.1″ colour touchscreen · Dimensions and weight: 27.4 × 28.8 × 11.1 cm – 2.5 kg (base); tablet 27.3 × 18.2 × 2.6 cm – 1.2 kg |
Key features
- Glasgow algorithm for advanced trace interpretation.
- 4 sampling modes: pre-sampling, real-time, periodic and trigger-based.
- Automatic detection and marking of pacemaker stimulation.
- RR auto-analysis, HRV (VFC) analysis, drug test and ECG event mode.
- Storage of up to 10,000 ECGs and export of patient data to USB memory.
- File formats: Carewell ECG, PDF, BMP, HL7, DICOM and SCP.
- Freezing of the ECG waveform on screen.
- Virtual keyboard and barcode scanning for entering patient data.
- Integrated thermal printer in the base and support for an external laser printer.
- Wireless transmission of ECG data via WiFi and mobile networks.
- External DC power supply and integrated rechargeable lithium battery.
- Selectable multilingual software: ES, FR, IT, PT, DE, EN, GR, PL.
Included accessories
- 1 IEC ECG cable, Ø 4 mm, with banana connector.
- 6 adult suction precordial electrodes (Ø 4 mm).
- 4 adult limb forceps.
- 1 roll of thermal paper, 20 m × 216 mm.
- 1 rechargeable lithium battery.
- 1 power adapter.
- Multilingual manual: GB, IT (FR, ES, PT, DE, GR, PL; RU available on the manufacturer’s website).
Product data
| Manufacturer | GIMA |
|---|---|
| Manufacturer reference | 54231 |
| Sales unit | 1 pc |
| Device type | Medical device (medical device) |
| Device class | II A |
| CND | Z12050301 |
| GMDN | 16231 |
| EAN | 8023279542318 |
4. Equivalences and Cross-References
This device corresponds to the factory reference Ref. 54231 (SKU Cahnos GM54231), belonging to the NEO ECG T180 series. The sister reference 54230 shares the user manual and platform with this model within the same ECG tablet family. There is no manufacturer-validated equivalence table cross-referencing this reference with codes from other manufacturers; comparable ranges by category (portable 12-lead electrocardiographs with touchscreen and banana connectors) must be validated case by case with our technical-commercial team before any substitution in a tender.
5. Selection tips and variants of interest
Check the connection system of your patient cable: this model uses banana-type connectors (IEC plug Ø 4 mm), compatible with clip and suction electrodes. For telemedicine and community medicine applications, WiFi and mobile network transmission allows the ECG to be sent remotely without cables; in the practice, the base with integrated thermal printer and the support for an external laser printer speed up the documentation of the record. Consider the consumable: the device uses a 216 mm wide thermal paper roll; always confirm the original reference when replenishing. For integration with electronic health record systems, the HL7, DICOM and SCP formats facilitate interoperability. Consult our technical-commercial department to validate the most suitable reference for your equipment fleet and volume supply conditions. Product without reference price: For large orders you can request a quote.













Reviews