Description
1. Reliable backup power for your Philips IntelliVue MP20 monitor
Keep your Philips IntelliVue MP20 vital signs monitor always operational with this rechargeable lithium-ion battery, compatible with original reference MFR 04605A. With 10.8 V and 6 Ah (64.8 Wh) capacity, this battery ensures that patient monitoring —ECG, SpO₂ pulse oximetry, blood pressure and other parameters— is not interrupted during intra-hospital transfers, ambulance transport or a power supply failure.
It is the replacement that medical electronics, maintenance and hospital purchasing departments need to have in stock to restore the monitor to its original battery life in minutes and without tools, at a significantly lower cost than the official factory spare part.
2. How this battery works in the monitor
Philips IntelliVue MP series monitors operate powered by the mains or, in portable mode, by their internal rechargeable battery. While the equipment remains plugged in, the battery is automatically kept charged; when disconnected, moved or during a power cut, the battery takes over without interrupting monitoring or patient data recording.
Lithium-ion (Li-ion) chemistry offers high energy density, stable 10.8 V voltage, virtually no memory effect and a long cycle life if the maintenance guidelines indicated at the end of this sheet are followed. A degraded cell manifests as loss of battery life, premature low battery warnings or shutdowns during transport; preventive replacement with this reference restores the equipment to its original performance.
3. Technical Specifications
| Characteristic | Specification |
| Reference Cahnos (SKU) | 88888945 |
| Original reference cited | 04605A (in official Philips documentation: M4605A / 989803135861) |
| Application | Philips IntelliVue MP20 vital signs monitor; pack format compatible with MP30 / MP40 / MP50 and MX400 / MX500 platforms |
| Type | Compatible brand battery (replacement) |
| Technology | Li-ion (lithium-ion), rechargeable |
| Nominal voltage | 10.8 V |
| Capacity | 6 Ah |
| Energy | 64.8 Wh |
| Dimensions | 125 × 68 × 50 mm (according to distributor data sheet; M4605A packs for the IntelliVue MP platform are documented on the market with approximate dimensions 149–150 × 89 × 20 mm, check the compartment before mounting) |
| Weight | Approximately 0.5 kg |
| Device manufacturer | Philips |
| Recharge | Automatic, integrated in the monitor when connected to the mains |
| Installation | Direct replacement in the equipment compartment, without tools |
| Maintenance | Maintenance-free; periodic inspection in medical electronics programmes |
| Presentation | 1 unit |
| Origin | China |
Technical note: reference 04605A is the transcription cited by the distributor; the official code that Philips assigns to the battery for the IntelliVue MP platform is M4605A (internal code 989803135861), consistently confirmed in specialist distribution. Always verify the reference and chemistry printed on the removed pack before replacement.
4. Equivalences and Cross-References
The official factory reference for the Li-ion pack of the IntelliVue MP platform (MP20/MP30/MP40/MP50/MX400/MX500) is M4605A, also identified with internal code 989803135861 and, in its most recent revision, 989803210521. Philips does not publish an official cross-reference table authorising third-party batteries; the listed alternatives share chemistry (Li-ion) and voltage (10.8 V), but come from another manufacturer and are classified as comparable by category, not 1:1 interchangeable.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Philips (equipment manufacturer) | Li-ion battery for IntelliVue MP20 / MP30 / MP40 / MP50 / MX400 / MX500 | M4605A = 989803135861 | Exact reference equivalence (same manufacturer) | This is the official reference that this battery reproduces: Li-ion, 10.8 V, 6 Ah (≈65 Wh). Correspondence confirmed in multiple specialist distribution sources and in the Cahnos catalogue itself (original version, ref. 88888486). |
| Philips (equipment manufacturer) | Most recent revision of the IntelliVue MP pack | 989803210521 | Exact reference equivalence (same manufacturer) | Updated revision of the same pack documented by specialist distribution with 10.7 V / 7,100 mAh; verify compatibility and mechanics with the monitor before replacing. |
| Cameron Sino (Medical) | Replacement battery for IntelliVue MP20 / MP30 / MP40 / MP50 | CS-PHM500MD | Comparable by category (not interchangeable 1:1) | Li-ion 10.8 V / 7,800 mAh; replaces M4605A according to the battery manufacturer’s data sheet. Higher nominal capacity and pack dimensions of 149 × 89 × 20 mm; check connector and compartment before mounting. |
| Axcom | Compatible battery for IntelliVue MP20 / MP30 / MP40 / MP50 | MB4605A | Comparable by category (not interchangeable 1:1) | Li-ion 10.8 V / 6.6 Ah; replacement for M4605A documented in European specialist distribution catalogue; check connector and compartment mechanics. |
There is no cross-reference table validated by the manufacturer (Philips) that matches reference M4605A / 04605A with third-party batteries; consult the technical department before using in a tender. Before any replacement, verify one by one: voltage (10.8 V), chemistry (Li-ion), connector and compartment dimensions.
5. Selection tips and variants of interest
Confirm the exact monitor and reference: this battery corresponds to Philips IntelliVue MP20 monitors and platforms with the same pack format (MP30, MP40, MP50, MX400, MX500). Verify the reference printed on the removed battery (04605A / M4605A / 989803135861) and the monitor model on its plate before ordering.
Do not confuse Philips families: IntelliVue MP2 / X2 monitors use the M4607A battery (10.8 V / 1 Ah), the Viridia M2/M3/M4 platform uses the M3046A (12 V Ni-MH) and the MX40 portable monitor mounts a 3.7 V pack: none is interchangeable with this battery, even though they are also Philips equipment.
Replace when symptoms of wear appear: clearly reduced battery life or low battery warnings a few hours after charging indicate the end of the battery’s useful life. After installation, connect the monitor to the mains and allow a full charge cycle to complete before an extended session in portable mode.
Storage and recycling: if the equipment is to remain unused for several weeks, keep the battery charged in a cool, dry place and recharge it periodically to avoid degradation of the Li-ion chemistry. Deliver the exhausted pack to a collection point in accordance with WEEE regulations; never place it in the general waste container.
At Cahnos we also have the original brand version for the same equipment (ref. 88888486). For institutional orders, tenders or ongoing supply contracts, contact our technical-commercial department indicating the model, serial number and reference of the installed battery; we will confirm compatibility and offer you special conditions for volume.





Reviews