Description
1. The power that keeps your Powerdrive operational
At Cahnos we offer the replacement rechargeable battery for the Synthes Powerdrive powered surgical instrument system, corresponding to the manufacturer’s original reference 530,200 (catalogue 530200; Ref. Cahnos 88888626). It is the spare part that keeps your drill, burr or screwdriver operational without depending on the mains supply, with full continuity in orthopaedic and trauma surgery. A practical solution for operating theatres, sterilisation services, biomedical engineering and medical electronics maintenance programmes, with fast delivery and spare part management in accordance with WEEE regulations. With the original battery withdrawn from the market, this compatible replacement is the most cost-effective way to keep your fleet of Powerdrive motors in service.
2. How this battery works in the Powerdrive system
The Synthes Powerdrive line (now DePuy Synthes) is a modular, rechargeable surgical motor system in which the interchangeable accumulator is the power source for the milling, drilling and screwdriving handpieces. This replacement uses nickel-cadmium (Ni-Cd) chemistry and delivers a nominal voltage of 12 V DC with a capacity of 0.6 Ah (≈7.2 Wh). Ni-Cd technology is a mature and robust solution for powered surgical instruments: it tolerates frequent charge and discharge cycles, withstands deep discharges and maintains a stable voltage curve during milling and drilling.
Recharging is carried out exclusively with the Synthes line universal battery charger (models 530,600 / 530,601), as confirmed by the manufacturer’s own manual: the UBC II universal battery charger (05,001,204) does not accept the Power Drive 530,200 battery and must not be used with it. Replacement is carried out directly: the battery is detached from the motor handpiece and the replacement fits into the same housing, without tools or adaptations. Important: Synthes batteries must not be sterilised; they are cleaned with a cloth and returned to the charger after each use.
3. Technical Specifications
Data from the replacement’s distribution data sheet (SKU 88888626), from the specifications registered for original reference 530,200 in the GUDID/FDA regulatory database and from DePuy Synthes technical documentation (universal charger line).
| Parameter | Specification |
| Reference Cahnos (SKU) | 88888626 |
| Original reference replaced | 530,200 / 530200 (Synthes GmbH; “Battery for Power Drive”, GUDID/FDA) |
| Device manufacturer | Synthes / DePuy Synthes (Powerdrive system) |
| Type | Compatible spare part (replacement rechargeable battery) |
| Technology / chemistry | Ni-Cd (nickel-cadmium), rechargeable |
| Nominal voltage | 12 V |
| Nominal capacity | 0.6 Ah (600 mAh) |
| Approximate energy | ≈ 7.2 Wh (12 V × 0.6 Ah) |
| Application | Synthes Powerdrive surgical instruments (milling, drilling, screwdriving) |
| Compatible chargers | Synthes universal battery charger 530,600 / 530,601 · Not compatible with UBC II (05,001,204) according to DePuy Synthes IFU |
| Installation | Direct replacement in the motor handpiece housing, without tools |
| Regulatory classification | Class I (FDA) · GMDN “Battery box” · FDA product code MOQ “Battery, Replacement, Rechargeable” · Primary DI 10886982239785 · Secondary DI H6795302000 |
| Commercial status of the original | Out of commercial distribution as an active line (GUDID/FDA) |
| Presentation | 1 unit |
| Recycling | Contains cadmium (Cd) · Directive 2006/66/EC · separate collection (WEEE) |
4. Equivalences and Cross-References
Original reference 530,200 is verified in the GUDID/FDA regulatory database: “BATTERY FOR POWER DRIVE” by Synthes GmbH (catalogue number 530200, GMDN battery box, class I, FDA product code MOQ “Battery, Replacement, Rechargeable”, primary DI 10886982239785, secondary H6795302000), currently out of commercial distribution as an active line. The device manufacturer does not publish an official cross-reference table with third-party batteries; the following cross-references share Ni-Cd chemistry, 12 V voltage and Powerdrive application (type 530200) with this replacement and have been checked against the replacement manufacturer’s data sheet (Akkuplanet / AKKUmed), the battery manufacturer’s catalogue (Omniacell Tertia) and the specialised distribution data sheet (Dislamp Medical), but they must be validated field by field (capacity, connector and housing dimensions) before replacement.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Synthes GmbH (device manufacturer) | Original pack “BATTERY FOR POWER DRIVE” | 530,200 (cat. 530200) | Comparable by category (not interchangeable 1:1) | Original reference of the equipment that this compatible replacement substitutes: same Ni-Cd chemistry and 12 V; registered in GUDID (primary DI 10886982239785, secondary H6795302000). Do not confuse with the large 14.4 V battery. |
| AKKUmed (Akkuplanet GmbH) | NC conversion for Synthes type 530200 | 110231-U (GTIN 4064295005344) | Comparable by category (not interchangeable 1:1) | Ni-Cd 12 V / 1.2 Ah, ≈30 × 50 × 75 × 101 mm and ≈0.47 kg, CE compliant according to the replacement manufacturer’s data sheet; same chemistry and voltage, higher capacity (1.2 vs 0.6 Ah). It is a conversion kit: requires sending the original battery. Check connector and fit. |
| Omniacell Tertia | Ni-Cd battery for Synthes Powerdrive | 1100323 | Comparable by category (not interchangeable 1:1) | Ni-Cd 12 V / 1.2 Ah, original reference code 530200 and dimensions 30 × 50 × 75 × 101 mm according to the replacement part manufacturer’s datasheet; same chemistry and voltage, higher capacity (1.2 vs 0.6 Ah). Check connector and housing dimension. |
| Dislamp Medical (specialised distribution) | Synthes Powerdrive surgical drill battery | BT42755Y | Comparable by category (not interchangeable 1:1) | Ni-Cd 12 V / 1.25 Ah and ≈0.455 kg according to distribution datasheet; compatible with the Synthes Powerdrive surgical drill (530200). Same chemistry and voltage, higher capacity (1.25 vs 0.6 Ah). Check connector and fit. |
There is no equivalence table validated by the device manufacturer (Synthes / DePuy Synthes) cross-referencing this reference with third-party batteries under 1:1 replacement conditions; the cross-references indicated have been checked against manufacturer and specialised distribution datasheets, but must be validated one by one (voltage, chemistry, capacity, connector and compartment dimensions) before replacement. Consult the technical department before use in tenders.
5. Selection tips and variants of interest
Verify the exact model of the device: this 12 V battery is intended for the Synthes Powerdrive system. Check the label on the removed battery: if it shows 530,200 or 530200, it corresponds to the original reference that this replacement replaces; contact our technical-commercial department to confirm the fit before placing the order. Do not confuse the 12 V pack with the large 14.4 V battery (532,033) or with the Colibri battery (532,003): they have different voltage, capacity and geometry and are not interchangeable (in our catalogue you will find the Colibri replacement under Ref. 88888625). Use the correct charging station: this replacement is Ni-Cd and is recharged with the Synthes universal battery charger (530,600 / 530,601); note that the UBC II charger (05,001,204) does not accept the 530,200 battery and must not be used for this pack. Replace upon signs of wear: clearly reduced runtime or loss of power during milling indicate the end of the battery’s service life; with Ni-Cd chemistry, perform full charge cycles periodically and avoid leaving the device for long periods with the battery completely discharged. Environmental management: take the exhausted battery to a specialised collection point (RAEE/WEEE); do not dispose of it in the general waste container. Institutional purchases: for drill fleets, maintenance contracts or public tenders, request special conditions, availability and documentation (declaration of conformity and safety data sheet) from our technical-commercial team.





Reviews