Description
1. The power that keeps your Synthes Colibri drill operational
At Cahnos we offer the replacement rechargeable battery for the Synthes Colibri I surgical drill, corresponding to the manufacturer’s original reference 532,003 (Ref. Cahnos 88888625). It is the replacement that ensures your equipment remains operational without depending on the mains supply: milling, drilling and screwdriving in orthopaedic and trauma surgery with total continuity. A practical solution for operating theatres, sterilisation services, biomedical engineering and maintenance programmes for powered instruments, with fast delivery and replacement management in accordance with WEEE regulations.
2. How this battery works in the Colibri drill
The Colibri line from Synthes (now DePuy Synthes) is a compact and modular surgical motor system powered by interchangeable batteries. The small 12 V DC accumulator (reference 532,003, with housing 532,002) is the power source for the Colibri I handpieces and, according to the manufacturer’s documentation, remains compatible with the Colibri II (532,101) handpiece.
This replacement uses nickel-cadmium (Ni-Cd) chemistry and delivers a nominal voltage of 12 V with a capacity of 0.6 Ah (≈7.2 Wh). Ni-Cd technology is a mature and robust solution for powered surgical instruments: it tolerates frequent charge and discharge cycles, withstands deep discharges and maintains a stable voltage curve during milling and drilling. Recharging is carried out with the Colibri line charger (universal battery charger), which also accepts the large 14.4 V battery (532,033). Replacement is carried out directly: the battery is detached from the motor handle and the replacement fits into the same housing, without tools or adaptations.
3. Technical Specifications
Data cross-checked against the technical distribution sheet for the replacement (same SKU 88888625) and against the specifications recorded for the original reference 532,003.
| Parameter | Specification |
|---|---|
| Reference Cahnos (SKU) | 88888625 |
| Original reference replaced | 532,003 (Synthes GmbH; catalogue number 532003) |
| Device manufacturer | Synthes / DePuy Synthes (Colibri line) |
| Type | Brand-compatible replacement (quality generic battery) |
| Technology / chemistry | Ni-Cd (nickel-cadmium), rechargeable |
| Nominal voltage | 12 V |
| Nominal capacity | 0.6 Ah (600 mAh) |
| Approximate energy | ≈ 7.2 Wh (12 V × 0.6 Ah) |
| Dimensions | 67 × 37 × 90 mm (according to the replacement distribution sheet; verify dimension against the removed part) |
| Approximate weight | ≈ 0.196 kg |
| Application | Synthes Colibri I surgical drill; compatible with Colibri II handpiece (532,101) |
| Recharge | Colibri line charger (universal battery charger; also accepts 532,033) |
| Installation | Direct replacement in the motor housing, without tools |
| Presentation | 1 unit |
4. Equivalences and Cross-References
The original reference 532,003 is verified in the GUDID/FDA regulatory database: «SMALL BATTERY DRIVE 12V BATTERY» from Synthes GmbH (catalogue number 532003, GMDN secondary battery, class I, product code HWE), currently out of commercial distribution. The device manufacturer does not publish an official cross-reference table with third-party batteries; the following cross-references share chemistry, voltage and housing with this replacement and have been cross-checked against manufacturer and specialised distribution data sheets, but must be validated field by field before replacement.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Synthes GmbH (device manufacturer) | Original pack «SMALL BATTERY DRIVE 12V BATTERY» (Colibri I) | 532,003 (cat. 532003) | Comparable by category (not interchangeable 1:1) | Original reference of the equipment that this compatible replacement replaces: same Ni-Cd chemistry, 12 V and 0.6 Ah; registered in GUDID (primary DI 10886982240347, secondary H6795320030). Do not confuse with the large 14.4 V battery (532,033). |
| AKKUmed | NC battery for Synthes Colibri I type 532,003 | 110232-A / AKKP-110232-A | Comparable by category (not interchangeable 1:1) | Ni-Cd 12 V / 0.7 Ah, ≈67 × 37 × 116 mm and ≈0.23 kg, CE compliant according to the replacement manufacturer’s data sheet; same chemistry and voltage as this reference, higher capacity (0.7 vs 0.6 Ah). Verify connector and housing dimension. |
| Axcom Battery Technology | Battery for Synthes type 532,003 | MB1373 (manufacturer’s specification sheet) | Comparable by category (not interchangeable 1:1) | Ni-Cd, 67 × 37 × 116 mm and ≈0.21 kg, high-quality contacts according to the manufacturer; same chemistry and voltage as this reference, capacity stated in specification MB1373. Check connector and fit. |
| Euro Energy (specialised distribution) | Medical battery compatible with Colibri Surgical Drill | 02078 | Comparable by category (not interchangeable 1:1) | Ni-Cd 12 V / 600 mAh, 67 × 37 × 91 mm and ≈200 g; same voltage, chemistry and capacity as this reference, slightly different height. Check dimension and connector. |
There is no equivalence table validated by the device manufacturer (Synthes / DePuy Synthes) that cross-references this reference with third-party batteries under 1:1 replacement conditions; the indicated cross-references have been checked against manufacturer datasheets and specialised distribution, but must be validated one by one (voltage, chemistry, capacity, connector and compartment dimensions) before replacement. Consult the technical department before use in tenders.
5. Selection tips and variants of interest
- Check the exact model of the device: this 12 V battery is indicated for the Colibri I (reference 532,003, housing 532,002) and is compatible with the handpiece Colibri II (532,101) according to the manufacturer’s documentation.
- Do not confuse the small 12 V pack with the large 14.4 V battery (532,033, with housing 532,032): they have different voltage and geometry and are not interchangeable. Check the label of the removed battery before ordering.
- If the removed battery shows the number 532,003 or 532003, it corresponds to the original reference that this replacement substitutes; contact our technical-commercial department to confirm the fit before placing the order.
- Respect the charger’s chemistry: this replacement is Ni-Cd and is recharged with the Colibri line charger; confirm that the centre’s charging station accepts the chemistry of the received batch.
- Replace upon signs of wear: clearly reduced runtime or loss of power during drilling indicate the end of the battery’s useful life. With Ni-Cd chemistry, perform full charge cycles periodically and avoid leaving the equipment for a long time with the battery completely discharged.
- Environmental management: take the spent battery to a specialised collection point (RAEE/WEEE); do not dispose of it in the general waste container.
- Institutional purchasing: for drill fleets, maintenance contracts or public tenders, request special conditions, availability and documentation (declaration of conformity and safety data sheet) from our technical-commercial team.





Reviews