Description
1. The power source that keeps the Avea ventilator always operational
The backup battery of the Avea ventilator from the Sebac / Bird platform (Viasys Healthcare) is a critical component in any intensive care unit: when the mains power fails or the patient must be transferred, the equipment must continue ventilating without interruption. This Ni-MH replacement, 24 V and 4.5 Ah (108 Wh), restores the Avea’s full autonomy for intra-hospital patient transport and for maintaining ventilation without interruptions during power cuts.
It is supplied per unit and installs directly into the equipment’s original housing, reusing the existing connector and wiring. It is a solution designed for institutional purchasing: reliable, rechargeable, maintenance-free and available so that the technical service always has a spare on hand and the ventilator never goes out of service.
2. How this battery works in the Avea ventilator
The Avea is an intensive care ventilator designed to provide continuous respiratory support to adult, paediatric and neonatal patients, so backup power is not optional: the equipment’s internal circuit keeps the pack charged while connected to the mains and automatically switches to battery when the main supply is interrupted or the patient is transferred.
This battery uses nickel-metal hydride (Ni-MH) chemistry, rechargeable and virtually maintenance-free, with a nominal voltage of 24 V and a capacity of 4.5 Ah (4500 mAh), which translates into 108 Wh of available energy. As it is the equipment’s own backup battery, replacement is straightforward: the depleted pack is removed and the new one is inserted into the same housing, without the need for adapters or modification of the wiring.
3. Technical Specifications
| Reference Cahnos (SKU) | 88889098 |
| Original reference (MFR) | 68339 (documented variants 68339A and 68339K) |
| Application | Avea ventilator from Sebac / Bird (Viasys Healthcare; Avea Vent, Aveo) |
| Battery type | Ni-MH (nickel-metal hydride), rechargeable |
| Nominal voltage | 24 V |
| Capacity | 4.5 Ah (4500 mAh) |
| Available energy | 108 Wh (24 V × 4.5 Ah) |
| Use | Battery life for intra-hospital transport and backup in the event of a power outage |
| Presentation | Unit |
| Maintenance | Maintenance-free (rechargeable) |
| Recycling | Management in accordance with WEEE regulations |
4. Equivalences and Cross-References
The original reference 68339 corresponds to the internal battery of the Avea ventilator from Bird Products Corporation / Viasys Healthcare (later CareFusion and Vyaire Medical). The 68339K variant is registered in the GUDID/FDA regulatory database with the literal description “Battery Assembly, 24V, 4.5AH, NiMH, AVEA” (primary identifier 10190752174801, company Vyaire Medical, Inc., class II, in commercial distribution). The third-party references listed below belong to the same product category (24 V Ni-MH battery for the Avea ventilator), have been cross-checked in battery manufacturer catalogues or specialised distribution catalogues, and are classified as comparable by category, not interchangeable 1:1: they require verification of the connector, the compartment dimensions and the equipment’s charging protocol before replacement.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Bird / Viasys Healthcare (CareFusion / Vyaire) | Original internal battery for the Avea ventilator | 68339 (variants 68339A and 68339K) | Comparable by category (not interchangeable 1:1) | Original part of the equipment that this replacement substitutes; the 68339K variant is registered in GUDID/FDA as a 24 V and 4.5 Ah Ni-MH battery for the Avea. Cross-check with the part number stamped on the installed battery and verify connector and dimensions. |
| R&D Batteries | Compatible replacement for Avea Vent / Aveo | 6068 | Comparable by category (not interchangeable 1:1) | Documented in the battery manufacturer’s catalogue as a replacement for 68339 / 68339A at 24 V and 4.5 Ah Ni-MH, with a published cross-reference table. Confirm connector, compartment dimensions and charging protocol. |
| Cameron Sino | Medical spare parts series for ventilators (AVEA) | CS-AVS683MD | Comparable by category (not interchangeable 1:1) | Documented in manufacturer listings as a replacement for 68339, 68339A, 68339K, AMED0013 and B11407; the published capacity is 4000 mAh (96 Wh) and the dimensions are 136 × 93 × 33 mm with 1033 g, so it differs in capacity from this reference (4.5 Ah). Verify capacity, connector and dimensions. |
| OSI Batteries / MK Battery | Avea ventilator battery replacement | B11407 / BMED11407 | Comparable by category (not interchangeable 1:1) | Documented cross-references for the Avea’s 24 V Ni-MH pack in specialised distribution tables and medical spare parts catalogues. Confirm connector and compartment dimensions before use. |
| Alpha Source | Avea Vent battery replacement | AS36068-1 | Comparable by category (not interchangeable 1:1) | Published cross-reference for the Avea’s 24 V Ni-MH 4.5 Ah pack in specialised distribution equivalence tables. Verify connector, dimensions and the equipment’s charging protocol. |
There is no equivalence table validated by the equipment manufacturer (Sebac / Bird / Viasys Healthcare / Vyaire) that authorises third-party batteries for the Avea ventilator; consult the technical department before use in tenders. Before a replacement, the following must be verified one by one: voltage (24 V), capacity (4.5 Ah), chemistry (Ni-MH), connector type and compartment dimensions.
5. Selection tips and variants of interest
Identify the exact model of your equipment: confirm that your ventilator is from the Avea platform (Sebac / Bird, Viasys Healthcare; also documented as Avea Vent and Aveo) and note the part number of the installed battery (68339, 68339A or 68339K) before ordering the replacement.
Do not interchange with other voltages or chemistries: each Bird ventilator accepts a specific voltage; do not use the 12 V pack from the VELA (ref. 88888553) or the 13.2 V pack from the Fabian (ref. 88888693), as the Avea’s charging circuit and firmware are calibrated for 24 V.
Ni-MH maintenance: avoid repeated deep discharges, keep the battery charged if the device is to remain idle for a prolonged period, and replace the pack when transport autonomy is appreciably reduced or the device warns of low battery.
Purchase planning: in units with a high rate of transfers, it is recommended to have a spare pack per device so that service is not interrupted during recharging. For institutional orders, tenders or continuous supply contracts, contact our technical-commercial department to confirm compatibility, lead times and a formal quotation.





Reviews