Description
1. Product Presentation
The 6V 3.4Ah battery (Ref. 88888299) is the power replacement indicated for the volumetric infusion pumps of the CareFusion IVAC 598 family, equipment widely used in hospitals, clinics and medical transport for the controlled administration of sera, drugs and intravenous solutions. It is a sealed lead-acid (SLA/VRLA) accumulator, rechargeable and maintenance-free, which restores the equipment’s full autonomy without specialised tools. A replacement solution designed for institutional purchasing: reliable, direct-fit and available so that the technical service can always keep the pump fleet operational.
2. Technical explanation
The IVAC 598 pump is a volumetric linear peristaltic pumping device that automatically regulates the flow of intravenous solutions at the programmed rate. When not connected to the mains — during patient transfers, supply interruptions or inter-hospital transport — all the control electronics and the pumping mechanism are powered exclusively by the internal battery. The official service manual for the device (Cardinal Health, ref. 1000SM00016) specifies a sealed lead-acid pack of 6V and 3.4Ah consisting of three cells of 1.8–2.4V in series, identified as part 129471 (BATT PACK, series 80/90).
This reference replicates exactly that configuration: it delivers a very stable voltage throughout the discharge — the pump receives a constant 6V until the reserve is exhausted — and recharges automatically as soon as the device is connected to the alternating current. According to the official documentation, a new, fully charged battery provides up to 6 hours of operation at 125 ml/h, with a recharge time from full discharge of approximately 24 hours, a charging current limited to 700mA and a maintenance voltage of about 7.02V. Replacement is carried out by accessing the battery compartment (two screws), disconnecting the connector and housing the new unit; a quick process that does not require advanced electronics knowledge.
3. Technical Specifications
| Characteristic | Specification |
|---|---|
| Reference Cahnos (SKU) | 88888299 |
| Battery type | Sealed lead-acid (SLA/VRLA) rechargeable, maintenance-free |
| Nominal voltage | 6 V (3 cells of 1.8–2.4 V in series) |
| Nominal capacity | 3.4 Ah |
| Reference for the equipment’s original battery | 129471 (BATT PACK, series 80/90 — IVAC 597/598 service manual, ref. 1000SM00016, Appendix B) |
| Cell format | NP2.8-6 (according to technical data sheet; approx. 140 × 38 × 60 mm; verify compartment) |
| Approximate weight | 0.59 kg (according to technical data sheet) |
| Recharge | Automatic when the pump is connected to mains (current limited to 700mA; maintenance voltage ≈7.02V) |
| Battery life | Up to 6 hours at 125 ml/h (new and fully charged battery, according to service manual) |
| Recharge time | Approximately 24 hours from full discharge |
| Documented variant | Series 598 with hood; confirm the equipment variant before ordering |
| Compatible equipment | CareFusion IVAC 598 volumetric pump (also documented for IVAC 580/581/590/591/597 and 281 GC regulator) |
| Installation | Directly in the equipment’s battery compartment, without specialised tools |
| Use | Intravenous infusion and perfusion in hospitals, clinics and medical transport |
| Presentation | 1 unit |
| Recycling | Manage in accordance with WEEE regulations |
Technical note: the device service manual (1000SM00016) considers the battery voltage test (Test 16) valid with readings above 5.7V and recommends replacing the accumulator if the voltage is below 6.0V or when the autonomy is noticeably reduced. The device’s charging circuit limits the current to 700mA and maintains the charging voltage at 7.02 ± 0.05V when idle; replacement must be followed by a full 24-hour charge.
4. Equivalences and Cross-References
Reference 88888299 is documented in the public catalogue of a medical battery manufacturer with ISO 13485 certification and CE declaration of conformity as a 6V 3.4Ah lead replacement for the IVAC series 598 pumps (hooded variant). The original battery of the device appears in the official service manual with part number 129471 (BATT PACK, series 80/90). Below are listed the references verified against manufacturer or specialised distribution datasheets; third-party ones are classified as comparable by category and require verification of connector, terminals and with/without hood variant before replacement.
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| CareFusion / Cardinal Health (Alaris Products) | Original device battery (BATT PACK, series 80/90) | 129471 | Exact reference equivalence (original part from the equipment manufacturer) | Official part listed in the IVAC 597/598 service manual (1000SM00016, Appendix B): sealed lead-acid 6V 3.4Ah. Reference 88888299 is the replacement that replaces it. |
| XCell (AccuCell) | Lead battery for IVAC 581/591/597/598 and regulator 281 GC | AKKP-110044 | Comparable by category (not interchangeable 1:1) | SLA 6V 3.4Ah maintenance-free, approx. 140 × 38 × 60 mm and 0.612 kg, verified in distribution datasheet (Galaxus/Digitec). Generic format: verify connector, terminals and with/without hood variant. |
| Eurosep (distributor) | Lead battery for IVAC/ALARIS 581/591/598 and INFUS 281 GC | BPAIV581 | Comparable by category (not interchangeable 1:1) | Distribution code for 6V 3.4Ah lead battery of the same pump family; it appears discontinued in the catalogue consulted: verify availability and connector. |
| Panasonic / Exide (AGM cell) | Medical cell 6V 3.4Ah for ALARIS/IVAC 580/581/590/591/598 | LCR063R5PG (cell variant LCR063R4PG) | Comparable by category (not interchangeable 1:1) | AGM lead cell 6V 3.4Ah, 134 × 34 × 60 mm, diagonal terminals, documented by specialised distributor for the same IVAC pump family; battery only, without hood: check connector and device variant. |
| UPG (Universal Power Group) | Generic SLA 6V cell | UB634 | Comparable by category (not interchangeable 1:1) | Sealed lead-acid/AGM 6V 3.4Ah maintenance-free battery with F1 terminals, generic industrial cell; not documented by its manufacturer for the IVAC family. Verify connector, terminals and with/without hood variant before replacing. |
Note on unverified codes: in the resale parts markets, codes 17048253 (battery) and 17095862 (battery with hood) circulate for the 598 series, without official confirmation from the device manufacturer; they are not listed as equivalents as they could not be verified against a primary source. There is no equivalence table validated by the device manufacturer (CareFusion / BD) that authorises the direct cross-reference between reference 88888299 and the third-party batteries cited; consult the technical department before using in tenders or replacing the battery of a device in service. Before a replacement, the following must be verified one by one: voltage (6V), capacity (3.4Ah), chemistry (sealed lead-acid), format and connector of the compartment, and the with or without hood variant of your unit.
5. Selection tips and variants of interest
- Identify the device variant: reference 88888299 is documented for the 598 series in its hooded variant (hood). Confirm whether your pump is with or without hood before placing the order, as there are cell-only versions on the market.
- Complete the fleet: the same 6V 3.4Ah battery is documented for the IVAC 580/581/590/591/597 pumps and the 281 GC regulator; verify the compartment and connector of each model.
- Respect the voltage and chemistry: the device is designed for sealed lead-acid 6V; do not install batteries of other chemistries (Ni-MH, Li-ion) or other voltages without prior technical validation, even if the connector looks similar.
- Preventive maintenance: give a full 24-hour charge after installation, after storage periods longer than 3 weeks and when receiving a new pump (recommendation from the service manual). Replace the battery when the autonomy decreases or the device warns of low battery (batt. / P4); Test 16 must exceed 5.7V.
- Recycling: lead batteries must be handed in at authorised collection points in accordance with the RAEE/WEEE regulations; they must not be disposed of as urban waste.
- Institutional orders: if you manage a mixed fleet of infusion pumps, at Cahnos we have spare parts for other families of pumps and infusion devices; contact the technical-commercial department to confirm compatibility, lead times and special conditions by volume.