Description
1. Commercial Introduction
At Cahnos we make available to you the rechargeable 7.2V and 1.9Ah battery for the volumetric pump B. Braun Infusomat FMS (Ref. Cahnos 88888062), a nickel-cadmium (Ni-Cd) accumulator that replaces the original reference 8715548 from the device manufacturer. This spare part keeps the infusion running when the device is not connected to the mains, preventing treatment interruptions in anaesthesia, intensive care, oncology, emergency and hospitalisation units. Available with delivery in 24/72 h and suitable for medical electronics maintenance programmes, tenders and institutional orders.
2. Technical explanation of the equipment and the battery
The B. Braun Infusomat FMS (Infusomat® fmS) is a high-precision volumetric infusion pump, designed for the controlled administration of large and small volumes (fluids, drugs, transfusions and enteral nutrition) in anaesthesia, intensive care and hospital ward settings. To guarantee therapy continuity during transfers or in the event of a power supply cut, the device incorporates an internal rechargeable battery that recharges automatically while the pump remains connected to the mains.
This battery is a rechargeable nickel-cadmium (Ni-Cd) pack, made up of 6 cells of 1.2V in series (Cs format) delivering a nominal voltage of 7.2V, with a nominal capacity of 1.9Ah (1900 mAh) and an approximate nominal energy of 13.7 Wh. It replaces the original reference from the device manufacturer, the 8715548. When the autonomy is clearly reduced or the device warns of low battery prematurely, it is time to replace the accumulator with this reference.
3. Technical Specifications
| Parameter | Specification |
| Reference Cahnos (SKU) | 88888062 |
| Original equipment reference | 8715548 (B. Braun); other codes of the same battery family of the Infusomat FMS cited on the market: 34502556 (original battery 7.2V / 1.2Ah, complete with base plate and screw) |
| Accumulator type | Ni-Cd (nickel-cadmium), rechargeable |
| Pack configuration | 6 cells of 1.2V in series (Cs format) |
| Nominal voltage | 7.2 V |
| Nominal capacity | 1.9 Ah (1900 mAh) |
| Nominal energy | ≈ 13.7 Wh |
| Approximate dimensions | 69 x 43 x 46 mm (per distribution technical data sheet) |
| Approximate weight | ≈ 0.31 kg |
| Compatibility | Volumetric infusion pump B. Braun Infusomat FMS (Infusomat® fmS) |
| Recharge | Automatic, integrated in the pump when connected to the mains |
| Installation | Direct replacement in the equipment compartment; verify fixing via the base plate and screw of the original pack |
| Use | Continuous infusion and perfusion, hospital, outpatient and medical transport settings |
| Presentation | 1 unit |
4. Equivalences and Cross-References
The 88888062 battery replaces the original reference 8715548 from the device manufacturer (B. Braun), a code consistently used by technical spare parts distribution to list the 7.2V / 1.9Ah (Cs format) battery of the Infusomat FMS and verified in distribution technical data sheets. Some distribution data sheets also cite the code 33175551 for this same pack, but this numbering appears in registration databases (GUDID/FDA) linked to the Vista Basic pump family from B. Braun, so it must be confirmed with the manufacturer before accepting it as valid for the Infusomat FMS. The alternatives from other battery manufacturers cited below have been verified against technical data sheets from battery manufacturers and authorised distribution catalogues; they are not official references from B. Braun and must be validated before replacement (voltage, capacity, dimensions, connector and fixing system according to batch).
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| B. Braun (device manufacturer) | Original Infusomat FMS/P battery (Ni-Cd, with base plate and screw) | 34502556 | Comparable by category (not interchangeable 1:1) | Original pack of the device, Ni-Cd 7.2V / 1.2Ah, approx. 66 x 43 x 44 mm, ≈ 0.25 kg; same chemistry and voltage as our ref. 88888062 but lower nominal capacity (1.2 Ah vs 1.9 Ah); code verified in the battery manufacturer’s technical data sheet and cited in GUDID in relation to the Vista Basic family from B. Braun; verify the numbering engraved on the installed battery. |
| Axcom Power Solutions | Battery for B. Braun Infusomat FMS / P | MB1125P | Comparable by category (not interchangeable 1:1) | Ni-Cd 7.2V / 1.2Ah, 44 x 66 x 43 mm, 0.25 kg, with declaration of conformity; replaces the original ref. 34502556; verified in the manufacturer’s data sheet; capacity lower than 1.9Ah |
| Cameron Sino | Medical battery for B. Braun Infusomat fmS / Infusomat P | CS-BRA110MD | Comparable by category (not interchangeable 1:1) | Ni-MH 7.2V / 3000mAh (21.6 Wh), ≈ 97 x 51 x 50 mm, ≈ 0.34 kg; replaces part numbers BRA110, 120089, MB1125P and 110089-S; verified in authorised distribution catalogues; different chemistry (Ni-MH vs Ni-Cd) and capacity, larger dimensions |
There is no equivalence table validated by the manufacturer (B. Braun) that cross-references this reference with third parties; the references indicated have been verified against technical data sheets from battery manufacturers, registration databases and authorised distribution catalogues, but the 1:1 replacement must be validated field by field (chemistry, voltage, capacity, dimensions, connector type, base plate and fixing screw) before being used in tenders. Consult the technical department.
5. Selection tips and variants of interest
- Confirm the exact pump model: this battery is the one intended for the volumetric pump B. Braun Infusomat FMS (original ref. 8715548). Do not use it in other models of the B. Braun family with a different voltage or connector (e.g., Perfusor Space, Compact, FM/D or FT/P).
- Respect the voltage: do not use batteries of other voltages (3.6V, 4.8V, 9.6V, etc.) in the Infusomat FMS even if the connector may look similar.
- Replace upon signs of wear: clearly reduced battery life or low-battery warnings within a few hours of charging indicate that the battery has reached the end of its service life.
- Always order by reference: quote code 88888062 or the original equipment reference (8715548) to avoid confusion with other batteries in the B. Braun family (the original 1.2Ah one is identified as 34502556).
- For mixed fleets: in our catalogue of batteries for infusion and perfusion pumps we have spare parts for Perfusor FT/P (88888066), Perfusor FM/D (88888064), Perfusor Compact (88888067) and Perfusor Space (88888697) from B. Braun, as well as for other market families (Codan, Arcomed, Fresenius, Terumo, etc.).
- Institutional orders: for tenders or continuous supply contracts, contact our technical-commercial department to confirm compatibility and exact reference with a formalised quotation.





Reviews