Description
Battery 3.6V 700mAh for B. Braun Compact perfusor
Keep your fleet of syringe pumps operational with the rechargeable 3.6V 700mAh battery for the B. Braun Perfusor Compact (Ref. Cahnos 88888067). This accumulator provides a stable and safe power source for the continuous administration of medication in hospitals, clinics, home care and medical transport, preventing treatment interruptions when the device is not connected to the mains. A reliable spare part, with direct fit and proven in demanding clinical environments.
Technical explanation of the equipment and the battery
The B. Braun Perfusor Compact is an electronic syringe infusion pump for professional use, designed to administer fluids and drugs (anaesthetics, sedatives, analgesics, catecholamines, anticoagulants, etc.) intermittently or continuously via clinically accepted routes. To ensure continuity of therapy even when the patient is mobilised or during transfers without mains power, the device has an internal rechargeable battery that automatically recharges while the pump remains connected to the mains.
This 3.6V and 700mAh (0.7Ah) battery is a nickel-cadmium (Ni-Cd) pack, made up of 1.2V cells in series, which delivers the nominal voltage required by the Compact perfusor and powers the syringe advance motor and control circuits. It replaces the original reference of the device manufacturer, 34501690 (B. Braun). When battery life is noticeably reduced or the device warns of low battery prematurely, it is time to replace the accumulator with this reference.
Technical specifications
| Parameter | Value |
|---|---|
| Reference Cahnos (SKU) | 88888067 |
| Accumulator type | Ni-Cd (nickel-cadmium), rechargeable |
| Nominal voltage | 3.6 V |
| Nominal capacity | 700 mAh (0.7 Ah) |
| Compatibility | Syringe pump B. Braun Perfusor Compact and Compact S |
| Original equipment reference | 34501690 (B. Braun); internal pack code 3/N600AAK cited on the market |
| Recharge | Automatic, integrated in the pump when connected to the mains |
| Installation | Direct replacement in the equipment compartment, without tools |
| Use | Continuous infusion and perfusion, hospital, ambulatory and medical transport settings |
| Presentation | 1 unit |
Equivalences and cross-references
The 88888067 battery replaces the original reference 34501690 of the device manufacturer (B. Braun, catalogue “Rechargeable battery pack PERF.COMPACT”). The alternatives from other battery manufacturers cited below have been cross-checked against technical data sheets from battery manufacturers and authorised distribution catalogues; they are not official references from B. Braun and must be validated before replacement (voltage, connector and pack dimensions according to batch).
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Axcom Power Solutions | Battery for B. Braun Perfusor Compact | MB1120 | Comparable by category (not interchangeable 1:1) | Ni-Cd 3.6V / 600mAh, 56 x 14 x 50 mm, 80 g, OEM 8714991, with declaration of conformity; verified in the manufacturer’s data sheet; capacity lower than 700mAh. |
| Cameron Sino | Medical battery for B. Braun Perfusor Compact | CS-BRA125MD | Comparable by category (not interchangeable 1:1) | Ni-MH 3.6V / 1500mAh, 56.36 x 49.24 x 24.30 mm; replaces references BRA125, 120015, BATT/110015 and SE-306122; verified in authorised distribution catalogue; different capacity and chemistry (Ni-MH vs Ni-Cd). |
In addition, brands such as Artisan Power offer compatible replacements for the Perfusor Compact (Ni-MH 3.6V / 2100mAh, which replace part numbers MB1120 and 3/N600AAK, manufactured under ISO 13485); their product codes are not cited in the table as they do not provide a stable reference in public sources. There is no equivalence table validated by the manufacturer (B. Braun) for this reference; consult the technical department before use in tenders.
Selection tips and variants of interest
- Confirm the exact pump model: this battery is the one indicated for the B. Braun Perfusor Compact / Compact S syringe perfusor (original Ref. 34501690). Do not use it in other models of the Perfusor family with different voltage or connector (e.g., Perfusor Space, FM/D or FT/P).
- Respect the voltage: do not use batteries of other voltages (7.2V, 4.8V, etc.) in the Perfusor Compact even if the connector may look similar.
- Replace upon signs of wear: clearly reduced battery life or low battery warnings a few hours after charging indicate that the accumulator has reached the end of its service life.
- Always order by reference: quote code 88888067 or the original equipment reference (34501690) to avoid confusion with other batteries in the B. Braun family.
- For mixed fleets: in our catalogue of batteries for infusion and perfusion pumps we have spare parts for Perfusor Space (88888697), Perfusor FT/P (88888066), Perfusor FM/D (88888064) and Infusomat FMS (88888062) from B. Braun, as well as for other families on the market.





Reviews