Description
Replacement battery for TAEMA Osiris 2 ventilator
Keep your TAEMA Osiris 2 ventilator always operational with this replacement battery of 9.6 V and 1 Ah in nickel-cadmium (Ni-Cd) technology, specific to the Air Liquide respiratory and ventilation family. It is a reliable and economical replacement that directly replaces the original reference KY50900 and ensures continuity of care in emergency care, intrahospital transport, critical care units and post-operative wards. A practical solution for maintenance services, biomedical engineering and hospital purchasing departments that need standardised and traceable spare parts.
How the equipment works
The Osiris 2 is a pneumatic ventilator with controlled frequency (from 6 to 40 cycles per minute, with adjustable inspiratory/expiratory ratio from 1/1 to 1/3) developed by TAEMA, the hospital subsidiary of the Air Liquide group, for use in transport, emergency care, intensive care and post-operative wards. It operates on pure oxygen or a mixture of ambient air and oxygen, and has internal rechargeable batteries that ensure its autonomy when the equipment is not connected to the mains; the orange pilot light on the panel warns when the remaining charge only allows less than 60 minutes of operation, the recommended time to connect the equipment to the mains or change the battery pack. The original TAEMA manual describes this pack as Ni-Cd of 700 mAh made up of 8 cells of 1.2 V in series; commercial replacements in this family are usually presented between 0.94 and 1 Ah. The batteries recharge automatically while the ventilator remains connected to the mains via its specific power supply box.
Technical specifications
| Characteristic | Value |
|---|---|
| Application | TAEMA Osiris 2 Ventilator / Osiris 2 Ventilator (Air Liquide) |
| Original reference | KY50900 |
| Nominal voltage | 9.6 V |
| Capacity | 1 Ah nominal (approx. 940 mAh) |
| Technology | Nickel-Cadmium (Ni-Cd) |
| Cells | 8 cells of 1.2 V in series |
| Dimensions | 57 x 29 x 50 mm |
| Weight | 0.160 kg approx. |
| Connection | Connector with fixing tab |
| Type | Rechargeable |
Equivalences and cross-references
Code verification: the reference KY50900 consistently appears as the original reference for the battery pack of the TAEMA Osiris 2 ventilator in the technical data sheet of the authorised distributor VLAD (document 88888629_EN.pdf, Ni-Cd battery 9.6 V 940 mAh for TAEMA Osiris 2 ventilator) and in catalogues of sector distributors (Surgicall, ref. 43.888.8629, manufacturer indicated as Air Liquide). Code 88888629 is the reference under which VLAD catalogues this same battery and Cahnos uses it as the internal reference for the spare part; it also appears linked to KY50900 in international sources of compatible spare parts for the Osiris 2 ventilator. There is no GUDID/FDA registration for KY50900; treat this reference as catalogue data, not as regulatory data.
| Manufacturer / Brand | Family / Model | Reference Code / SKU | Equivalence Level | Technical Differences / Remarks |
|---|---|---|---|---|
| Cameron Sino | CS-TRS200MD (compatible with Taema Osiris 2 / Osiris 2 Ventilator) | CS-TRS200MD | Comparable by category (not interchangeable 1:1) | Replaces KY50900 / 88888629; same voltage of 9.6 V and same compatibility with Osiris 2, but Ni-MH chemistry, capacity of 2000 mAh (double) and approximate dimensions of 58.4 x 51.4 x 29.3 mm. Verify the charging electronics of the equipment before replacing. |
| AolsteCell | KY50900 Medical Battery for Osiris 2 Ventilator | AOM10983 | Not verified | Commercial replacement advertised for KY50900/88888629 in Osiris 2; the distributor’s data sheet presents contradictory data (voltage 7.2 V vs. 9.6 V and text mentioning lithium-ion cells versus declared Ni-MH, 2000 mAh, 8 cells, ABS+PC housing). Confirm voltage, chemistry and connector with the manufacturer before replacement. |
There is no equivalence table validated by the manufacturer (Air Liquide/TAEMA) for this reference; consult the technical department before using in tenders. In the market there are also compatible replacements for KY50900 from VLAD/Exalium Premium and from the Spectra range: check voltage, chemistry, capacity and connector in each technical data sheet before replacement.
Selection tips and variants of interest
- Always verify the voltage (9.6 V) and the connector type of the ventilator before replacing; the Osiris 2 compartment accepts the pack with fixing tab.
- Ni-Cd batteries are sensitive to the memory effect: perform complete charge and discharge cycles periodically and keep the ventilator connected to the mains when not in use to preserve the charge.
- For the Osiris family, Cahnos also offers the battery for the Osiris 3 ventilator (Ni-MH, 9.6 V, 1.7 Ah; ref. KY564600, Ref. Cahnos 88888630), which is not interchangeable with this unit without prior technical verification of the equipment.
- If you are looking for greater autonomy, consider Ni-MH replacements of 2000 mAh only if the equipment’s charging electronics allow it; if in doubt, consult the technical service.
- For transport and emergency equipment, it is recommended to have a reserve battery and check the low battery indicator before each departure.





Reviews
There are no reviews yet.