Description
1. Reliable power for your LIFEPAK 9 defibrillator
The battery Ref. 88888500 is the power replacement that keeps Physio-Control LIFEPAK 9, 9A, 9B and 9P defibrillators operational, classic resuscitation and monitoring equipment that continue to serve in ambulances, hospitals and emergency services. It is a sealed lead-acid (SLA/VRLA) cell of 16 V and 2.5 Ah, leak-proof and maintenance-free, which ensures the necessary power for heart rhythm monitoring and the delivery of shocks when the device operates without mains connection. An essential spare part for maintenance services, biomedical engineering and hospital purchasing departments that need standardised, traceable and economical spare parts.
2. How this battery works in your defibrillator
LIFEPAK 9 defibrillators use a sealed, rechargeable lead-acid cell as an internal power source. The battery delivers a nominal voltage of 16 V with a capacity of 2.5 Ah (≈40 Wh), sufficient to power the ECG monitoring electronics and the shock charging circuit during operation without mains power. Its sealed (VRLA) construction allows spill-proof mounting and replacement is carried out in a few minutes: the battery is inserted into the rear compartment of the device and is secured by its bayonet connector, with positive polarity alignment guaranteed. While the defibrillator remains connected to the mains, the battery recharges automatically and is ready for the next use. Like any sealed lead cell, it offers a long service life if kept charged and not subjected to repeated deep discharges; when the autonomy is noticeably reduced or the device warns of low battery prematurely, it is time to replace it.
3. Technical Specifications
| Characteristic | Specification |
|---|---|
| Reference Cahnos (SKU) | 88888500 |
| Manufacturer reference (MFR) | 803704-3 / 803704-03 (also Medtronic 21300-002259) |
| Battery type | Sealed rechargeable lead-acid (SLA/VRLA), leak-proof, maintenance-free |
| Nominal voltage | 16 V |
| Nominal capacity | 2.5 Ah (2,500 mAh / ≈40 Wh) |
| Approximate dimensions | 127.4 × 68.0 × 68.0 mm (according to compatible spare part manufacturers’ data sheets) |
| Approximate weight | ≈1.68 kg (1,676 g) |
| Mounting / connector | Rear compartment with bayonet connector and aligned polarity |
| Compatibility | Physio-Control (Medtronic) LIFEPAK 9, 9A, 9B and 9P defibrillators |
| Rechargeable | Yes, automatic recharging integrated in the device when connected to the mains |
| Application | ECG monitoring and defibrillation in emergency care, ambulances and medical transport |
| Presentation | 1 unit |
Technical note: the electrical data (16 V / 2.5 Ah) correspond to the original specification of reference 803704-03 verified in specialised distribution sources; the dimensions and weight come from data sheets of compatible spare part manufacturers and must be checked against the compartment of the specific unit before mounting.
4. Equivalences and Cross-References
Ref. 88888500 corresponds to the standard sealed lead-acid battery format of 16 V / 2.5 Ah used in LIFEPAK 9 defibrillators (includes Cyclon EnerSys type cells in the original pack). The references from other manufacturers listed below belong to this same product category, have been verified in technical data sheets from battery manufacturers or authorised distribution catalogues and are classified as comparable by category, not 1:1 interchangeable: before replacement, the voltage (16 V), capacity (2.5 Ah), chemistry (sealed lead-acid), connector type (bayonet) and compartment dimensions must be verified.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Cameron Sino | Medical battery for Medtronic Lifepak 9 / 9A / 9B / 9P | MED-CS-MPD910MD | Comparable by category (not interchangeable 1:1) | SLA 16 V / 2,500 mAh, dimensions 127.40 × 68.00 × 68.00 mm, black; replaces original references 21300-002259 and 803704-03; verified in Cameron Sino authorised distribution catalogue. |
| OSI Batteries | Biomedical battery for LifePak 9 / 9P (Upgrade) | B11053 | Comparable by category (not interchangeable 1:1) | SLA 16.0 V / 2.5 Ah that replaces original part 803704-03; documented for LifePak 9 and 9P upgrade; verify connector and pack dimensions before replacement. |
| Peakbatt | Replacement battery for Physio-Control Lifepak 9 | Compatible ref. 21300-002259 / 803704-03 | Comparable by category (not interchangeable 1:1) | SLA 16.0 V / 2,500 mAh / ≈40 Wh, dimensions 127.40 × 68.00 × 68.00 mm and ≈1,676 g; replaces 21300-002259 and 803704-03; verified in online distribution; confirm exact reference before purchasing. |
There is no equivalence table validated by the manufacturer (Physio-Control / Medtronic) that authorises third-party batteries for reference 803704-03; consult the technical department before using in tenders. Before a replacement, each of the following must be verified one by one: voltage (16 V), capacity (2.5 Ah), chemistry (sealed lead-acid), connector type (bayonet) and compartment dimensions (127.4 × 68.0 × 68.0 mm approx.).
5. Selection tips and variants of interest
Identify the exact model of your equipment: this 16 V / 2.5 Ah battery is documented for the LIFEPAK 9, 9A, 9B and 9P defibrillators. Do not interchange it with batteries from other Physio-Control series with a different voltage or connector: for example, the LIFEPAK 5 uses a 12 V / 1.9 Ah Ni-Cd cell (ref. 88888495), the LIFEPAK 12 in lead version uses 12 V / 2.5 Ah (ref. 88888504) and the LIFEPAK 20 uses 12 V / 3 Ah Ni-MH (ref. 88881006 / 88888505). Always check the compartment and the connector before replacing.
Prevention and maintenance: SLA/VRLA cells self-discharge slowly at rest; if the defibrillator is to remain switched off for long periods, keep it connected to the mains periodically to preserve the charge. Replace the battery when the runtime is appreciably reduced or the equipment warns of low battery, and check the condition of the connector at each replacement. Always order by reference: quote code 88888500 or the original reference 803704-3 / 21300-002259 to avoid confusion with other batteries in the LIFEPAK family.
For institutional orders, tenders or ongoing supply contracts, contact our technical-commercial department to confirm compatibility with your fleet of defibrillators, delivery times and a formal quotation.





Reviews