Description
1. HeartSine Pediatric-Pak battery: the all-in-one paediatric replacement
Keeping a defibrillator ready for action is a responsibility that allows no delay, and in the paediatric population every minute counts even more. With the HeartSine Pediatric-Pak (Pedi-Pak) battery 18 V / 1.5 Ah (Ref. Cahnos 88881023, factory reference PAD-PAK-02 / 11516-000004) you will have the original energy replacement for the semi-automatic external defibrillators (AEDs) of the HeartSine Samaritan PAD family.
This cartridge is the paediatric version of HeartSine’s innovative PAD-PAK system: an all-in-one module that integrates the device battery and the pre-connected pair of defibrillation electrodes in a single piece, with a single expiry date and a single component to replace. It is the solution that occupational risk prevention services, schools, sports centres, clinics, administrations and medical electronics departments need to keep their AED fleet available for cardiac emergencies in children, backed by the original manufacturer part.
2. How the HeartSine PAD-PAK system works
HeartSine Samaritan PAD defibrillators (models 300P, 350P, 360P, 450P and 500P depending on market) operate with a consumable cartridge called PAD-PAK: a module that contains the pair of defibrillation electrodes and the system battery. By integrating both consumables into a single piece, maintenance is radically simplified: there is only one expiry date (4 years from date of manufacture) and a single replacement after activation or depletion.
The Pediatric-Pak version (pink cartridge) incorporates paediatric electrodes and, when inserted into the AED, the device automatically adjusts the shock energy to ≈50 J, the recommended energy for children aged 1 to 8 years or up to 25 kg (55 lb). The battery uses lithium-manganese dioxide (LiMnO₂) chemistry, a non-rechargeable primary lithium technology with high energy density and very low self-discharge, ideal for devices that remain on standby for long periods. It delivers a nominal voltage of 18 V with 1.5 Ah capacity, sufficient for ≈60 shocks or ≈6 hours of continuous monitoring; when the remaining charge equals 10 shocks or less, the defibrillator emits an acoustic warning to plan replacement.
3. Technical Specifications
| Parameter | Specification |
| Reference Cahnos (SKU) | 88881023 |
| Original manufacturer reference | PAD-PAK-02 (11516-000004) — Pediatric-Pak™ (HeartSine Technologies Ltd) |
| Product type | All-in-one cartridge: system battery + pair of paediatric defibrillation electrodes |
| Technology | Primary non-rechargeable lithium (lithium-manganese dioxide, LiMnO₂) |
| Nominal voltage | 18 V |
| Capacity | 1.5 Ah (1,500 mAh) |
| Yield | ≈60 shocks or ≈6 hours of continuous monitoring (HeartSine specification) |
| Paediatric energy | ≈50 J (the AED adjusts the energy upon detecting the Pediatric-Pak) |
| Shelf life | 4 years from date of manufacture (single expiry date) |
| Paediatric population | Children from 1 to 8 years or up to 25 kg (55 lb); do not use on children under 1 year |
| Low battery warning | Audible device warning when ≈10 shocks remain |
| Compatibility | HeartSine Samaritan PAD 300P, 350P, 360P, 450P (350P/360P/450P in official catalogue; 300P and 500P in specialised distribution) |
| Regulatory registration | GUDID/FDA UDI-DI 05060167120473 (model PAD-PAK-02); CE Class IIb (notified body 0123) |
| Presentation | 1 unit |
| Transport of dangerous goods | UN3090 (lithium metal battery) |
| Recycling | Return the depleted unit to a WEEE collection point; do not dispose of with general waste |
Compatibility by defibrillator model:
| AED model | Compatible PAD-PAK cartridges |
| Samaritan PAD 300P | PAD-PAK-01 / PAD-PAK-03 (adult); Pediatric-Pak PAD-PAK-02 documented in specialised distribution |
| Samaritan PAD 350P | PAD-PAK-01 / PAD-PAK-03 / PAD-PAK-07 (adult) and PAD-PAK-02 (paediatric) |
| Samaritan PAD 360P | PAD-PAK-01 / PAD-PAK-03 / PAD-PAK-07 (adult) and PAD-PAK-02 (paediatric) |
| Samaritan PAD 450P | PAD-PAK-01 / PAD-PAK-07 (adult) and PAD-PAK-02 (paediatric) |
| Samaritan PAD 500P | PAD-PAK-03 (adult); Pediatric-Pak PAD-PAK-02 documented in specialised distribution |
4. Equivalences and Cross-References
The reference Cahnos 88881023 corresponds to the original battery of the HeartSine Pediatric-Pak system (factory reference PAD-PAK-02 / 11516-000004), registered in the GUDID/FDA regulatory database with UDI-DI 05060167120473 under the model «Pad-Pak-02» and the brand «Pediatric-Pak», described as a non-rechargeable paediatric defibrillation electrode for semi-automatic AEDs. The manufacturer does not market this battery as a standalone part: the official replacement route is the complete cartridge (battery + electrodes), so the manufacturer’s own references listed below use this same 18 V / 1.5 Ah lithium battery but with adult electrodes, and are classified as comparable by category, not interchangeable 1:1 in terms of clinical indication.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| HeartSine (equipment manufacturer) | Adult PAD-PAK cartridge (battery + adult electrodes) | PAD-PAK-01 (11516-000003) | Comparable by category (not interchangeable 1:1) | Same non-rechargeable 18 V / 1.5 Ah (LiMnO₂) lithium battery integrated with adult electrodes; indicated for patients over 8 years or over 25 kg, compared to 1-8 years / up to 25 kg for the Pediatric-Pak. Adult energy ≈150 J versus ≈50 J paediatric. Register in GUDID (05060167120466). Mechanically interchangeable in the AED; verify clinical indication before replacing. |
| HeartSine (equipment manufacturer) | Adult PAD-PAK cartridge (commercial variant) | PAD-PAK-03 (EAN 5060167121234) | Comparable by category (not interchangeable 1:1) | Commercial variant of the adult cartridge with the same non-rechargeable 18 V / 1.5 Ah lithium battery; documented for Samaritan PAD 300P, 350P, 360P and 500P. Adult electrodes, not paediatric; does not clinically replace the Pediatric-Pak. |
| HeartSine (equipment manufacturer) | PAD-PAK aviation cartridge (adult) | PAD-PAK-07 (11516-000027) | Comparable by category (not interchangeable 1:1) | Same 18 V / 1.5 Ah lithium battery with adult electrodes certified TSO-C142a for aeronautical use; GUDID registration 05060167124488. Indicated for adults; not the 50 J paediatric variant. |
There is no manufacturer-validated equivalence table authorising third-party cartridges or batteries for reference PAD-PAK-02; consult the technical department before use in tenders. The HeartSine cartridge integrates the electrode connector and the device’s cartridge management, so generic lithium batteries from other AEDs (e.g. cartridges and batteries from ZOLL, Defibtech, Philips HeartStart or Physio-Control Lifepak) are comparable only at the product category level, never interchangeable 1:1 with the Pediatric-Pak: they differ in chemistry, voltage, connector, geometry and integrated paediatric functionality.
5. Selection tips and variants of interest
- Confirm the exact model of the device: this cartridge is the one indicated for HeartSine Samaritan PAD 300P, 350P, 360P, 450P defibrillators (and 500P in certain markets). Check the model on the AED label and the installed cartridge code (PAD-PAK-02 / 11516-000004) before ordering.
- Respect the paediatric indication: the Pediatric-Pak is designed for children aged 1 to 8 years or up to 25 kg (55 lb) and delivers ≈50 J. Do not confuse it with the adult PAD-PAK-01 cartridge (adult electrodes, ≈150 J): although they mechanically fit the same device, the clinical indication is different and they must not be interchanged.
- Track a single date: the advantage of the PAD-PAK system is that the battery and electrodes share an expiry date (4 years from manufacture). Plan preventive replacement before expiry and after any real use of the device in a resuscitation.
- Heed the low battery warning: when the AED indicates approximately 10 shocks or fewer remain, replace the cartridge immediately to avoid compromising device availability.
- Storage and transport: as this is a lithium metal battery (UN3090), transport is carried out as dangerous goods; store in a dry place, between 0 °C and 50 °C, away from heat sources. The Pediatric-Pak contains a magnetic component (≈6500 gauss): avoid storing it near sensitive magnetic media.
- Environmental management: deliver the exhausted unit to a collection point in accordance with WEEE regulations; do not dispose of it in the general waste container.
- Institutional orders and tenders: for tenders, mixed fleets or continuous supply contracts, contact our technical-commercial department indicating the defibrillator model and the installed cartridge code; we will confirm compatibility, traceability and a formalised quotation.





Reviews
There are no reviews yet.