Description
1. Backup power for your Defibtech defibrillator
At Cahnos we offer you the replacement battery that corresponds to the original unit of the Defibtech Lifeline AED defibrillators, identified in our catalogue with reference 88888845 and with factory code DBP-1400 (also referred to as DCF-200 in distribution channels). It is a non-rechargeable primary lithium cell of 15 V and 1.4 Ah, designed to keep the AED permanently in “ready-to-use” mode for up to 5 years, without charging routines or maintenance.
It is the essential replacement for emergency services, ambulances, civil protection, workplaces, sports facilities and public access defibrillation points that manage equipment from the Lifeline AED / DSA-DEA family, ensuring an immediate response to cardiac arrest.
2. Technical explanation: how this battery works
The automated external defibrillator (AED) Lifeline AED operates exclusively on battery: while at rest, the device performs daily, weekly, monthly and quarterly self-tests that verify the condition of the cell, the electronics and the electrodes. In an emergency, the stored energy powers the rhythm analysis electronics, the voice prompts and the capacitor charge/discharge stage to deliver biphasic therapy.
This battery uses non-rechargeable primary lithium chemistry (lithium-manganese dioxide type, Li-MnO₂), single-use: its very low self-discharge allows it to remain installed for years with full response capacity. The pack also includes a 9 V auxiliary battery (model DAC-410) that independently powers the device’s self-tests, leaving the main energy reserved for assistance. When depletion approaches, the defibrillator warns via indicator light and acoustic signal; replacement is carried out by removing the depleted battery and inserting the new one into the compartment, without tools. Due to its disposable nature, stock management must be carried out by expiry date (FIFO rotation).
3. Technical Specifications
| Parameter | Specification |
|---|---|
| Reference Cahnos (SKU) | 88888845 (distribution code) |
| Manufacturer reference (MFR) | DBP-1400, also referred to as DCF-200 (Defibtech) |
| Battery type | Primary lithium cell, non-rechargeable (single-use) |
| Chemistry | Lithium (lithium-manganese dioxide type, Li-MnO₂) |
| Nominal voltage | 15 V |
| Nominal capacity | 1.4 Ah (1400 mAh) |
| Stored energy | ≈ 21 Wh (15 V × 1.4 Ah) |
| Standby autonomy | Up to 5 years installed (includes device self-tests) |
| Shock capacity | Up to 125 shocks or 8 hours of continuous monitoring |
| Auxiliary self-test battery | 9 V battery model DAC-410 (included in the pack) |
| Rechargeable | No |
| Maintenance | Maintenance-free; replace by expiry date |
| Compatibility | Defibtech Lifeline AED (DDU-100), Lifeline AUTO (DDU-120) and DSA-DEA (DDU-1XX series) |
| Low battery warning | Visual indicator and audible signal |
| Installation | Direct replacement in the compartment, no tools required |
| Regulatory registration | GUDID/FDA: UDI-DI 00815098020362 and 10815098020369 (DEFIBTECH, L.L.C., brand “DDU-1XX Series Battery Pack”, model DBP-1400, publication date 24/10/2015, Class III per FDA) |
| Presentation | 1 unit |
| Recycling | Management in accordance with WEEE regulations (regulated lithium transport) |
4. Equivalences and Cross-References
The official reference for the pack is DBP-1400, confirmed in the GUDID/FDA regulatory database in the name of DEFIBTECH, L.L.C. under the brand “DDU-1XX Series Battery Pack” with description “Primary battery” and FDA Class III classification. The designation DCF-200 corresponds to the same assembly and is the code under which it is distributed in certain channels (exact reference equivalence within the same manufacturer Defibtech). Defibtech does not publish an equivalence table authorising third-party batteries for the Lifeline AED family; the listed alternatives share the same type (lithium battery for AED) but come from other manufacturers and are classified as comparable by category, not interchangeable 1:1.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Recell (BatteryMart / AED Battery Exchange) | DBP-1400-ABE battery for Lifeline AED DDU-100 and DDU-120 | DBP-1400-ABE | Comparable by category (not interchangeable 1:1) | Aftermarket replacement that claims up to 125 shocks, 5 years of shelf life and 4 years of warranty; 15 V Li-MnO₂ chemistry, without the DAC-410 auxiliary battery standard in all versions. Validate voltage, connector and compartment housing against the removed unit. |
| AolsteCell | Replacement for Defibtech DDU-100, DBP-2800 and DBP-1400 | AolsteCell DBP-1400 | Comparable by category (not interchangeable 1:1) | Third-party replacement battery that claims compatibility with the Lifeline AED family; verify voltage (15 V), capacity (1.4 Ah), connector and compartment geometry before replacement. |
There is no equivalence table validated by the manufacturer (Defibtech) that cross-references the DBP-1400 part number with third-party batteries; consult the technical department before using in a tender. Before accepting an alternative battery, each of the following must be verified one by one: voltage (15 V), capacity (1.4 Ah), chemistry (non-rechargeable primary lithium), presence of the 9 V auxiliary battery for self-test (DAC-410), connector and compartment dimensions. Batteries from other AEDs on the market (e.g. ZOLL AED Pro 8000-0860-01, Philips HeartStart FR3 989803150161, CU Medical CUSA0601F or Welch Allyn 1852) share the same type but are not interchangeable due to the proprietary connector and management electronics.
5. Selection tips and variants of interest
- Identify the exact model: this battery is for the Defibtech Lifeline AED (DDU-100), Lifeline AUTO (DDU-120) and DSA-DEA defibrillators of the DDU-1XX series. It is not compatible with the Lifeline VIEW / ECG / PRO family (DDU-2000 series), which uses the DBP-2003 / DCF-2003 pack.
- Check the printed reference: before ordering, confirm the DBP-1400 code on the removed battery; if your device works with another pack, do not force the replacement.
- Expiry management: as this is a non-rechargeable primary battery, establish FIFO stock rotation and replace the unit before its expiry date even if the device has not recorded any shocks. Respond to the self-test low battery warnings with a new, verified unit.
- Higher-capacity variant in the same family: if your Lifeline AED requires a greater number of shocks per cycle, consider the Defibtech DBP-2800 battery with 7 years / 300 shocks (ref. Cahnos 88888931), with the same chemistry and 15 V voltage; confirm compatibility with your unit beforehand.
- Accessories: also check the condition and expiry date of the device’s defibrillation electrodes to keep the defibrillation point fully operational.
- Recycling: dispose of the exhausted battery at the authorised WEEE collection points in accordance with current regulations. For volume orders or tenders, contact the Cahnos technical department stating the model, serial number and reference of the installed battery; we will confirm compatibility before placing the order.