Description
1. Reserve power for your CU Medical i-PAD defibrillator
Keep your CU Medical i-PAD NF1200 external defibrillator (MRF family) always ready to act with this 12V 4.2Ah lithium battery that reproduces the original reference CUSA0601F (Ref. Cahnos 88888831, SKU 0159031). It is a direct replacement part with which you will restore full operability of the device, eliminate battery warnings from the self-test and ensure an immediate response in the event of cardiac arrest. A reliable, cost-effective solution designed for emergency services, first responders, ambulances, sports centres, businesses and public-access defibrillation.
2. How this battery works: technical explanation
The i-PAD is an automated external defibrillator (AED) designed to remain in standby mode and respond instantly when removed from its base. If cardiac arrest is suspected, the device analyses the patient’s heart rhythm through the defibrillation electrodes and decides whether a shock should be delivered; if so, it charges the capacitor with the stored energy and guides the rescuer through voice and visual prompts.
This battery is a primary lithium-manganese dioxide (Li-MnO₂) cell, non-rechargeable and single-use, which supplies the 12V working voltage to the analysis electronics, the instruction speaker and the discharge stage, without depending on the mains power supply. While the device remains on standby, the internal self-test periodically checks the battery status; as it approaches depletion, the indicator light warns that replacement is required. Being a primary lithium cell with very low self-discharge, it keeps the device operational for long periods without maintenance, although it does not accept recharging: replacement is carried out with a new unit in the device compartment.
3. Technical Specifications
Technical data verified against specialised distribution documentation (kingsener, BatteryWeb, Batteries4pro, Akkumed, Battery Specialists) for the original reference CUSA0601F:
| Parameter | Detail |
|---|---|
| Manufacturer reference (original) | CUSA0601F (also cited as 110604-O and WM-63010) |
| Reference Cahnos (SKU) | 88888831 (SKU 0159031) |
| Type / technology | Primary lithium-manganese dioxide (Li-MnO₂) battery, non-rechargeable, single-use |
| Nominal voltage | 12 V |
| Capacity | 4.2 Ah (depending on distribution: 4050–4650 mAh; ≈48.6 Wh) |
| Approximate dimensions | 164.34 x 60.70 x 31.45 mm (depending on distribution; verify compartment) |
| Approximate weight | ≈0.49 kg |
| Rechargeable | No (primary battery; replace with a new unit) |
| Shock capacity | Minimum 200 shocks at 150 J |
| Shelf life | Up to 5 years in standby (depending on usage conditions and frequency of self-tests) |
| Operating temperature | From -20 °C to +60 °C |
| Protection | IP54 rating |
| Compatibility | CU Medical i-PAD NF1200 Saver, i-PAD NF1201 Saver, i-PAD NF1200 Training, i-PAD SP1 / SP2 and Auto-Def (DSA i-PAD Auto-Def) |
| Mounting type | Direct replacement in the equipment compartment, without tools |
| Presentation | 1 unit |
| Recycling | Do not dispose of as municipal waste; recycle in accordance with WEEE regulations (regulated lithium transport) |
4. Equivalences and Cross-References
The original reference stamped on this battery is CUSA0601F, consistently confirmed in specialised distribution datasheets (kingsener, BatteryWeb, Batteries4pro, Akkumed, Battery Specialists, Vintrons) and in the internal reference register. The designations 110604-O and WM-63010 are repeatedly cited in those same sources as other codes for the same OEM assembly (exact reference equivalence within the same manufacturer CU Medical): same Li-MnO₂ chemistry, 12V and 4.2Ah. CU Medical does not publish an official equivalence table authorising third-party batteries for the i-PAD NF1200; the references listed below share the same type (non-rechargeable primary lithium batteries for AEDs) and are classified as comparable by category, not 1:1 interchangeable, due to different connector, management electronics and capacities:
| Manufacturer / Brand | Family / Model | Reference Code / SKU | Equivalence Level | Technical Differences / Remarks |
|---|---|---|---|---|
| ZOLL | AED Pro non-rechargeable battery | 8000-0860-01 (Cahnos 88888634) | Comparable by category (not interchangeable 1:1) | Same Li-MnO₂ chemistry, 12V and 4.2Ah, non-rechargeable; proprietary connector and management electronics of the AED Pro. |
| Philips | HeartStart FR3 | 989803150161 (Cahnos 88888783) | Comparable by category (not interchangeable 1:1) | Non-rechargeable 12V / 4.2Ah lithium for the FR3 AED; connector and power management exclusive to Philips. |
| Welch Allyn | DSA AED 10 | 1852 / 001852 (Cahnos 88888647) | Comparable by category (not interchangeable 1:1) | Primary lithium cell 12V but 1.4Ah; lower capacity and different compartment format from the i-PAD. |
| Defibtech | DSA-DEA 5 Years | DBP-1400 (Cahnos 88888845) | Comparable by category (not interchangeable 1:1) | Primary lithium battery for AED at 15V / 1.4Ah; voltage and connector different from the i-PAD (12V). |
| Defibtech | DSA-DEA 7 Years | DBP-2800 (Cahnos 88888931) | Comparable by category (not interchangeable 1:1) | Primary lithium battery 15V / 2.8Ah, up to 300 shocks or 7 years; not interchangeable with the i-PAD. |
There is no manufacturer-validated equivalence table authorising third-party batteries for the CU Medical i-PAD NF1200; consult the technical department before use in tenders. Before accepting an alternative battery, each of the following must be verified one by one: voltage (12V), capacity (4.2Ah), chemistry (primary non-rechargeable Li-MnO₂), pack dimensions and compartment connector.
5. Selection tips and variants of interest
Identify the exact model of your device: this battery is the one indicated for the CU Medical i-PAD NF1200 (MRF family) and its variants NF1201, NF1200 Training training units, SP1, SP2 and Auto-Def. Check three parameters on the label of the removed cell: voltage (12V), technology (primary non-rechargeable lithium) and capacity (4.2Ah).
Do not confuse it with other AED batteries: the batteries for ZOLL AED Pro (8000-0860-01), Philips HeartStart FR3 (989803150161), Welch Allyn AED 10 (1852) or Defibtech DSA-DEA (DBP-1400 / DBP-2800) are not interchangeable with the i-PAD due to connector, voltage and management electronics, even though they share the Li-MnO₂ chemistry.
Expiry management: as it is a non-rechargeable primary battery, establish a FIFO stock rotation and replace the unit before its expiry date even if the device has not recorded any shocks. Respond to the autotest low battery warnings with a new, verified unit.
Accessories: also check the condition and expiry date of the i-PAD defibrillation electrodes to keep the device fully operational.
Recycling: dispose of the exhausted battery at the authorised WEEE collection points in accordance with current regulations. For volume orders or tenders, contact the Cahnos technical department stating the model, serial number and reference of the fitted battery; we will confirm compatibility before formalising the order.